1 x Etekcity Lasergrip 1080 Infrared Thermometer
1 x 9V DC Battery
1 x User Manual
Measurement Range: -58°F ~ 1022°F (-50°C ~ 550°C)
Accuracy: ±2% / ±2°C
Resolution: 0.1°F / °C
Maximum Output: < 1 mW
Wavelength: 630 - 670 nm
Distance-Spot Ratio: 12:1
Response Time: < 500 ms
Emissivity (Adjustable): 0.1-1.0
Battery: DC 9V
Display Icon Key
NOTE: The Lasergrip cannot measure the temperature of objects behind glass. Inaccuracy may also occur when exposed to steam, dust or any other contaminants in the air.
HOLD for continuous temperature reading
RELEASE to lock the temperature result
NOTE: The Lasergrip LCD will display 'OH' when the measured temperature is above the measurable range, and will display 'OL' when the measured temperature is below the measurable range.
To switch between units of temperature measurement, press the °C/°F button at any time while the Lasergrip is on.
To activate and deactivate the laser, press the laser pointer On/Off button at any time while the Lasergrip is on.
To activate and deactivate the display backlight, press and hold measurement trigger and then press the display backlight On/Off button at the same time.
To adjust the emissivity value before measurement, press the emissivity adjustment mode button, then use the UP and DOWN
buttons to navigate to the desired value. You may refer to the emissivity reference chart for correct surface emissivity values. To transfer into Max Temperature Mode, press the mode button again.
NOTE: The Lasergrip cannot measure the temperature of objects behind glass. Inaccuracy may also occur when exposed to steam, dust or any other contaminants in the air.
A new pre-installed 9V battery comes in protective plastic wrap inside your Lasergrip. Follow the steps below to properly install or replace the battery.
NOTE:
The Lasergrip 1080 measures surface temperature on the basis of distance to spot diameter ratio (D:S). As the distance between the thermometer and the surface increases, the total surface area measured will also increase. With a distance to spot ratio of 12:1 the surface area measured has a diameter of roughly 1/12 the distance.
For the most accurate results, make sure the target has a surface area of twice the corresponding spot diameter. Insufficient surface area will result in inaccurate results. The recommended distance to hold the Lasergrip from the surface of measurement is 14.17 in (36 cm). This creates a spot measurement area of 1.18 inches (3 cm) in diameter.
Distance (D) to Spot (S) size D:S = 12:1
The emissivity of a material is its efficiency in emitting thermal energy. Non-reflective surfaces have a higher emissivity (closer to 1) than reflective surfaces (closer to 0). Inaccurate results may occur when measuring reflective surfaces such as glass, polished wood, and granite.
To take accurate temperature measurement of reflective surfaces with low emissivity, adjust the emissivity value before measurement using the reference chart. Or, place a strip of masking tape over the surface and allow for it to adjust to the temperature of the surface for approximately 30 minutes. Measure the surface, scanning the taped section, eliminating the issue of inaccuracy.
The device certification information is labeled directly on the device. The sticker is located on the left side of the Lasergrip.
The device warning and aperture safety information are also labeled directly on the device; the sticker is located on the right side of the Lasergrip.
Any updates to the product information (date of manufacturing and manufacturer address) will be added as adhesive overlays.
Questions or Concerns?
support@etekcity.com
(855)-686-3835
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Download Etekcity Lasergrip 1080 - Infrared Thermometer Manual
Advertisement
Need help?
Do you have a question about the Lasergrip 1080 and is the answer not in the manual?
Questions and answers