Restore factory defaults
Enable Remote Monitoring Service
QuantStudio
5 Real-Time PCR Instrument User Guide (for Human Identification) | MAN0017162 A.0
™
Chapter 4 Configure the instrument and manage instrument profiles
IMPORTANT! Back up the instrument before restoring factory defaults (see "Backup
or restore the instrument" on page 48 ).
When you restore an instrument to its factory defaults:
·
The following items are deleted:
·
All instrument profiles
·
All files stored on the instrument, including all EDS files
·
All custom dye calibrations, including calibrations for ABY
·
RNase P verification
·
The following items are not deleted:
·
The most recent valid ROI/uniformity and background calibrations
·
The most recent system dye calibrations
In the home screen:
1. Touch
Settings4Instrument Settings4Restore Factory Defaults.
2. Touch Restore Factory Defaults.
3. Power Off, then power On the instrument to apply the change.
After restoring factory defaults, perform the "Initial start up" tasks.
Note: Enabling the Remote Monitoring Service allows the instrument to
automatically notify Thermo Fisher Scientific support teams in real time of potential
instrument issues. The service monitors and sends general instrument data, but the
service does not monitor or send customer data.
In the home screen:
1. Touch
Settings4Maintenance and Service4Monitoring.
2. Slide the control On to enable the Remote Monitoring Service.
3. Touch OK.
4
Restore factory defaults
™
™
and JUN
dyes
31