Measurement accuracy

When using the Measurement and Analysis (M&A) package, it is
important to keep in mind the different aspects that affect the
accuracy of the measurements. These include acoustical
properties, patient echogenicity, measurement tools and
algorithms, scanner setup (especially Field-of-view or Range
settings), probe type used, and operator inputs.
The accuracy of each measurement is highly dependent on
image quality. Image quality is highly dependent on system
design, operator variability, and patient echogenicity. The
operator variability and patient echogenicity are independent of
the ultrasound system.
One of the largest potential sources of error is operator
variability. A skilled operator can reduce this by optimizing the
image quality for each type of measurement. Clear identification
of structures, good probe alignment and correct cursor
placement is important. Because of pixel resolution, the
accuracy of a measurement decreases with decreasing distance
on screen. Therefore it is important when scaling the object on
the screen to avoid measuring objects that are too small.
See also 'Optimizing Measurement Accuracy' on page 8-99 for
recommended techniques.
The accuracy in lateral direction is limited by the beam width and
the beam positioning. The radial accuracy is mainly limited by
the acoustic pulse length.