Use with a 56100A Scalar Network Analyzer

7-3
Use with a 56100A
Scalar Network
Analyzer
NOTES
The MG369XA's GPIB address
should be set to 5 (the default ad-
dress setting) for operation with a
56100A SNA. To verify or change
the GPIB address setting refer to
Configuring the GPIB on page 3-80.
The 56100A SNA will only accept
and display nine video markers, F1
thru F9, from the MG369XA.
When performing amplifier testing
only use the MG369XA power
level, L1.
7-10
The MG369XA is directly compatible with the Anritsu Model 56100A
Scalar Network Analyzer (SNA). The following paragraphs provide in-
structions for connecting the signal generator to the 56100A SNA so
that is can be used as a signal source for the analyzer. Operating in-
structions for the network analyzer can be found in the Model 56100A
Scalar Network Analyzer Operation Manual, P/N 10410-00193.
Figure 7-2. MG369XA to 56100A SNA Connections
Connect the MG369XA signal generator to the
Connecting
56100A SNA as shown in Figure 7-2.
the MG369XA
to the 56100A
Step 1.
Step 2.
Step 3.
Use With Other Instruments
Connect one end of the Auxiliary I/O ca-
ble (P/N 806-7) to the 56100A rear panel
AUX I/O connector. Connect the other end
of the cable to the MG369XA rear panel
AUX I/O connector.
Connect one end of the dedicated system
bus cable (P/N 2100-1) to the 56100A rear
panel DEDICATED GPIB connector. Con-
nect the other end of the cable to the
MG369XA rear panel IEEE-488 GPIB con-
nector.
Turn on the instrument and the 56100A.
The system is now ready to operate.