MC9R /
Before operating it is critical to reset the bed
To reset the MC9R/Encore
Ensure resident/patient is not in the bed.
With the decks up or down and bed horizontal at any height (i.e., bed not in Trendelenburg or Reverse
Trendelenburg), simultaneously press and hold both HI-LO Up and HI-LO Down buttons from either the
Hand Pendant (buttons
seconds, decks will lower (if raised) and the bed will move to the fully lowered position (if bed is raised).
Continue holding both buttons down until the Hand Pendant LED & backlight, Foot End Controls LEDs flash
(if installed, underbed lights will flash as well). Flashing lights signal a successful bed reset.
Button A
Fig 1
12
Series
®
control box
A &
B) -or- from the Foot End Controls (buttons
Hand Pendant LED
Button B
Fig 2
:
C &
D). After approximately five
Foot End Controls LEDs