Page 2
Warranty A-dec warrants its products against defects in material or workmanship for one year from time of delivery (except for handpieces which have a warranty period of six months). A-dec’s sole obligation under the warranty is to provide parts for the repair, or at its option, to provide the replacement product (excluding labor).
Page 3
Decade 1021 Vac Back Chair Decade 1021 Vac Back Chair...
Decade 1021 Vac Back Chair CONTENTS Operation About Your Decade 1021 Vac Back Chair ..2 Controls ............2 Master On/Off Toggle ........ 2 On/Off Indicator ........3 Instruments and Accessories ......3 Care Instructions ..........4 (Refer also to the Equipment Asepsis Owner's Guide, Publication No.
Decade 1021 Vac Back Chair OPERATION About Your Decade 1021 Vac Back Chair Your Decade 1021 Vac Back chair is available in two configurations: 1. As part of an integrated delivery system that is connected into the regulated air and water supplies of an existing delivery system;...
Decade 1021 Vac Back Chair On/Off Indicator (Free Standing Configuration Only) When the Master On/Off toggle is in the ON position, the On/Off Indicator will be red, indicating that the unit is pressurized. (This is not a lamp; the indicator is operated by air pressure.)
Decade 1021 Vac Back Chair Figure 3. Solids Collector Solids Collector. The solids collector, located on the chair back, collects solids and prevents them from entering the central vacuum system (see Figure 3). For operating instructions, refer to your Assistant’s Instrumentation Owner's Guide ( A-dec Publication No.
Decade 1021 Vac Back Chair Adjustments For system air and water adjustments, refer to your Floor Boxes Owner’s Guide (A-dec Publication No. 85-2611-00). Arm Adjustment If the instrument holder arm on your chair starts to droop when all the instruments are in place, or if the arm swings too freely, an adjustment is necessary.
Decade 1021 Vac Back Chair Syringe Air and Water Pressure Syringe air and water flow are controlled by two pinch valves in the chair back (see Figure 5). Remove the four chair back cover screws, then remove the cover to access the pinch valves (see Figure 4 on page 5).
Decade 1021 Vac Back Chair MAINTENANCE Troubleshooting Given proper care, your Vac Back chair will provide years of reliable service. In the event that something does go wrong with the unit, this troubleshooting guide will tell you what to do before calling to have it serviced.
Page 11
Decade 1021 Vac Back Chair No Air from the Syringe 1. Is the Master On/Off toggle in the ON position? Yes: Proceed with Step 2. No: Move the toggle to the ON position and retest. 2. Is the air compressor ON and operating properly? Yes: Proceed with Step 3.
Page 12
(refer to page 4). Is the screen clogged with debris? Yes: Clean the screen and retest. No: The vacuum supply to the Decade 1021 Vac Back Chair is clogged. Contact your authorized A-dec dealer for assistance. If the preceding steps failed to correct the problem, contact your authorized A-dec dealer for further recommendations.