Büchi Labortechnik AG
10 Help with faults
10.1 Troubleshooting
10.1.1 Troubleshooting general
Operation Manual Mini Spray Dryer S-300
Problem
Possible cause
Instrument cannot
No electric connection.
be switched on.
Peristaltic pump
The rollers are not in contact
does not apply sol-
with the feed tubing.
vent.
Feed tube defective.
Product is delivered
The pressure of the rollers
after the spray flow
on the running surface is too
is switched on al-
weak.
though the pump is
switched off.
Nozzle is blocked.
Product is too concentrated.
Encrustation on nozzle exit.
Nozzle is defective (e.g. bent
nozzle needle).
Irregular or pulsed
Leaks in the spray nozzle.
spraying.
Product drips in
No spray flow.
spray cylinder.
Insufficient spray flow.
Help with faults | 10
Action
Establish an electrical
u
connection. See Chapter 5.2
"Establishing electrical
connections", page 24.
Lift the hose base.
u
Adjust the peristaltic pump
u
bed. See Chapter 7.8
"Adjusting the peristaltic pump
bed", page 45.
Replace the feed tube.
u
Adjust the peristaltic pump
u
bed. See Chapter 7.8
"Adjusting the peristaltic pump
bed", page 45.
Use a lower concentration in
u
pump.
Clean the nozzle. See
u
Chapter 9.10 "Cleaning the
nozzle", page 64.
Replace nozzle or defective
u
part.
Check seals in the spray
u
nozzle.
If necessary, replace the seals.
u
Open spray gas valve.
u
Check pressure of the spray
u
gas in the supply tube (5–8
bar).
67/90