Fuses and similar must never be removed or replaced while the appliance is connected to a power source.
The electrical terminal box must never be opened while the appliance is connected to a power source.
The compressor starting equipment must never be dismantled while the appliance is connected to a power source.
Whenever electrical components are dismantled or replaced, the appliance must be moved to an area in which there is no risk of
ignition caused by the electrical components or gases contained in the appliance.
Never use the cabinet if the plug is damaged. The cabinet should be examined by a Gram BioLine service technician in such cases.
When setting up in an ordinary scenario that is not subject to regulations for Zone 2:
The appliance may be connected in accordance with applicable local heavy current regulations.
In both cases:
Use a three-wire plug, if the power outlet is intended for a three-wire plug, a the lead in green / yellow insulation should be
connected to the ground terminal.
Power must be connected via a wall socket. The wall socket should be easily accessible.
All earthing requirements stipulated by the local electricity authorities must be observed. The cabinet plug and wall socket should
then give correct earthing. If in doubt, contact your local supplier or authorized electrician.
Technical Support –
In the event of technical difficulties always
contact Gram BioLine technical support or
a Gram BioLine authorized service partner.
Never dismantle the terminal box or any
other electrical component.
For Ex environments –
Special conditions for safe use may apply
to this product when installing in an
EN 60079-14 environment.
Please, see corresponding Ex certificate for
specifications.