In order for EZ-IO® Intraosseous Vascular Access System products to perform properly, the following conditions are recommended. Failure to comply with these conditions will void any applicable warranties.
Do not immerse or use excessive amount of liquid when performing cleaning and disinfecting. In the unlikely event of a driver failure, remove the EZ-IO® Power Driver, grasp the needle set by hand, and advance the needle set into the medullary space while twisting the needle set.
If your clinical environment requires sterilization, the EZ-IO® Power Driver can be sterilized using the STERRAD® 100S System. STERRAD® Systems are products of ©ADVANCED STERILIZATION PRODUCTS, Division of Ethicon Inc., a Johnson & Johnson company.
No modification of this equipment is allowed.
Declaration-Electromagnetic Emissions | |||||
The EZ | |||||
Emission Test | Compliance | Compliance | |||
RF Emissions CISPR 11 | Group 1 | The EZ-IO® Power Driver uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause interference in nearby electronic equipment. | |||
RF Emissions CISPR 11 | Class B | The EZ-IO® Power Driver is suitable for use in all establishments, including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes. | |||
Harmonic emissions IEC 61000-3-2 | Not applicable | ||||
Voltage fluctuations/flicker emissions IEC 61000-3-3 | Not applicable |
Declarations - Electromagnetic Immunity | |||
The EZ-IO® Power Driver is intended for use in the electromagnetic environment specified below. The customer or the user of the EZ-IO® Power Driver should assure that it is issued in such an environment. | |||
Immunity test | IEC 60601 test level | Compliance Level | Electromagnetic Environment-Guidance |
Electrostatic discharge (ESD) IEC 61000-4-2 | +/- 6 kV Contact +/- 8 kV air | +/- 6 kV Contact +/- 8 kV air | Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material the relative humidity should be at least 30%. |
Electrical fast Transient/burst IEC 61000-4-4 | +/- 2 kV for power supply lines +/- 1 kV for input. output lines | Not applicable (battery powered) Not applicable (no I/O lines) | Mains power quality should be that of a typical commercial or hospital environment. |
Surge IEC 61000-4-5 | +/- 1 kV differential mode +/- 2 kV common mode | Not applicable (battery powered) | Mains power quality should be that of a typical commercial or hospital environment. |
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11 | < 5% Ur (>95% dip in Ur) for 0.5 cycles 40% Ur (60% dip in Ur) for 5 cycles 70% Ur (30% dip in Ur) for 25 cycles <5% Ur (95% dip in Ur) for 5 sec | Not applicable (battery powered) | Mains power quality should be that of a typical commercial or hospital environment. If the user of the EZ-IO® Power Driver requires continued operation during power mains interruptions, it is recommended that the power driver be powered from an uninterruptible power supply or a battery. |
Power frequency (50/60Hz) magnetic field IEC 61000-4-8 | 3 A/m | 3 A/m | Power frequency magnetic fields should be at levels characteristic of typical location in typical commercial or hospital enviroment. |
NOTE: Ur is the a.c. mains voltage prior to application of the test level. |
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Download Teleflex Arrow EZ-IO, 9040, 9058 - Power Driver Manual