Process Documentation - Miele PG 8536 AD LFM SST Operating Instructions Manual

Laboratory glassware washer
Table of Contents

Advertisement

Process documentation

The lab washer provides the option to
document the cleaning processes
(process documentation).
External process documentation
software or an external printer can be
used to output the process
documentation.
The rear panel of the lab washer is
equipped with two RJ45 sockets used
for connecting process documentation
software or a printer.
For instructions on how to configure
the interface, refer to the section on
PC/Printer functions in the
"Programming Manual".
Process documentation with external
software
The "Ethernet" socket is intended for the
transmission of process protocols and
temperature curves to an external
process documentation software.
An Ethernet cable (straight-through) is
required for the Ethernet connection of
the lab washer to suitable network
components (e.g. hub, switch).
A twisted Ethernet cable (cross-over) is
required for the direct Ethernet
connection between the lab washer
and an external device, e.g., a PC.
48
Process documentation with external
printer
The "Printer" plug is intended for
connecting an external printer to a
serial interface.
An RJ45 / Sub-D adapter plug is
supplied for the connection of the lab
washer to an external printer via an
RS 232 interface.
For instructions on how to set the
external printer functions, refer to the
section on PC/Printer functions in the
"Programming Manual".
When connecting a printer or PC,
please note the following:
– use only approved printers,
– select a printer or PC that is suited to
the installation application,
– the length of the extension cord from
the serial interface to the printer or
PC must not exceed 32' 10" (10 m),
and 328' (100 m) from the Ethernet
interface.
For more information regarding
suitable printers and software
solutions, please contact Miele.