Page 1
System 9 Sterile Battery Charger 9110-120-000 Instructions For Use (01)07613327553505 ENGLISH (EN) 2022-08 700000544314 Rev-AD www.stryker.com...
Fuse (IEC 60417 Ref 5016) NOTE - Contact Stryker for a complete list of accessories. Protective earth; protective ground (IEC 60417 Ref To obtain a safe combination, the charger must be used with System 9 5019) batteries and a Stryker-approved power cord.
Charger The System 9 Sterile Battery Charger (Figure 1) has 6 bays to charge System 9 Battery Packs (large or small) through inductive (wireless) charging. The battery packs can be charged before sterilization while (1) unwrapped or after sterilization while (2) wrapped in sterilization wrap. The charger has an illuminated display panel that shows charger, bay, and battery status, including cycle counts.
Confirm with a facility IT representative that the Ethernet receptacle has access to the Internet. See the network requirements in the Specifications section. For more information, contact your Stryker sales representative. No Charge – 1 flashing green bar Error State An amber light(s) on the display panel indicates an error state for the battery or charger.
System 9 Processing Instructions manual supplied with the handpiece wrapped flat and tight across the bottom of the battery To operate the charger in standard mode, no further action is needed.
4. Dry the equipment with a clean, non-linting cloth or medical-grade 2. The number of charge cycles for the selected battery displays in blue compressed air. above the cycle button. 5. Inspect the battery charger and power cord for damage; see Inspection. Figure 8 – Check cycle count www.stryker.com...
Always store and transport the product within the specified environmental limitations. ▪ Always save the original packaging container for reuse. Failure to comply may result in damage during transport to the Stryker customer service center. Disposal/Recycle WARNING - Always follow the current local recommendations and/or regulations governing environmental protection and the risks associated with recycling or disposing of the product at the end of its useful life.
Have qualified personnel confirm that the facility power receptacle is working properly. If the charger remains unresponsive, contact Stryker for service. All 6 bay status indicators flash amber to indicate a charger error. Reset the cord connections on the back of the battery charger and at the facility power receptacle;...
Standard TCP/IP over Ethernet (IEEE 802.3 10/100 Mbs) Required Network Configuration: DHCP server, Mask and Gateway provided by site IT admin; Access to Stryker’s Cloud Technical Specifications of the Connection: Standard TCP/IP over Ethernet (IEEE 802.3 10/100 Mbs), utilized industry standard https (port 443) web services communication to Stryker’s Cloud.
Page 11
EQUIPMENT IS CONNECTED TO AN EQUIVALENT RECEPTACLE MARKED ‘HOSPITAL ONLY’ OR ‘HOSPITAL GRADE’ ” or equivalent wording. Agency Approval: CSA Certified for Canada and US or UL Recognized for Canada and US. Environmental Limitations Operation Storage and Transportation Temperature: Humidity: Atmospheric Pressure: www.stryker.com...
Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. To ensure compliance with FCC and ISED RF exposure requirements this device must be installed to provide a minimum of 20cm between the device and people. www.stryker.com...
Guidance and manufacturer’s declaration - electromagnetic emissions The System 9 Sterile Battery Charger (REF 9110-120-000) is intended for use in the electromagnetic environment specified below. The customer or the user of the System 9 Sterile Battery Charger (REF 9110-120-000) should assure that it is used in such an environment.
Page 14
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the System 9 Sterile Battery Charger (REF 9110-120-000) is used exceeds the applicable RF compliance level above, the System 9 Sterile Battery Charger (REF 9110-120-000) should be observed to verify normal operation.
Page 16
(269) 323-7700 (800) 253-3210 Stryker European Operations Limited Anngrove, IDA Business & Technology Park Carrigtwohill, Co Cork T45 HX08 Ireland Stryker EMEA Supply Chain Services B. V. Frans Maasweg 2 Venlo 5928 SB, The Netherlands UK Responsible Person Stryker UK Ltd.