Message Display With Numeric Error Codes; Table 9-1 Universal Hand Control And Surgical Table Compatibility Mismatch; Table 9-2 Display Screen Numeric Error Codes - Steris 5095 Operator's Manual

General surgical table
Table of Contents

Advertisement

Table 9-1. Universal Hand Control and Surgical Table Compatibility Mismatch
Scenario*
Table is not in Beach Chair Mode, but Control and Table function correctly.
1
When Beach Chair Mode (option) is engaged, Control displays BEACH CHAIR
MODE ERROR (see BEACH CHAIR MODE Section).
When table is powered on, user is prompted to select [applicable]
2
Compatibility Mode.
Control display shows BAD VERSION message but otherwise Control and Table
function correctly. When table is powered on, user is prompted to select
3
[applicable] Compatibility Mode.
Control and Table shut down after approximately seven seconds.
4
NOTE: * For scenarios 1, 2, and 3, The Refer to Instructions symbol (at right) appears in
lower right corner of display.

9.5 Message Display With Numeric Error Codes

Universal Hand Control displays numeric error messages beside Wrench symbol when V2 table control board is
installed in a Surgical Table. Numeric error codes persist during a key press for articulation related errors. Numeric error
codes persist until the Surgical Table is powered off for errors related to system failures.
error code display.
Table 9-2
F F i i g g u u r r e e 9 9 - - 4 4 . . E E x x a a m m p p l l e e o o f f E E r r r r o o r r C C o o d d e e D D i i s s p p l l a a y y - - E E r r r r o o r r C C o o d d e e 2 2 0 0 S S h h o o w w n n
Wrench #
2
ALARM - Commands received
from both Primary and
Auxiliary Hand Controls.
7
ALARM - SLIDE blocked.
9
ALARM - Table batteries
depleted.
46
56404485_C
lists error codes (wrench codes) and their meanings.
Table 9-2. Display Screen Numeric Error Codes
Condition
Only one Hand Control (Primary or Auxiliary) is required to
perform user guided task.
Press STOP. Slide function is blocked causing an over-voltage
situation, address obstruction.
Plug table into facility power.
STERIS 5095 General Surgical Table
Section 9 — Surgical Table Operation
Corrective Action
1.
Power table off by
pressing STOP button.
2.
Power table on by
pressing any key except
STOP.
3.
Select [correct table]
Mode when prompted.
1. Power table on by pressing
any key except STOP.
2. Select [correct table] Mode
when prompted.
Figure 9-4
shows a typical
Action