Further settings
78

Additional functions

You can use this menu to customise the current programme to suit
technical requirements and the wash load or to reset all additional
functions to the factory default settings.
Additional specialist knowledge is required to alter programme
settings and this should therefore be undertaken only by experienced
users or by Miele Service.
Changing programme parameters on a validated machine will
necessitate a renewed performance validation.
Open the
Further settings
the  button and then switch it on again with the  button whilst
keeping the  button held in.
Open the menu as follows:
Further settings
Additional functions
Reset
All parameters that have been set under "Additional functions" will
be reset to the factory default setting.
Increased water level
The water level will be increased for all programmes.
Interim rinse
All programmes that this option is intended for will have an extra
interim rinse stage (see the programme chart).
Dispensing system
Venting the dispensing system and renaming it.
Temperature / time
Adjust the temperature and holding time for the Main wash 1 or
Final rinse 1 programme block.
Use the  and  arrow buttons to select an option and confirm
your selection with OK.