FILE ACCESS CODE . . . . . . . . . . . 1 to 9999
The File Access Code allows configuration file retrieval from the USB port when
the Program Lock Code is locked. This allows a process recipe to be entered
into Cygnus 2 from a USB storage device without unlocking the Program Lock
Code. If the File Access Code has not been enabled, USB port retrieval is not
allowed when a Program Lock Code is present. This parameter has no effect if
there are no lock codes, or the access code is set to zero. Default value is 0.
When pressing CLEAR on power-up, the File Access Code is cleared when the
parameters are cleared, not when the lock codes are cleared. The File Access
Code is not stored on nor retrieved from the USB port.

6.9 AUDIO / VISUAL Page Set Up

Figure 6-8 General audio/visual page
AUDIO FEEDBACK . . . . . . . . . . . . . Yes (1), No (0)
Default is No. Yes Activates Audio Feedback, signifying keyboard interaction.
LCD DIMMER TIME . . . . . . . . . . . . . 0 to 99 min
Determines the time before the backlight is turned off to conserve display life.
Default is 0 minutes which disables the dimmer function.
NOTE: The backlight is ON when a process is running. It is only OFF when
NOTE: All keys previously active in the dimmed page remain active. Pressing
time elapsed since last key press and in READY or IDLE and
end-of-process exceeds the LCD Dimmer Time setting.
any active key performs that key's function and restores the display.