Engler Sonus II Instruction Manual

Ultrasonic dental scaler

Advertisement

Quick Links

Sonus II
Ultrasonic Dental Scaler
Instruction Manual
Engler Engineering Corporation
1099 East 47th Street, Hialeah, Florida 33013
800-445-8581 – 305-688-8581 – FAX 305-685-7671
Web site: www.englerusa.com Support site: www.engler411.com
10/ 09 / 2014 - E

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Sonus II and is the answer not in the manual?

Questions and answers

Summary of Contents for Engler Sonus II

  • Page 1 Sonus II Ultrasonic Dental Scaler Instruction Manual Engler Engineering Corporation 1099 East 47th Street, Hialeah, Florida 33013 800-445-8581 – 305-688-8581 – FAX 305-685-7671 Web site: www.englerusa.com Support site: www.engler411.com 10/ 09 / 2014 - E...
  • Page 2: Table Of Contents

    Table of Contents READ BEFORE YOU START......................3 COMPANY PROFILE...........................4 INTRODUCTION..........................5 INSTALLATION INSTRUCTIONS.......................7 SCALER OPERATING INSTRUCTIONS..................10 SCALER MAINTENANCE........................12 WATER FILTER CLEANING INSTRUCTIONS.................16 SADDLE VALVE ASSEMBLY ......................17 WATER TANK INSTRUCTIONS.......................20 WATER TANK CARE & MAINTENANCE..................21 CARE AND MAINTENANCE OF YOUR PORTABLE WATER TANK..........21 TECHNICAL DATA..........................22 DIMENSIONS.............................22...
  • Page 3: Read Before You Start

    This is normal for all ultrasonic scalers. Never let the scaling tip touch soft tissue or skin, Engler Engineering Corporation is not responsible for any damage caused by improper use of this device and / or its accessories.
  • Page 4: Company Profile

    A.D.S. 2000, microprocessor controlled anesthesia delivery system / ventilator, Sentinel V.R.M., respiratory monitor. We manufacture all of the inserts and tips used in the Engler products as well as many others on the market today in the 18K and 25K frequency range.
  • Page 5: Introduction

    As a result, the tip vibrates at this ultrasonic frequency with an amplitude of 0.001 to 0.004 in. (25.4 um to 102 um). In designing our unique Engler tips, water flows internally through the tip as it vibrates. As the bubbles in the lavage are bactericidal, the energy released collapses and destroys the bacterial cell walls. The water flowing internally through the tip, effectively cools the area and assists in removing any debris from the operative site.
  • Page 6 Multi turn water control, counterclockwise to Display Power control open, clockwise to close. Handpiece holder Scaler permanent connection (not a quick disconnect, do not twist) Scaling Nose Scaler handpiece cone (Ultrasonic stack inside) on first run, remove stack until water flows then install stack. Resettable fuse Power inlet (plug power cord Water quick disconnect...
  • Page 7: Installation Instructions

    INSTALLATION INSTRUCTIONS Before installing or operating your new Sonus II, carefully read and follow all of the instructions. IMPORTANT: This device must be connected to a clean, filtered, water supply, capable of delivering 30 to 60 PSI., 0 to 4.2 kg/cm2) of water input pressure. This unit comes with an In Line Water Filter (part # A52030).
  • Page 8 Female Quick Disconnect (part #44300) - This is the female mating connector to the Male Quick Disconnect supplied with the Sonus II . Use this to create a custom water installation utilizing ¼ “ I.D. water tubing. 44300 Female quick disconnect 2.
  • Page 9 DO NOT remove or bypass the ground pin from the power cord. Doing so will void the warranty. IMPORTANT: Your Sonus II has been equipped with a universal switching power supply and will not require adjustments in this regard. See technical data for specifications.
  • Page 10: Scaler Operating Instructions

    Do not use this device if the water temperature is too hot. Engler Engineering Corporation will not be liable for damage due to improper use of this device.
  • Page 11 Ultrasonic scaling procedures are not intended for soft tissue. CAUTION: Contact with Soft Tissue Will Cause Burns. Engler Engineering Corporation will not be liable for damage or injury due to improper use of this device.
  • Page 12: Scaler Maintenance

    SCALER MAINTENANCE FINAL PROCEDURES AT THE END OF EACH DAY: 1. Turn the unit off. 2. Remove the tip and nose cone and sterilize. 3. Disconnect the unit from its water source or turn off the water supply. 4. Clean and disinfect all surfaces. Ultrasonic Scaling Tips: IMPORTANT: The scaling tips should be thoroughly cleaned and free of blood, tissue, or any other debris before sterilization.
  • Page 13 INSERT (TRANSDUCER) INSTALLATION / REMOVAL INSTRUCTIONS You have purchased a precision instrument. Please handle gently - It is easily damaged TO REMOVE THE STACK: 1. Unscrew and remove the scaling tip by turning the plastic nosecone counterclockwise using the tip wrench if necessary.
  • Page 14 Some tips may no longer be available, please call Engler Engineering for more information...
  • Page 15 3. Check and / or replace O-ring and Filter Disc in the Inline Filter. 4. Water line may be twisted or kinked. 5. Water restriction in unit: contact Engler Engineering Corporation repair department. 6. If using a Portable Water Bottle, check water level then pump to pressurize the bottle.
  • Page 16: Water Filter Cleaning Instructions

    WATER FILTER CLEANING INSTRUCTIONS SHOULD BE PERFORMED AT LEAST QUARTERLY 1. Turn off water supply to unit or disconnect the male QD from the female QD. 2. Unscrew filter by firmly holding Point “A” in your left hand and Point “B” in your right hand. (Refer to Photo) Next, unscrew by simultaneously rotating your left hand away from you and rotating your right hand toward you.
  • Page 17: Saddle Valve Assembly

    SADDLE VALVE ASSEMBLY Installation Instructions To connect the saddle valve to an existing water source: Be sure the tubing is copper and be sure to attach it to the cold water supply. The copper tube should be between 3/8” and 1-3/8”. Tighten the saddle valve's securing screws evenly.
  • Page 19 It is very important that you mount the saddle valve after a shut off valve. ENGLER ENGINEERING CORPORATION WILL NOT BE HELD LIABLE FOR ANY DAMAGE INCLUDING, BUT NOT LIMITED TO LEAKAGE CAUSED BY IMPROPER INSTALLATION OF OUR PRODUCTS.
  • Page 20: Water Tank Instructions

    WATER TANK INSTRUCTIONS PT-1 Water Tank DIRECTIONS: 1. Release Air Pressure by PULLING and TURNING pressure relief valve, (black knob) located on the side of the bottle. 2. Remove pump and cap assembly. 3. Fill tank with distilled water or medicated solution up to the “FILL LINE” mark (approx. 2/3) . Do NOT fill beyond this line.
  • Page 21: Water Tank Care & Maintenance

    4. Black particles found in water bottle indicates that the pump assembly is deteriorating. Order new pump assembly from Engler Engineering. Pump assembly is pre-lubricated when new.
  • Page 22: Technical Data

    TECHNICAL DATA SCALER: Input Voltage: 100 - 250VAC (UNIVERSAL POWER SUPPY) Input Frequency: 50 / 60 Hz. Current (Amperes) 2.3 Amps Transducer Style: Magnetostrictive Operating frequency: 17,500 - 19,000 Hz Ultrasonic Generator Data: Advanced Auto-Tuned Technology. Variable Power Control Water: Input Pressure: 30 PSI (min) 60 PSI (max) DIMENSIONS NET WEIGHT:...