Displayport phy 2.1 receiver calibration and test user guide & method of implementation using anritsu mp1900a bert and high performance real-time oscilloscope with displayport 2.1 receiver calibration and test automation software
Do you have a question about the GRL-DP21-SINK-AN and is the answer not in the manual?
Questions and answers
Summary of Contents for Granite River Labs GRL-DP21-SINK-AN
Page 1
Granite River Labs DisplayPort PHY 2.1 Receiver Calibration and Test User Guide & Method of Implementation (MOI) Using Anritsu MP1900A BERT and High Performance Real-Time Oscilloscope with GRL-DP21-SINK-AN DisplayPort 2.1 Receiver Calibration and Test Automation Software Published on 09 December 2022...
1 INTRODUCTION 2 RESOURCE REQUIREMENTS ....................10 QUIPMENT EQUIREMENTS ....................12 OFTWARE EQUIREMENTS 3 INSTALLING AND SETTING UP GRL-DP21-SINK-AN SOFTWARE GRL-DP21-SINK-AN S ............13 OWNLOAD AND OFTWARE 3.1.1 On the Anritsu MP1900A BERT ................. 13 3.1.2 On the PC Used for GRL Framework Installation ............ 14 GRL-DP21-SINK-AN S ....
Page 4
6 TEST RESULTS AND REPORTS USING GRL-DP21-SINK-AN 6.1.1 DUT Information ....................... 48 6.1.2 Results Summary Table ................... 48 6.1.3 Calibration & Test Result Details ................50 ......................51 ELETE ESULTS 7 SAVING AND LOADING TEST SESSIONS 8 APPENDIX A: MANUAL CTS SINK CALIBRATION AND TEST PROCEDURE 53 UHBR 10 (10.0 G...
Logo certification and listing on the DP Integrators List. This User Guide & MOI explains how to set up and use the GRL-DP21-SINK-AN software to automate receiver calibration and compliance testing for DisplayPort 2.1 Sink conformance. The...
To open the software application, follow the procedure in the following section. Download and Set Up GRL-DP21-SINK-AN Software Install, launch and set up the GRL-DP21-SINK-AN software on a PC or an oscilloscope (where GRL- DP21-SINK-AN is referred to as ‘Controller PC’ or ‘Scope’ respectively in this MOI & User Guide): 1.
Configure the GRL-DP21-SINK-AN Software Before Calibration and Testing 3.2.1 Enter Calibration/Test Session Information Select from the menu to access the Session Info page. Enter the information as required for the test session that is currently being run. The information provided will be included in the test report generated by the software once tests are completed.
After calibration has completed, testing the DisplayPort Sink DUT for CTS compliance can be performed. The GRL-DP21-SINK-AN application supports automated testing of a DisplayPort Sink device if the device supports standard DPCD through the AUX channel or a compatible AUX controller is connected.
Test Results and Reports Using GRL-DP21-SINK-AN The Report page has all the results from all the test runs displayed. If some of the results are not desired, they can be individually deleted by using the Delete button. Also for a PDF report, select the Generate report button.
Saving and Loading Test Sessions The GRL-DP21-SINK-AN software enables Calibration and Test Results to be created and maintained as a ‘Live Session’ in the application. This allows you to quit the application and return later to continue where you left off.