Application Method - Haier HXC-106 Operation Manual

Blood bank refrigerator
Table of Contents

Advertisement

Application method

Temperature display
The refrigerator is set to reach appropriate temperature automatically before leaving the factory:
4℃ ±1℃ (4℃ ±1.5℃ for HXC-1308/1308B) .
Temperature display
NO.
Operation
Button operation
1
Plug in, and power-on
Press the temperature
2
display button
Press the temperature
3
display button
Press the temperature
4
display button
Repeat the operation,
5
starting from 2
Caution
Displayed temperature on the front panel is the temperature of the temperature measurement box
located either in the upper area or the bottom area of the refrigerator. The displayed temperature is
not always 4℃ . It however, shows the bulk temperature inside the temperature measurement boxes.
Light on
Display
Display Mode
Average temperature
of temperature
Average temperature
measurement box
up
low
Temperature of upper
Temperature of
temperature
upper temperature
measurement box
measurement box
up
low
Temperature of lower
Temperature of
temperature
lower temperature
measurement box
measurement box
up
low
Average temperature of
temperature
Average temperature
measurement box
up
low
16
Automatic recovery of alarm
The audible alarm of this series of refrigerator has the following characteristics.
•When the system is in a n alarm mode, press the Buzzer Cancel key on the display board and the
audible alarm is silenced temporarily. Remote alarm condition will not be aff ected.
Light off
•After 30 minutes, if the alarm conditions still persist, the audible alarm will resume. On HXC-
1308/1308B, the resumption time for the audible alarm is 20 minutes.
Remote alarm terminals
The refrigerator of this series has the remote alarm terminals:
•Remote alarm terminals are located on the switch panel of the
electrical control box.On HXC-1308/1308B, they are located at the
upper right corner on the back side of the unit. Remote alarm can
be activated via the actuation of the remote alarm terminal contacts.
The terminals are rated for 30V DC and 2 A.
•Remote terminals include a common terminal, a normally Open
terminal and a normally close terminal. Once the unit is in alarm
mode, the terminals will reverse their status and hence activate the
remote alarm.
•Selecting and installing a remote alarm device is a customer's responsibly. The alarm
Caution
device must be connected with proper alarm terminals.
•When there is a loss of power supply, the remote alarm will be activated even though
the battery switch is not turned on.
•The remote alarm port coordinates with the audible alarm on the refrigerator. Thus, the
remote alarm can be stopped by pressing the key of buzzer (outage alarm cannot be
canceled).
COM
NO
NC
25