AD33, AD34, AD35 and AD36 are waterproof EC and TDS testers, that also measure temperature. The housing has been completely sealed against humidity.
All EC and TDS readings are automatically temperature compensated, and temperature values can be displayed in °C or °F units.
EC (or TDS) calibration is automatic at one point, while temperature range is factory calibrated and can only be adjusted by the user.
The AD33P probe supplied with the meters, is interchangeable and can be easily replaced by the user.
The encapsulated temperature sensor allows fast and accurate temperature measurement and compensation.
To prolong battery life, these models are provided with an auto-off feature that turns them off after 5 minutes of non-use.
Moreover, when batteries become too weak to ensure reliable readings the meter automatically turns off.
Each meter is supplied complete with:
Range
0.0 to 60.0°C / 32.0 to 140.0°F
0 to 1999 µS/cm (AD33)
0 to 1999 ppm (AD34)
0 to 199.9 µS/cm (AD35)
0 to 199.9 ppm (AD36)
Resolution
0.1°C / 0.1°F
1 µS/cm (AD33)
1 ppm (AD34)
0.1 µS/cm (AD35)
0.1 ppm (AD36)
Accuracy (@20°C/68°F)
±0.5°C / ±1°F
±2% f.s. (EC/TDS)
Calibration
EC & TDS: automatic, 1 point
Temperature: adjustment
Probe
AD33P (included)
Battery Type / Life
4 x 1.5V button type
Approx. 150 hours of use
Auto-off
After 5 minutes of non-use
Environment
0 to 50°C (32 to 105°F);
RH 100%
Dimensions/Weight
175.5 x 39 x 23 mm / 100 g
AD33P | Spare EC probe for AD33, AD34, AD35, AD36 |
A70031P | 1413 µS/cm solution, 20 ml, 25 pcs |
A70030P | 12.88mS/cm solution, 20 ml |
A70032P | 1382 ppm (mg/l) solution, 20 ml, 25 pcs |
Note: Before taking any measurement, make sure the meter has been calibrated.
Note: If measurements are taken in different samples successively, rinse the probe thoroughly to eliminate crosscontamination. After cleaning, rinse the probe with some of the sample to be measured.
For better accuracy, frequent calibration of the tester is recommended. In addition, calibration should be performed whenever the probe is replaced, after testing aggressive chemicals and where high accuracy is required.
Calibration Procedure
Note: The calibration solution value must be within the meter measurement range.
These meters are factory calibrated for temperature readings, and the user can only perform an adjustment using an accurate reference thermometer.
Procedure
The probe can be easily replaced by unscrewing the body as shown below.
When the batteries become too weak to ensure reliable readings, the "Eb" message appears on the LCD, then the meter turns off.
Batteries must be replaced.
Unscrew and release the probe body. Take out the battery compartment and carefully replace all four batteries while paying attention to their polarity.
Reattach and tighten the probe body properly to ensure a watertight seal.