FL250 FLOORLINE BED
USER MANUAL

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the FL250 and is the answer not in the manual?

Questions and answers

Summary of Contents for aidacare FL250

  • Page 1 FL250 FLOORLINE BED USER MANUAL...
  • Page 2: Table Of Contents

    4. Specifications 5. Accessories 6. Electronic Functions 7. Assembly Thank you for choosing this Aidacare FL250 Bed. This bed has been specifically designed to offer 8. General Operation comfort, support and safety. It is important that you read this user manual 9.
  • Page 3: Symbols Used In This Manual

    INTRODUCTION The symbols below are used throughout this owner’s manual and on the The Aidacare FL250 Bed is an electrically operated Floorline Bed offering product to identify warnings and important information. It is very important functions and positional movements that assist caregivers and maximise for you to read them and understand them completely.
  • Page 4: Safety Precautions

    • The maximum safe working load of the FL250 Bed is 250kg and includes the ALWAYS transport/relocate beds using the EasyMove Transport mass of the patient, mattress and any accessories fitted to the bed = 250kg System when uninhabited.
  • Page 5: Overview

    3. OVERVIEW 4. SPECIFICATIONS The Aidacare FL250 Bed has been designed specifically for the aged care and 0º – 60º hospital environment. Please take the time to familiarise yourself with the key NOTE: Backrest and knee bend components shown below: 0º...
  • Page 6: Accessories

    5. ACCESSORIES A range of FL250 Bed accessories are available. In addition to the below listed Code Description accessories, Aidacare specialises in other complimentary products such as foam mattresses, pressure care surfaces, over bed tables, side lockers and adjustable day chairs. Please contact Customer Service on 1300 133 120 for further information.
  • Page 7: Electronic Functions

    • If the power lead shows any sign of damage, replace immediately • Ensure safety key is removed when handset is not in use 6.1 HANDSET OPERATION FL250 Beds are powered by four electric IP54 rated actuators, which are controlled by the Handset or Attendant Control Panel. Adjustment occurs only Auto-Contour: Conveniently positions patient into cardiac chair while a button is pressed.
  • Page 8: Assembly

    Chair Position: Raises backrest and legrest into cardiac wrapped for further protection. The FL250 Bed should arrive to you in pristine chair orientation and then tilts bed forward into an upright condition, ready for immediate assembly and installation.
  • Page 9 Undo Velcro strap to release the power lead ALWAYS use appropriate manual and powered handling techniques Note: beds with serial number FL250-V2-2204-00000 onwards allow for the and equipment, including the EasyMove Transport System. power cord to be extended if required.
  • Page 10 – which is the bed frame lining up with the shaft locator point 18. Fasten each Holder at the locator points by tightening the hand wheel on the bed frame. This ensures greater stability and rigidity of the Head & Foot Boards aidacare.com.au...
  • Page 11 Remove Bed From Base Carton (B) ASSEMBLY OF FL250 BED (without access to power source) 19. Unlock central locking brake at the foot end by pressing the green pedal 20. Roll the bed out of the flattened carton and position or move the bed as desired Unpack Bed, Mount To EasyMove Transport System: 21.
  • Page 12 4. Line up the Head Board Bracket Pegs with the Head Board Holder keyholes The FL250 Bed Head and Foot Board attachment system is designed to allow 5. Insert, ensuring both sides are fully inserted, push the Head Board down...
  • Page 13 Item 2a – Hardware for Single Bed: • 18 x Attachment Bolts and Washers The FL250 Side Panels are designed to match the FL250 Head and Foot Board options for a complete surround finish. They can also be customised to match Item 2b –...
  • Page 14 The FL250 Self Help Pole and FL250 IV Pole are designed to attach to the FL250 Head Board and therefore will move with any Head Board movement – up, down, tilt etc.
  • Page 15 7.5 KING SINGLE CONVERSION KIT ASSEMBLY Recommended Tools: • Power Drill The FL250 King Single Conversion Kit is designed to retrofit a Single FL250 bed • Right Angled Power Drill Adaptor into a King Single. King Single FL250 beds are available separately (BEB046150).
  • Page 16 3. Remove the 2 bolts from the rotation plate at the base of the folding bar different for Single vs King Single FL250 Beds. Therefore TWO items or SKU’s 4. Attach the folding bar to the mounting plate with these 2 bolts. Fasten tightly must be supplied / sold to correctly fit the Side Panels: 5.
  • Page 17 The FL250 Split Folding Side Rails are supplied as a left/right pair and require this position installation on the bed.
  • Page 18 The FL250 EasyMove Transport System is designed to facilitate safe and NOTE: BEB046590 is compatible with beds with serial number “V1” ONLY easy transport of an FL250 bed over longer distances or through tighter BEB046592 is compatible with beds with serial number “V2” ONLY doors and passage ways.
  • Page 19 11. Ensure the alignment pin to the left of the axle inserts into the bed frame and safety clip to the right of axle is engaged – you will hear a click as it engages over the frame 12. Repeat for second trolley aidacare.com.au...
  • Page 20 27. When the bed frame is vertical, re-engage the pivot locks by pulling and twisting 28. Repeat the above steps in reverse to reposition the bed surface horizontally in preparation for lowering aidacare.com.au...
  • Page 21 4. Assemble USB port to the underside of the backrest using the mattress retainer legs as studs and secure in place with handwheels (bed serial number prior to FL250-V2-2204-24505) or using the M4 x 16mm bolts (bed serial number after FL250-V2-2204-24505)
  • Page 22: General Operation

    (Only applicable to beds with serial number prior is locked in place to FL250-V2-2204-24505) 2. To disengage brakes push down on the green pedal until the brake system 9. Open control box and connect the splitter cable to the port shown...
  • Page 23: Storage & Movement

    Handle With Care FL250 Beds should be stored in a clean, dry, well ventilated area. Do not drop and avoid shock or violent impact when transporting.
  • Page 24: Cleaning

    10. CLEANING Moving your FL250 Bed Before cleaning always disconnect the bed from the mains Before moving the bed, press the ExitSafe position button on the handset power source. or ACP until it is in position WARNING 2. Release the brakes using the green pedal NOTE: It is strongly recommended that protective clothing be used when cleaning.
  • Page 25: Maintenance

    FL250 Bed. Calibration issues can occur when power outages happen or when beds are If inspection reveals that the safety of the bed is jeopardised in any way, it unplugged from the wall during use.
  • Page 26: Warranty & Service

    Date: replaced with non-original Aidacare approved parts. If any faults are detected upon receipt of this product please phone Aidacare. Any faults that are detected during normal use should be reported to Aidacare immediately to determine if warranty conditions apply and if so, the necessary repair or replacement work...
  • Page 27 AIDACARE PTY LTD Building 3A, 1 Moorebank Avenue, Moorebank NSW Australia 2170 1300 133 120 | aspirecare.com.au MADE IN PRC ABUM10.61 281122...