Bands are no longer in contact (16)
Pressure is too high.
Lower the pressure by rotating the tonometer knob
in the opposite direction.
Correct setting (17)
The inside borders of the fluorescein bands are con-
tacting each other.
© HAAG‑STREIT AG, 3098 Koeniz, Switzerland ‑ HS‑Doc. no. 1500.7006000‑04270 – 2022 – 09
NEDERLANDS
PORTUGUÊS
7 Technical data
7.1 Applanation tonometer AT 900
Measuring force
generated
Installation
Measurement range 0 ― 80 mm Hg
Measurement devi-
ation
Backlash width
Net weight
ESPAÑOL
ITALIANO
FRANÇAIS
through leverage weight.
AT 900 model R
Can be fitted on the peg on the microscope.
AT 900 model T
Can be fitted on the guide plate on the swivelling
peg for the microscope and illumination arm.
AT 900 model BQ
Assembly on the microscope arm.
The measurement deviation in the measuring prism is in the measur-
ing range from 0 – 58.84 mN and amounts to a maximum of ±1.5%
and to a minimum of ±0.49 mN of the nominal value.
≤0.49 mN
AT 900 model R
0.73 kg (without accessories)
AT 900 model T
0.49 kg (without accessories)
AT 900 model BQ
0.78 kg (without accessories)
AT 900 model R
AT 900 model T
DEUTSCH
ENGLISH
AT 900 model BQ
15