Advertisement
Before using the voltage tester DUSPOL® digital plus: Please read the operating manual carefully and always observe the safety instructions!
Attention:
After maximum load (i.e. after a measurement of 30 seconds at AC 500 V/ DC 750 V), the voltage tester must not be used for a duration of 240 seconds!
The voltage tester is marked with international electric symbols and symbols for indication and operation with the following meaning:
The DUSPOL® digital plus is a two-pole voltage tester according to IEC 61243-3 with digital display. As supplementary function, the voltage tester is equipped with a measuring point illumination and display illumination as well as with a phase and phase-sequence indication. For all these functions, the voltage tester requires a built-in battery (2x micro LR03/ AAA). Determining the phase of external conductors and the phase-sequence of a three-phase mains is only possible provided the neutral is earthed. The voltage tester is designed for DC and AC voltage tests within the voltage range of 5 V up to AC 500 V/ 1.5 up to DC 750 V. It can be used to perform polarity tests at DC voltage.
The voltage tester consists of the test probes L1 and L2
and a connecting cable
. The test probe L1
is equipped with a LC display
. Both test probes are provided with push buttons
. From a voltage of 6 V onwards, the voltage tester switches on automatically. For voltage tests below 6 V, the voltage tester has to be switched on by briefly pressing the push button
of test probe L2
. The voltage tester only works properly with the batteries (inside test probe L1
) being intact and inserted correctly. The display
indicates voltages within the nominal voltage range of 5 V up to AC 500 V/ 1.5 up to DC 750 V. Exceeding the limiting value for low voltages (ELV, AC 50 V, DC 120 V) is additionally indicated on the display.
By pressing both push buttons, the voltage tester switches to a lower internal resistance (suppression of inductive and capacitive voltages). Furthermore, a vibrating motor (motor with a flyweight) is put under voltage. From approximately 200 V this motor is set in rotation. With the voltage increasing, the motor's speed and vibration increases as well so that additionally by means of the handle of test probe L2 the voltage value can be estimated roughly (e.g. 230/ 400 V). The duration of the test with a lower internal resistance of the device (load test) depends on the value of the voltage to be measured. To prevent excessive warming of the voltage tester, it is equipped with a thermal protection (reverse control). With this reverse control, the speed of the vibrating motor decreases as well.
The measuring point illumination can be activated by means of the push button of test probe L1
with the device being switched on. Depending on the brightness, the LCD background illumination is activated automatically.
Hold Function
If you press and hold the push button of test probe L1
during a voltage test, the last measuring value is indicated and is flashing. The voltage tester can be separated from the unit under test and the measuring value can be read (DATA-HOLD). The value can be deleted by releasing the push button.
Display field
The display field consist of an LC display. This display indicates the exceeding of the upper limiting value for low voltages (ELV)
, the phase, the phase-sequence
and
, the exact voltage value
, the polarity at DC voltage
and
as well as a symbol for weak batteries
. The measuring range for continuous voltage measurement is set automatically. Up to 80 V, the value is indicated with a decimal place. For higher values, the decimal place is not indicated.
Do not use the voltage tester unless all functions are operating correctly!
Please make sure that you touch the voltage tester at the insulated handles of test probes L1 and L2
only! Do not cover the display and do not touch the contact electrodes!
At AC voltage from 5 V onwards, the display indicates the „plus" and „minus" symbol and
. Furthermore, the three-digit indication
indicates the measuring value (up to approx. 80 V with decimal place!).
Note:
The reading of the LC display might be impaired due to unfavorable light conditions.
Never touch the contact electrode of test probe L2 during the single-pole test (phase test)!
If the „"-symbol appears in the upper part of the LC display
, the tester is in contact with the live phase of an AC voltage on this point of the unit under test.
Note:
The reading of the LC display might be impaired due to unfavorable light conditions, protective clothing or in insulated locations.
Please make sure that you touch the voltage tester at the insulated handles of test probes L1 and L2
only! Do not cover the display and do not touch the contact electrodes!
Please make sure that you touch the voltage tester at the insulated handles of test probes L1 and L2
only! Do not cover the display and do not touch the contact electrodes!
When contacting the two contact electrodes with two phases of a three-phase mains connected in clockwise rotation, the LC display
indicates the symbol "
" (phase-sequence clockwise)
. If for two phases the rotation is counter-clockwise, the LC display indicates the symbol "
" (phase-sequence counter-clockwise)
.
The phase-sequence test always requires a counter-test! For this purpose, the measurement has to be performed again with reversed contact electrodes . During the counter-test, the LC display must indicate the opposite phase-sequence. If in both cases, the LC display indicates a clockwise phase-sequence, the earthing is too weak!
Note:
The reading of the LC display might be impaired due to unfavorable light conditions, protective clothing or in insulated locations.
Do not set the voltage tester under voltage with the battery compartment being open!
The energy supply of the DUSPOL® digital plus is done by means of two built-in micro batteries (LR03/ AAA). Battery replacement is necessary as soon as the display shows the battery symbol "" (weak battery)
. The symbol appears in case the battery voltage is below 2.75 V. If the battery voltage drops below a value of approx. 2.5 V, the battery symbol is flashing.
Indicating the battery voltage
Switch the voltage tester on by briefly pressing the push button of test probe L2
. After approx. 10 seconds, the value of the battery voltage is indicated for 1 second. (Example:
)
How to replace the batteries:
Take a screw driver and open the battery compartment (next to the cable outlet) by a 1⁄4-turn in direction of the arrow (counter-clockwise). The slot is now vertical and the battery compartment with the batteries can be removed.
Remove the discharged batteries from the battery compartment. Insert the new batteries with correct polarity (see marking) into the battery compartment. Put the battery compartment with the batteries back onto the handle and lock it by 1⁄4-turn in clockwise direction (slot must be horizontal and the marking points are opposite!). Make sure not to damage the O ring. If necessary, it has to be replaced.
Battery disposal:
Do not dispose of batteries with the household garbage. You as a consumer are legally obliged to return used batteries. You can return used batteries to public collection facilities in your community area or return them to any retail outlet selling similar batteries. Avoid using batteries containing dangerous substances!
Clean the exterior of the housing with a clean dry cloth (exception: special cleansing cloths). Do not use solvents and/or abrasives to clean the voltage tester. Make sure not to contaminate the battery compartment and the battery contacts with leaking battery electrolyte. Should such electrolyte contamination or white deposits occur near the battery or the battery housing, these must also be removed with a dry cloth. In case of wear or damaging of the O ring of the battery compartment, the voltage tester does not comply with the indicated protection class anymore (protection against dust and water). In this case, the O ring must be replaced.
The O ring can be ordered under the BENNING piece number 772871. Moisten a new O ring with Vaseline or talcum so that the battery compartment can be locked and unlocked easily.
voltage | time |
230 V | 30 s |
400 V | 9 s |
750 V | 2 s |
Attention!
The voltage tester cannot be operated with the batteries being empty! Remove the batteries from the device at a longer storage!
BENNING Elektrotechnik & Elektronik GmbH & Co. KG
Münsterstraße 135 - 137 • D - 46397 Bocholt
Telefon ++49 (0) 2871 - 93 - 0 • Fax ++49 (0) 2871 - 93 - 429
www.benning.de • E-Mail: duspol@benning.de
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Download Benning DUSPOL digital plus - Two-pole voltage tester Operating Manual
Advertisement
Need help?
Do you have a question about the DUSPOL digital plus and is the answer not in the manual?
Questions and answers