Table of Contents READ BEFORE YOU START......................3 COMPANY PROFILE...........................4 INTRODUCTION..........................5 INSTALLATION INSTRUCTIONS.......................7 SCALER OPERATING INSTRUCTIONS..................10 SCALER MAINTENANCE........................12 WATER FILTER CLEANING INSTRUCTIONS.................16 SADDLE VALVE ASSEMBLY ......................17 WATER TANK INSTRUCTIONS.......................20 WATER TANK CARE & MAINTENANCE..................21 CARE AND MAINTENANCE OF YOUR PORTABLE WATER TANK..........21 TECHNICAL DATA..........................22 DIMENSIONS.............................22...
This is normal for all ultrasonic scalers. Never let the scaling tip touch soft tissue or skin, Engler Engineering Corporation is not responsible for any damage caused by improper use of this device and / or its accessories.
A.D.S. 2000, microprocessor controlled anesthesia delivery system / ventilator, Sentinel V.R.M., respiratory monitor. We manufacture all of the inserts and tips used in the Engler products as well as many others on the market today in the 18K and 25K frequency range.
As a result, the tip vibrates at this ultrasonic frequency with an amplitude of 0.001 to 0.004 in. (25.4 um to 102 um). In designing our unique Engler tips, water flows internally through the tip as it vibrates. As the bubbles in the lavage are bactericidal, the energy released collapses and destroys the bacterial cell walls. The water flowing internally through the tip, effectively cools the area and assists in removing any debris from the operative site.
Page 6
Multi turn water control, counterclockwise to Display Power control open, clockwise to close. Handpiece holder Scaler permanent connection (not a quick disconnect, do not twist) Scaling Nose Scaler handpiece cone (Ultrasonic stack inside) on first run, remove stack until water flows then install stack. Resettable fuse Power inlet (plug power cord Water quick disconnect...
INSTALLATION INSTRUCTIONS Before installing or operating your new Sonus II, carefully read and follow all of the instructions. IMPORTANT: This device must be connected to a clean, filtered, water supply, capable of delivering 30 to 60 PSI., 0 to 4.2 kg/cm2) of water input pressure. This unit comes with an In Line Water Filter (part # A52030).
Page 8
Female Quick Disconnect (part #44300) - This is the female mating connector to the Male Quick Disconnect supplied with the Sonus II . Use this to create a custom water installation utilizing ¼ “ I.D. water tubing. 44300 Female quick disconnect 2.
Page 9
DO NOT remove or bypass the ground pin from the power cord. Doing so will void the warranty. IMPORTANT: Your Sonus II has been equipped with a universal switching power supply and will not require adjustments in this regard. See technical data for specifications.
Do not use this device if the water temperature is too hot. Engler Engineering Corporation will not be liable for damage due to improper use of this device.
Page 11
Ultrasonic scaling procedures are not intended for soft tissue. CAUTION: Contact with Soft Tissue Will Cause Burns. Engler Engineering Corporation will not be liable for damage or injury due to improper use of this device.
SCALER MAINTENANCE FINAL PROCEDURES AT THE END OF EACH DAY: 1. Turn the unit off. 2. Remove the tip and nose cone and sterilize. 3. Disconnect the unit from its water source or turn off the water supply. 4. Clean and disinfect all surfaces. Ultrasonic Scaling Tips: IMPORTANT: The scaling tips should be thoroughly cleaned and free of blood, tissue, or any other debris before sterilization.
Page 13
INSERT (TRANSDUCER) INSTALLATION / REMOVAL INSTRUCTIONS You have purchased a precision instrument. Please handle gently - It is easily damaged TO REMOVE THE STACK: 1. Unscrew and remove the scaling tip by turning the plastic nosecone counterclockwise using the tip wrench if necessary.
Page 14
Some tips may no longer be available, please call Engler Engineering for more information...
Page 15
3. Check and / or replace O-ring and Filter Disc in the Inline Filter. 4. Water line may be twisted or kinked. 5. Water restriction in unit: contact Engler Engineering Corporation repair department. 6. If using a Portable Water Bottle, check water level then pump to pressurize the bottle.
WATER FILTER CLEANING INSTRUCTIONS SHOULD BE PERFORMED AT LEAST QUARTERLY 1. Turn off water supply to unit or disconnect the male QD from the female QD. 2. Unscrew filter by firmly holding Point “A” in your left hand and Point “B” in your right hand. (Refer to Photo) Next, unscrew by simultaneously rotating your left hand away from you and rotating your right hand toward you.
SADDLE VALVE ASSEMBLY Installation Instructions To connect the saddle valve to an existing water source: Be sure the tubing is copper and be sure to attach it to the cold water supply. The copper tube should be between 3/8” and 1-3/8”. Tighten the saddle valve's securing screws evenly.
Page 19
It is very important that you mount the saddle valve after a shut off valve. ENGLER ENGINEERING CORPORATION WILL NOT BE HELD LIABLE FOR ANY DAMAGE INCLUDING, BUT NOT LIMITED TO LEAKAGE CAUSED BY IMPROPER INSTALLATION OF OUR PRODUCTS.
WATER TANK INSTRUCTIONS PT-1 Water Tank DIRECTIONS: 1. Release Air Pressure by PULLING and TURNING pressure relief valve, (black knob) located on the side of the bottle. 2. Remove pump and cap assembly. 3. Fill tank with distilled water or medicated solution up to the “FILL LINE” mark (approx. 2/3) . Do NOT fill beyond this line.
4. Black particles found in water bottle indicates that the pump assembly is deteriorating. Order new pump assembly from Engler Engineering. Pump assembly is pre-lubricated when new.