INSTRUCTIONS FOR USE CONGRATULATIONS ON YOUR PURCHASE OF A PRODUCT FROM NOUVAG. We are pleased that you have chosen a quality product from NOUVAG and thank you very much for the trust you have placed in us. These instructions for use will familiarize you with the device and its functions so that you can apply and use them correctly.
INTENDED PURPOSE MEDICAL INDICATIONS The HighSurg 30 is a control unit which is used in combination with an electronic motor, handpieces, sterile single use tubing set (independent medical devices) to perform surgery in the following medical indications: // Plastic Surgery...
The use of the devices outside of the intended purpose is prohibited. Non intended modifications to the control unit and accessories are not allowed. Use of third party‘s devices and accessories not indicated by NOUVAG is not allowed. Repairs must only be performed by authorized NOUVAG service centres.
SETUP DEVICE AND ACCESSORIES SETUP ¬ Place the HighSurg 30 and all required accessories and instruments on an even, non-slip surface and make sure you have good access to all controls. ¬ The installation of the device near other devices is prohibited due to EMC.
Use only NOUVAG tube sets REF 6024 and REF 6025, otherwise the correct function cannot be guaranteed. Check the expiry date of the tubing set and ensure that the packaging is not damaged. Using non-sterile tubing sets can result in serious infection and, in extreme cases, can cause death.
Page 14
STEP 2 TEST OF HANDPIECE AND CONTRA ANGLE To make sure the handpieces work in perfect condition, the HighSurg 30 can check them in a test run for their torque behavior. This is a simple procedure, which helps to detect disorders and defects in the handpieces at an early stage, to ensure safety during application to the patient.
When switching the HighSurg 30 off, all the settings, made by the user are automatically stored in each program place. At switch off from the HighSurg 30 all the settings made by the user are automatically stored in each pro- gram place.
Page 20
HIGHSURG 30 INSTRUCTIONS FOR USE OPERATION The HighSurg 30 can recognize the type of plugged-in motor. This enables the adaption of the motors and their safe operation. SPEED RANGE FACTORY MOTOR 21 50’000 rpm DEFAULT DEFINITION Motor 21, 50’000 rpm / Min. speed 300 –...
Do not use tube set when packaging is already opened or damaged! Do not use tubing set if expired. Use only NOUVAG tubing sets with REF 6024 and REF 6025. Sterility cannot be guaranteed by reusing and re-sterilization of tubing sets. The characteristics of the mate- rial may change resulting in serious infections or, in worst case, the death of the patient.
Results shall be documented. The service manual, wiring diagrams, and descriptions are available upon request from the manufacturer. NOUVAG offers a safety inspection service for its customers. Addresses can be found in the appendix of this instructions for use under .
6 months, wear parts are excluded from the warranty. During this warranty period, NOUVAG agrees to either repair or replace the product at its option if the product fails to function properly under normal use and service and such failure is due solely to a defect in workmanship or materials.
Electromagnetic compatibility (EMC) Remark: The Product subsequently referred to herein always denotes the HighSurg 30. Changes or modifications to this product not expressly approved by the manufacturer may result in increased emissions or decreased immunity performance of the product and could cause EMC issues with this or other equipment. This product is designed and tested to comply with applicable regulations regarding EMC and shall be installed and put into service according to the EMC information stated as follows.