  • Page 3 AMD LASERS™, LLC. Operating Instructions Manual Information in this document is subject to change without notice. ©2009 AMD LASERS™, LLC. All rights reserved. www.amdlasers.com - 2 -...
  • Page 4: Table Of Contents

    - 11 - TOOTH-WHITENING - 11 - Delivery System - 12 - PICASSO Handpiece - 12 - Directions for Stripping and Cleaving PICASSO Fiber - 13 - OPERATION INSTRUCTIONS - 14 - Turn ON PICASSO - 14 - Turn OFF PICASSO...
  • Page 5 Daily Maintenance - 18 - Contamination Control Procedures - 18 - Cleaning Instructions for PICASSO Surgical Handpiece and the Fiber Optic Cable - 18 - High-level Disinfection Instructions for the Fiber Optic Cable - 18 - Steam Sterilization for PICASSO Surgical Handpiece...
  • Page 6: Introduction

    PIv3809 INTRODUCTION PICASSO Dental Soft Tissue Laser is a surgical device at the cutting edge of technology, designed for a wide variety of oral soft tissue and tooth whitening procedures. PICASSO utilizes a solid state diode as a laser energy source. The energy is delivered to the operating area by means of a delivery system consisting of a flexible fiber connecting the laser source and the handpiece.
  • Page 7: Safety Features

    One beep will sound to indicate that the system is ready for use. Footswitch PICASSO will not emit laser energy until the user presses down on the footswitch. Emergency Stop Press the red Emergency Stop button to instantly turn off the unit. To reset the Emergency Stop button, simply press RESET.
  • Page 8: Safety Classification

    PIv3809 Safety Classification The following safety classifications are applicable to the device: Laser Radiation – Class 4 • Type of protections against electrical shock – Class 1 • • Degree of protection against electrical shock – Type B Applied Part •...
  • Page 9: Equipment Description

    EMERGENCY STOP BUTTON Disables the unit in the event of an emergency. Push once to disable Picasso. Push again to enable. HANDPIECE AND TIPS Handpiece and optional tips for a variety of operation HANDPIECE HOLDER...
  • Page 10: Picasso Setup

    PIv3809 PICASSO Setup For Details refer to "Quick Start Guide" (Figure 4). See page 7 for identification of components. 1. Place unit in a clean, dry, and well ventilated area. 2. Verify circuit breaker is in OFF position. 3. Connect key into key switch. Verify it is in the OFF position.
  • Page 11: Language Selection

    5 steps: Mute->Quiet->Normal->Loud ->Very Loud Setting PICASSO PICASSO IS PRESET TO THE FOLLOWING PARAMETERS: Custom presets The user is able to select parameters as POWER, INTERVAL, DURATION, CONT or REP modes according to the operation. Then, press the SAVE button to store the selected parameters to PRESETS 1-4.
  • Page 12: Power Settings And Use

    PIv3809 STEP 2: Select desired preset 1-4 (Figure 5, #2) STEP 3: POWER. Select desired power settings 0.1 – 7 watts using the buttons (Figure 5, #3). STEP 4: INTERVAL / DURATION. Select desired interval and duration (figure 5, #4 and #5) (available only on REP mode).
  • Page 13: Delivery System

    AMD LASERS™, LLC. Operating Instructions Manual Delivery System NOTE: The fiber optic cable, handpiece, and tips are shipped non-sterile. The Delivery System consists of: Fiber Optic Assembly • NOTE: The standard fiber optic cable assembly is a 400µm fiber. Other sizes are available upon request.
  • Page 14: Directions For Stripping And Cleaving Picasso Fiber

    PIv3809 Directions for Stripping and Cleaving PICASSO Fiber PICASSO fiber should be cleaved after each procedure. To cleave the fiber: 1. Loosen the proximal end of the Handpiece by unscrewing the Handpiece base (Figure 8). 2. Push fiber optic cable into the base of the handpiece until fiber appears from the tip. Pull fiber approx.
  • Page 15: Operation Instructions

    AMD LASERS™, LLC. Operating Instructions Manual OPERATION INSTRUCTIONS Turn ON PICASSO • Turn circuit breaker to ON position. See Set-UP Page 8 for Circuit Breaker Location Rotate key clockwise to ON position. The unit will turn on • and begin a self-diagnostic routine for 4-5 seconds (Figure 12).
  • Page 16: Picasso Specifications

    PIv3809 PICASSO SPECIFICATIONS General Dimensions L x W x H -> 9.5'' x 6'' x 6.2'' (240 x 150 x 160mm) • • Weight 6.6 lbs (3 kg) Electrical • Operating Voltage AC 110/230V 50/60 Hz ± 10% Current Rating 3A 1.5A...
  • Page 17: Contraindications, Warnings, And Precautions

    Adjacent Structures PICASSO is designed to remove soft tissue. Therefore, always be aware of adjacent structures and substructures during procedures. Be extremely careful not to inadvertently penetrate or ablate underlying or adjacent tissues. Do not direct energy towards hard tissues, such as tooth or bone. Do not direct energy towards amalgam, gold or other metallic surfaces.
  • Page 18 CLINICAL APPLICATIONS Introduction To efficiently remove tissue, it is imperative to understand the nature of the PICASSO. PICASSO operates unlike traditional devices. The techniques mentioned below must be practiced and perfected to ensure efficient operation. Please study this section carefully, practice on model tissues, and attend a quality diode laser-training seminar before using this device in a clinical situation.
  • Page 19 , LLC to discuss extended services contracts and annual maintenance options. Daily Maintenance Use a disinfectant to wipe down the front panel of PICASSO system after each procedure. Do not use bleach or abrasive cleansers. Contamination Control Procedures The contamination control suggested for PICASSO fiber optic cable, handpiece, and interchangeable tips is the system sterilization method.
  • Page 20 , LLC. Storage PICASSO should be stored in a cool, dry place when not in use. Cover the unit when not in use for extended periods of time. Store the system in a place where it will not be accidentally hit.
  • Page 21 AMD LASERS™, LLC. Operating Instructions Manual CALIBRATION PICASSO is factory calibrated and certified for use. Recalibration of the unit should be performed every 12 months and should be performed only by authorized personnel. To calibrate the unit, ™ contact your authorized representative or AMD LASERS , LLC at 1.866.999.2635.
  • Page 22 Handpiece • Fiber Optic Assembly • • Laser Safety Glasses* o *Prescription eyewear is available. Please contact your local authorized ™ representative or AMD LASERS , LLC for additional information. Key(s) • • Remote Interlock Plug Power Cord • Footswitch •...
  • Page 23 ™ LASERS , LLC will repair, or, at its option, replace the defective parts without charge. AMD LASERS LLC reserves the right to make such an examination and to make necessary repair/replacement in its own factory, at any authorized repair station, or at the purchaser/user’s place of installation. In the ™...
  • Page 24 DAMAGES OF ANY KIND INCLUDING, BY NOT LIMITED TO, DAMAGES FOR LOSS OF REVENUE, LOSS OF BUSINESS OR BUSINESS OPPORTUNITY, OR OTHER SIMILAR FINANCIAL LOSS ARISING OUT OF OR IN CONNECTION WITH THE PERFORMANCE, USE, OR INTERRUPTED USE OF THE “PICASSO” SYSTEM(S) ™...
  • Page 25 AMD LASERS™, LLC. Operating Instructions Manual - 24 -...
  • Page 26 PIv3809 25 -...