Appendix 5 : Technical data
POWER SUPPLY
Supply Voltage
Fuse Rating (panel)
Fuse Rating (plug)
SAFETY CONDITIONS
Electric Shock Protection
Pressure Relief Valve
STERILISING AND TEST DATA
Sterilising Time
Drying Time (approximate)
Total cycle time
Operating Pressure
MAXIMUM LOADS
Maximum load per standard tray
Maximum weight per half tray
Maximum total load
WATER AND RESERVOIR
Water Reservoir capacity/ Use per cycle
Temperature of drained water/ flow rate
Heat transmitted by autoclave in one hour
DIMENSIONS AND WEIGHT
Depth
Width
Height
Usable chamber space
Weight (empty)
Weight (full reservoirs and full load)
SHIPPING DIMENSIONS AND WEIGHT (includes unit, accessories and packing)
Depth
Width
Height
Weight
STANDARDS COMPLIANCE
BS EN ISO 15223-1
BS EN ISO 20417
ENVIRONMENTAL REQUIREMENTS (contact Eschmann if required to operate outside these conditions)
Transport and Storage: (Note: Water must be drained)
Ambient temperature range
Relative humidity range
Atmospheric pressure range
Operation:
Location
Ambient temperature range
36
BS EN 13060
BS EN ISO 14971
Supply 230V (±10%) or 220V (-6%, +15%)
See rating plate. 50/60 Hz a.c.
For use with alternating current
2 x F10A, 250V (Part No. MUW218616)
13A
Class 1
Set pressure: 2.85 barg
3 minutes 20 seconds at 134/137°C
The sterilisation time can be increased to a maximum of
30 minutes by an Eschmann trained engineer.
17 minutes
11 mins (Empty, no drying)
303.5 kPa (abs) at 134°C, 331.1 kPa (abs) at 137°C
1.5 kg
0.75 kg
4.5 kg
2500 ml/ 600ml (Proportion re-cycled)
°
81
C/ 1.2 l/min
720 kJ
468 mm
460 mm
360 mm
190 x 110 x 288mm
26.5kg
33.5kg
605mm
655mm
468mm
31.5kg
BS EN 61326
BS EN 62366
-10°C to +40°C
30-70%RH (Non-condensing)
700 millibars to 1060 millibars
Indoor use only
+5°C to +40°C
BS EN 13445
BS EN 61010
490725-05
Need help?
Do you have a question about the Little Sister SES 2010 and is the answer not in the manual?
Questions and answers
Cycle keeps stopping with error code 2 displayed. I turn off then re-start by pressing the on switch and 134Degree btton at the same time. It re-starts but then cuts out again after short time with the sme error code displayed
Error code 2 on the Eschmann Little Sister SES 2010 indicates "Load Not Sterilised." To fix this:
1. Refer to Section 11 for further details.
2. Press OK to unlock the door when indicated.
3. Be cautious, as the load, chamber, and door may be very hot.
4. Wait for the chamber to stabilise before opening the door.
Check the cycle details to determine the cause of failure and restart the cycle if necessary.
This answer is automatically generated
error number 028
touchscreen not responding