4 Mounting
Dam gauge
Deep well level
14
Fig. 11: Gauge measurement dam, sensor mounting on bracket
Fig. 12: Gauge measurement deep well, sensor mounting on lid
4.5
Measurement setup - Flow
In general, the following must be observed while mounting the sensor:
•
Mounting the sensor on the upstream or inlet side
•
Installation in the centre of the flume and vertical to the liquid
surface
•
Distance to the overfall orifice or Venturi flume
•
Min. distance to the max. height of damming for optimum accu-
racy: 250 mm (9.843 in)
Detailed project planning data can be found at the channel manufac-
turers and in the technical literature.
The following examples serve as an overview for flow measurement.
At smaller distances the measuring accuracy is reduced, see Technical data.
1)
1)
VEGAPULS C 21 • SDI-12
Need help?