Autoclave Vacuum Sterilizer Contents Chapter One – Introduction (1-1) How to use the manual (1-2) Terminology and signs Chapter Two – General Information (2-1) Commitments of the Company (2-2) Guarantee (2-3) Persons Authorized to Use the Device Chapter Three – Introduction to the Device (3-1) General Introduction of the Autoclave Device (3-2) Introduction of Elite Premium Dental Autoclave (3-2-1) Device Specifications...
Page 4
Please impart us with your constructive guidance, so that we can benefit from your comments, recommendations, and gain knowledge about your needs. We hope that you will find using Elite Premium to be an enjoyable and successful experience, Polaris Engineering and Manufacturing...
Introduction (1-1) How to use the manual This manual presents the instructions about use, installation, and maintenance of Autoclave made by Polaris Co. It should be noted that: • The product should be used according to the instructions of this manual. Thus, all parts of it should be read carefully before starting and installing the tool.
Autoclave Vacuum Sterilizer (1-2) Terminology and signs Knowledge of signs meanings is of great importance. A list of signs has been introduced in the following for initial recognition and reference to them if necessary. Danger! Indicates compulsory warnings. Warning! Indicates functional recommendations. Forbidden! Indicates forbidden activities.
Autoclave Vacuum Sterilizer (2-1) Company liabilities Polaris Co. is not liable for any problem involving the follow- ing: Failure to follow the instructions of this manual (incorrect use of the product), repair by an unauthorized person and part replacement without coordination with the after-sales service department, failure due to power fluctuations.
Page 10
General Information • Authorized representatives of Polaris Co. These persons are endorsed by the company to work on the device under any operational circumstances. They are also authorized to do any electrical and mechanical adjustment/ repair, maintenance program and authorized parts replacement.
Device Introduction (3-1) General Introduction of the Autoclave Device What is an Autoclave? Generally speaking, an autoclave is a device with a steam con- tainer used from sterilizing laboratory or hospital materials and equipment. Autoclave is used for medical, dental, medicinal, food industries, microbial environments culture and veterinary medicine purposes etc.
Page 13
Autoclave Vacuum Sterilizer 1- Vertical autoclave (laboratory autoclave) 2- Horizontal autoclave (gravity autoclave, pre-vacuum autoclave, pulsomatic autoclave) or hospital autoclave 3- Desktop autoclave (dentistry autoclave) (3-1-3) Classification of Dental Autoclaves Autoclaves can be classified in the field of dentistry regarding their application range (and method of sterilization) into three general classes: N Type: for sterilizing unwrapped solid loads.
(3-2) Introduction of Elite Premium Dental Autoclave Picture 1 – Elite Premium Dental Autoclave Elite Premium is a B type dental autoclave and a new generation of sterilizers which provides an extensive range of application and assures a high performance authenticity by using the vacuum environment technology and steam pressure.
Page 15
Autoclave Vacuum Sterilizer •Equipped with heat and pressure calibration system •Equipped with a robotic lock system with a uniform appearance •Equipped with an integrated tank made out of stainless steel •Equipped with two water outlet and inlet valves on the front and back of the device •Equipped with a printer installed on the device for printing the sterilization process details...
into the clean water tank •Manual water supply option (3-2-2) Description of the Main Components of the Device LCD and control panel set Device door equipped with Printer robotic lock Use only distilled water Anti-bacterial filter Dirty water output Distilled water input...
Autoclave Vacuum Sterilizer Output current guiding canal from relief valve Fuses and power cables Distilled water input Distilled water output Dirty water output (3-2-3) Description of Device Panel FF: FRESH FULL Elite Premium B18 Select program FE: FRESH EMPTY UNIVERSAL WF: WASTE FULL QUICK B QUICK S...
(3-2-4) Technical Specifications All our products at Polaris Co. are manufactured according to the latest global technology at the highest quality. We use the latest design and production methods to increase the lifespan of our products and to improve our quality.
Autoclave Vacuum Sterilizer (3-2-5) Auxiliary Equipment Part Units Water flask Hose (attached to flask) Tray Manual Fuse Automatic water connection hose Dirty water connection hose Tray handle O-ring Printer paper Hose head Tray holder handle Tray holder Table 2 List of Auxiliary Equipment of the Device...
Page 20
Autoclave Transportation and installation...
The device should remain in a vertical position during movement. For more safety and prevention of any injury due to neglecting the technical details, refer to section 4-1. Failure to follow the aforementioned instructions results in device being damaged, and Polaris Co, won’t be held liable.
(4-3) environmental circumstances It is important to choose an appropriate environment during installation and operation for personal safety, correct operation and long lifetime. Workplace temperature: 0-28oC Maximum humidity: 50% the distance between the device and the wall is suposed to be at least 20cm from each side.
Page 23
Autoclave Vacuum Sterilizer In order to permanently connect the device to the distilled water tank, put the device in a place close to the tank. Connect the distilled water tank with a hose to the clean water input on the back of the device and then push the down button. The pump will work for 3 minutes and in case distilled water is required to remove the FE alarm, push the down button again.
Page 24
Maximum Possible Level for Engineering Parameter Water Loaded into the Device Silicon dioxide 1 mg /l Iron 0.2 mg /l Cadmium 0.005 mg /l Lead 0.05 mg /l Heavy metals 0.1 mg /l Chlorine 2 mg /l Phosphate 0.5 mg /l Transfer coefficient at 20 `C 15µS Evaporation residue...
Page 26
Production application Press Power to turn the device on. The following will be displayed on the monitor. The device is now ready to use and you may open the door and insert the items. Elite Premium B18 Picture 8 – Device Monitor You can enter the program selection menu or settings by pressing Set.
Autoclave Vacuum Sterilizer (5-1) Application Programs Five functional programs according to national standards 4860 and 4863 and EN 13060 are capable of sterilizing all loads with complex or porous shapes (made out of metal, polymers which are capable of sterilization and medical cloths). Please fully read the following instructions before using the device and select the proper program according to load type: Program Title...
Page 28
Production application Prion 20.5 Only singled 5 kg layered or multi layered tools, suspected of contamination with dangerous diseases such as Hepatitis, Bovine Spongiform Encephalopathy and AIDS, porous surgery tools with protein-based contamination is already removed. Bowie & Indicator Dick Vacuum Without tool Test...
Page 29
Autoclave Vacuum Sterilizer Bowie & Dick Program: This program is for assessing steam penetration level. It is strongly recommended to run this program every day before using the device and check the integrity of the device’s perfor- mance by studying the related indicator. If the indicator color change is not proper, you may not use the device.
The Vacuum Test program is for testing the performance of the device when creating the required vacuum level. When Vacuum Error is shown on the monitor, first empty the container from any loads and the run the Vacuum Test program. If there are any problems with the vacuum level the device will show an error and you should stop using the device immediately and contact the after sale services department of the company.
Autoclave Vacuum Sterilizer Elite Premium B18 2000.00.15 22:55:57 000 3 024.8 0.87 b TT:00:02 ST:00:00 Low water level! Cycle Failed Vac.Test STRT/STOP Picture 13 2. If the water inside the tank is low and you choose and select a program (by pressing Enter), Water Level Low will be shown on the monitor and unless at least …..
Select program Time: 04:00:00 B&D VACUUMTEST Date:2000/00/00 TIME SET TEMP CALIB PRESS. CALIB EMEREGENCY OPENING Picture 14 (5-4) Temperature and Pressure Calibration You can press Set and use Temp Calib or Press Calib options from the program menu for temperature and pressure calibra- tion adjustments.
Always use non-flammable, non-corrosion and non-toxic agents. (6-2) Safety Measures of the device To improve safety, Polaris Co. always uses raw materials with guaranteed long-term quality through advanced production procedures. The quality control process and carried out in three...
Page 36
Please remember that items such as the protective door on the device and the arm and also the automatic door lock along with multiple electronic controllers are all parts of the safety modules of the device. In order to assure maximum safety, steam pressure is checked electronically in real time.
Maintenance (7-1) Clean-Up Clean the body of the device with a dry cloth. If needed, you may dampen the cloth with a small dose of water or a non-fat solvent (only for the body). The user should always keep the device clean and free from dirt, water or other unnoticed solvents.
Autoclave Vacuum Sterilizer before the defect is resolved. Do not present the device to unauthorized repairers without the approval of the company. Reminder Dear user, please remember to put the device to rest for 15 to 30 minutes after each working cycle and then start using the device again.
Page 40
Maintenance Message (Error) Status Solution Door is open! Door is open. Close the door. Door lock error! Door is not locked. Open the door and close it again. Low water level! Distilled water tank Run water loading is near empty. operation.
Page 41
Autoclave Vacuum Sterilizer Power Failure Error! Power was lost Run the cycle during the previous again after power is cycle. supplied. END CYCLE Work is complete Open the door for or the device is not 3 seconds and ready for work yet. close it.
Page 42
Code Description Code Description Vacuum valve (pre radiator) Pressure sensor Vacuum valve after radiator Pressure switch Steam generator valve (in steam Air antibacterial current path) filter Air valve (for completing vacuum in Water filter container) Vacuum pump Water filter Water pump (from water tank to Lattice filter steam generator) Water pump (from exterior distilled...
Page 43
Autoclave Vacuum Sterilizer Calling the aftersales services Unit of Polaris Co. +49 (0) 21033373500 +49 (0) 21033373550 www.polarisgmbh.com Service@polarisgmbh.com Email...