Replacement Kit Rotor Control HS
2.

Stator inductance test

This test is suitable to detect internal stator coil short-circuits.
Depending on the location of a possible short-circuit (number of shorted wires), the resistance of the coil still
shows a value within tolerance but the inductivity is out of tolerance. This leads to an increased stator current
and can destroy the rotor control unit.
stator link
U - V
= 1 – 2 ≈
101/103 –
107/109
V - W
208/210 –
= 2 – 3 ≈
202/204
- Stator links not removed.
- Stator wires U-V-W disconnected.
This check can be performed with a low-cost inductance
measuring device obtainable locally or can be ordered at the
SLI Hamburg with the 12NC 4512 101 77141.
Please send the measuring results referring to the tube type and serial number together with generator serial
number and customer/site data to:
Dietmar Morgenstern Service Innovation XRT
mailto:dietmar.morgenstern@philips.com
e.g.
XYZ hospital, Town, Country
SRO 25 50 serial number 89123 – 654123
Generator Optimus 50 970023
Inductance:
U--V = 56; V--W = 35
Appreciate your help and many thanks in advance.
Rk RoCo HS Optimus
SRO stator (mH)
wire
-10%
51.7
57
30.6
34
© 2004 Philips Medical Systems
ALL RIGHTS RESERVED
+10%
62.7
37.4
(04.0)