Introduction
This monitor can be used by multiple users. This blood pressure monitor uses the oscillometric method of blood pressure measurement. This means this monitor detects your blood movement through your brachial artery and converts the movements into a digital reading.
Safety Instructions
This instruction manual provides you with important information about the OMRON BP7250 Blood Pressure Monitor. To ensure the safe and proper use of this monitor, READ and UNDERSTAND all of the safety and operating instructions. If you do not understand these instructions or have any questions, contact 1-800-634-4350 before attempting to use this monitor. For specific information about your own blood pressure, consult with your physician.
Intended Use
This device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population. The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.
Environments of Use: Home Patient Population: Adult
Receiving and Inspection
Remove this monitor and other components from the packaging and inspect for damage. If this monitor or any other components is damaged, DO NOT USE and contact 1-800-634-4350.
Symbols Glossary
For symbol information, visit:
OmronHealthcare.com/symbols-glossary
Contraindications
- This monitor is contraindicated for use in ambulatory environments.
- This monitor is contraindicated for use on aircraft.
Read the Important Safety Information in this instruction manual before using this monitor.
Follow this instruction manual thoroughly for your safety.
Keep for future reference. For specific information about your own blood pressure, CONSULT WITH YOUR PHYSICIAN.
![]()
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
- DO NOT use this monitor on infants, toddlers, children or persons who cannot express themselves.
- DO NOT adjust medication based on readings from this blood pressure monitor. Take medication as prescribed by your physician. ONLY a physician is qualified to diagnose and treat high blood pressure.
- DO NOT use this monitor on an injured arm or an arm under medical treatment.
- DO NOT apply the arm cuff on your arm while on an intravenous drip or blood transfusion.
- DO NOT use this monitor in areas containing high frequency (HF) surgical equipment, magnetic resonance imaging (MRI) equipment, computerized tomography (CT) scanners. This may result in incorrect operation of the monitor and/or cause an inaccurate reading.
- DO NOT use this monitor in oxygen rich environments or near flammable gas.
- Consult with your physician before using this monitor if you have common arrhythmias such as atrial or ventricular premature beats or atrial fibrillation; arterial sclerosis; poor perfusion; diabetes; pregnancy; pre-eclampsia or renal disease. NOTE that any of these conditions in addition to patient motion, trembling, or shivering may affect the measurement reading.
- NEVER diagnose or treat yourself based on your readings. ALWAYS consult with your physician.
- To help avoid strangulation, keep the air tube and AC adapter cable away from infants, toddlers or children.
- This product contains small parts that may cause a choking hazard if swallowed by infants, toddlers or children.
Data Transmission
- This product emits radio frequencies (RF) in the 2.4 GHz band. DO NOT use this product in locations where RF is restricted, such as on an aircraft or in hospitals. Turn off the Bluetooth monitor and remove batteries and/or unplug the AC adapter when in RF restricted areas. For further information on potential restrictions refer to documentation on the Bluetooth usage by the FCC.
AC Adapter (optional accessory) Handling and Usage
- DO NOT use the AC adapter if this monitor or the AC adapter cable is damaged. If this monitor or the cable is damaged, turn off the power and unplug the AC adapter immediately.
- Plug the AC adapter into the appropriate voltage outlet. DO NOT use in a multi-outlet plug.
- NEVER plug in or unplug the AC adapter from the electric outlet with wet hands.
- DO NOT disassemble or attempt to repair the AC adapter.
Battery Handling and Usage
- Keep batteries out of the reach of infants, toddlers or children.
![]()
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient, or cause damage to the equipment or other property.
- Stop using this monitor and consult with your physician if you experience skin irritation or discomfort.
- Consult with your physician before using this monitor on an arm where intravascular access or therapy, or an arterio-venous (A-V) shunt, is present because of temporary interference to blood flow which could result in injury.
- Consult with your physician before using this monitor if you have had a mastectomy.
- Consult with your physician before using this monitor if you have severe blood flow problems or blood disorders as cuff inflation can cause bruising.
- DO NOT take measurements more often than necessary because bruising, due to blood flow interference, may occur.
- ONLY inflate the arm cuff when it is applied on your upper arm.
- Remove the arm cuff if it does not start deflating during a measurement.
- DO NOT use this monitor for any purpose other than measuring blood pressure.
- During measurement, make sure that no mobile device or any other electrical device that emits electromagnetic fields is within 12 inches (30 cm) of this monitor. This may result in incorrect operation of the monitor and/or cause an inaccurate reading.
- DO NOT disassemble or attempt to repair this monitor or other components. This may cause an inaccurate reading.
- DO NOT use in a location where there is moisture or a risk of water splashing this monitor. This may damage this monitor.
- DO NOT use this monitor in a moving vehicle such as in a car.
- DO NOT drop or subject this monitor to strong shocks or vibrations.
- DO NOT use this monitor in places with high or low humidity or high or low temperatures. Refer to section 10.
- During measurement, observe the arm to ensure that the monitor is not causing prolonged impairment to blood circulation.
- DO NOT use this monitor in high-use environments such as medical clinics or physician offices.
- DO NOT use this monitor with other medical electrical (ME) equipment simultaneously. This may result in incorrect operation of the monitor and/or cause an inaccurate reading.
- Avoid bathing, drinking alcohol or caffeine, smoking, exercising and eating for at least 30 minutes before taking a measurement.
- Rest for at least 5 minutes before taking a measurement.
- Remove tight-fitting, thick clothing from your arm while taking a measurement.
- Remain still and DO NOT talk while taking a measurement.
- ONLY use the arm cuff on persons whose arm circumference is within the specified range of the cuff.
- Ensure that this monitor has acclimated to room temperature before taking a measurement. Taking a measurement after an extreme temperature change could lead to an inaccurate reading. OMRON recommends waiting for approximately 2 hours for the monitor to warm up or cool down when the monitor is used in an environment within the temperature specified as operating conditions after it is stored either at the maximum or at the minimum storage temperature. For additional information of operating and storage/transport temperature, refer to section 10.
- DO NOT use this monitor after the durable period has ended. Refer to section 10.
- DO NOT crease the arm cuff or the air tube excessively.
- DO NOT fold or kink the air tube while taking a measurement. This may cause an injury by interrupting blood flow.
- To unplug the air plug, pull on the plastic air plug at the base of the tube, not the tube itself.
- ONLY use the AC adapter, arm cuff, batteries and accessories specified for this monitor. Use of unsupported AC adapters, arm cuffs and batteries may damage and/or may be hazardous to this monitor.
- ONLY use the approved arm cuff for this monitor. Use of other arm cuffs may result in incorrect readings.
- Inflating to a higher pressure than necessary may result in bruising of the arm where the cuff is applied.
NOTE: refer to "If your systolic pressure is more than 210 mmHg" in section 3 of this instruction manual for additional information.
Data Transmission
- DO NOT replace batteries or unplug the AC adapter while your readings are being transferred to your smart device. This may result in incorrect operation of this monitor and failure to transfer your blood pressure data.
AC Adapter (optional accessory) Handling and Usage
- Fully insert the AC adapter into the outlet.
- When unplugging the AC adapter from the outlet, be sure to safely pull from the AC adapter. DO NOT pull from the AC adapter cable.
- When handling the AC adapter cable: DO NOT damage it. DO NOT break it. DO NOT tamper with it. DO NOT forcibly bend or pull it. DO NOT twist it. DO NOT use it if it is gathered in a bundle. DO NOT pinch it. DO NOT place it under heavy objects.
- Wipe any dust off of the AC adapter.
- Unplug the AC adapter when not in use.
- Unplug the AC adapter before cleaning this monitor.
Battery Handling and Usage
- DO NOT insert batteries with their polarities incorrectly aligned.
- ONLY use 4 "AA" alkaline or manganese batteries with this monitor. DO NOT use other types of batteries. DO NOT use new and used batteries together. DO NOT use different brands of batteries together.
- Remove batteries if this monitor will not be used for a long period of time.
- If battery fluid should get in your eyes, immediately rinse with plenty of clean water. Consult with your physician immediately.
- If battery fluid should get on your skin, wash your skin immediately with plenty of clean, lukewarm water. If irritation, injury or pain persists, consult with your physician.
- DO NOT use batteries after their expiration date.
- Periodically check batteries to ensure they are in good working condition.
Know Your Monitor
Contents/Product Includes
Monitor, 4 "AA" batteries, Wide-Range D-Ring Cuff, instruction manual, quick start guide
Monitor and Arm Cuff
Front of the monitor
![Omron - 5 Series - Front of the monitor Front of the monitor]()
- Display
- [Forward/Backward] buttons
- [Connection] button
(Date/Time) - [Memory] button
- [START/STOP] button
Left side, back and bottom of the monitor
![Omron - 5 Series - Left side, back and bottom of the monitor Left side, back and bottom of the monitor]()
- Air jack
- AC adapter jack
(for optional AC adapter) - Battery compartment
Arm cuff
(Upper arm circumference: 9" - 17" (22 - 42 cm))
![Omron - 5 Series - Arm cuff Arm cuff]()
- Air plug
- Air tube
Display and Symbols
![Omron - 5 Series - Display and Symbols Display and Symbols]()
![]() | ![]() | Memory symbol Appears when viewing readings stored in the memory. |
![]() | ![]() | Average value symbol Flashes when viewing the average of the latest 2 or 3 readings taken within the most recent 10 minute timeframe. |
![]() | ![]() | Sync symbol Flashes/appears when your data needs to be transferred because the stored memory is either almost, or completely full. Once you pair your monitor with your smart device, transfer your readings immediately before the monitor deletes the oldest reading. Up to 60 readings can be stored in the internal memory of your monitor. |
![]() | Systolic blood pressure reading |
![]() | Diastolic blood pressure reading |
![]() | ![]() | Deflation symbol Appears during cuff deflation. |
![]() | ![]() | Heartbeat symbol Flashes while taking a measurement. |
![]() | Date/time display |
![]() | ![]() | Bluetooth ON symbol Appears when your readings are being transferred. |
![]() | Bluetooth OFF symbol Appears when Bluetooth is disabled. |
| ![]() | High symbol Appears if your systolic blood pressure is 130 mmHg or the diastolic blood pressure is 80 mmHg. Refer to subsection 1.4 for more information. |
![]() | | Irregular heartbeat symbol Appears when irregular rhythm is detected 2 or more times during a measurement. The symbol with the number appears as below: Irregular heartbeat is detected 2 times:![]() Irregular heartbeat is detected 3 times or more:![]() |
![]() or | An irregular heartbeat rhythm is defined as a rhythm that is 25% less or 25% more than the average rhythm detected while your monitor is measuring blood pressure. |
![]() | ![]() Irregular heartbeat symbol may be displayed when the monitor detects arm movements. Refer to section 7 for more information. If it continues to appear, we recommend you to consult with and follow the directions of your physician. |
![]() | ![]() | Movement error symbol Appears with your reading when your body is moving during a measurement. If it appears, remove the arm cuff and wait for 2 to 3 minutes. Then, reapply the cuff, remain still and take another measurement. |
| ![]() | OK symbol Flashes when your monitor is connected to your smart device or readings are transferred successfully. |
![]() | ![]() | Low battery symbol Flashes when batteries are low. |
![]() | Depleted battery symbol Appears when batteries are depleted. |
| Pulse display / Memory number Pulse rate appears after the measurement. When pressing the button, the memory number will also appear for approximately 1.5 seconds before the pulse rate shows again on the display. |
Classification of BP (Blood Pressure)
ACC/AHA 2017 Hypertension guidelines define categories of BP in adults as below.
Categories of BP in Adults*
BP Category | Systolic BP | Diastolic BP |
Normal | < 120 mmHg and < 80 mmHg |
Elevated | 120 - 129 mmHg and< 80 mmHg |
Hypertension |
Stage 1 | 130 - 139 mmHg or 80 - 89 mmHg |
Stage 2 | 140 mmHg or 90 mmHg |
* Individuals with Systolic BP and Diastolic BP in 2 categories should be designated to the higher BP category.
BP indicates blood pressure (based on an average of 2 careful readings obtained on 2 occasions).
Source: ACC/AHA 2017 High Blood Pressure Clinical Practice Guideline
![]()
- NEVER diagnose or treat yourself based on your readings. ALWAYS consult with your physician.
Preparation
Installing Batteries
- Push down the hook of the battery cover and pull downward.
![Omron - 5 Series - Installing Batteries Installing Batteries]()
- Insert 4 "AA" batteries as indicated in the battery compartment.
- Close the battery cover. As soon as batteries are correctly installed, the year flashes on the display. Refer to sub-section 2.2 to set the date and time manually, or refer to sub-section 4.1 to pair your monitor to set the date and time automatically.
![]()
Note
- When the "
" symbol flashes on the display, it is recommended you replace batteries. - To replace batteries, turn your monitor off and remove all batteries. Then replace with 4 new batteries at the same time.
- Replacing batteries will not delete previous readings.
- The supplied batteries may have a shorter life span than new batteries.
- Disposal of used batteries should be carried out in accordance with local regulations.
Setting the Date and Time Note
Note
- Set your monitor to the correct date and time before taking a measurement for the first time.
- As soon as batteries are correctly installed, the year flashes on the display of your monitor.
- The date and time will automatically be set when the monitor is paired with "OMRON connect US/CAN" app.
Blood Pressure Measurement Tips
To help ensure an accurate reading, follow these directions:
- Stress raises blood pressure. Avoid taking measurements during stressful times.
- Measurements should be taken in a quiet place.
- It is important to take measurements at the same times each day. Taking measurements in the morning and in the evening is recommended.
- Remember to have a record of your blood pressure and pulse readings for your physician. A single measurement does not provide an accurate indication of your true blood pressure. You need to take and record several readings over a period of time.
![]()
- Avoid bathing, drinking alcohol or caffeine, smoking, exercising and eating for at least 30 minutes before taking a measurement.
- Rest for at least 5 minutes before taking a measurement.
Applying the Arm Cuff Note
Note
- The following steps are for applying the arm cuff to your left arm. When you take a measurement on your right arm, follow the cuff wrap instructions for a right arm use mentioned under "To take a measurement on your right arm" located at the end of this sub-section.
- The blood pressure can differ between the right arm and the left arm, and the measured blood pressure readings can be different. OMRON recommends to always use the same arm for measurement. If the readings between both arms differ substantially, check with your physician to determine as to which arm to use for your measurements.
![]()
- Remove tight-fitting, thick clothing from your arm while taking a measurement.
- Plug the arm cuff into your monitor by inserting the air plug into the air jack securely until it clicks.
![Omron - 5 Series - Applying the Arm Cuff Note - Step 1 Applying the Arm Cuff Note - Step 1]()
- Place your hand through the cuff loop. Pull the cuff until it reaches your upper left arm.
![]()
Note
- The bottom edge of the arm cuff should be 0.5 inch (1 to 2 cm) above the inside elbow. The air tube is on the inside of your arm and aligned with your middle finger.
![Omron - 5 Series - Applying the Arm Cuff Note - Step 2 Applying the Arm Cuff Note - Step 2]()
- Make sure that the air tube is positioned on the inside of your arm and wrap the cuff securely, so it can not move around your arm.
![]()
To take a measurement on your right arm
When you take a measurement on your right arm, the air tube should run along the side of your elbow, along the bottom of your arm. Be careful not to rest your arm on the air tube.
![]()
Sitting Correctly
To take a measurement, you need to be relaxed and comfortably seated in a room at a comfortable temperature. Place your arm on the table.
![Omron - 5 Series - Sitting Correctly Sitting Correctly]()
- Sit in a comfortable chair with your back and arm supported.
- Keep your feet flat and your legs uncrossed.
- The arm cuff should be placed on your arm at the same level as your heart, with the arm resting comfortably on a table.
Using Your Monitor without a Smart Device
Taking a Measurement
You can use your monitor without pairing to a smart device. To pair your monitor with a smart device.
Note
- To stop the measurement, press the [START/STOP] button once to deflate the arm cuff.
![]()
- DO NOT use this monitor with other medical electrical (ME) equipment simultaneously. This may result in incorrect operation of the monitor and/ or cause an inaccurate reading.
- Remain still and DO NOT talk while taking a measurement.
- Press the [START/STOP] button. All symbols appear on the display before starting the measurement.
- Remain still and do not talk until the entire measurement process is complete.
The "
" symbol flashes at every heartbeat.
After your monitor has detected your blood pressure and pulse rate, the cuff automatically deflates. Your blood pressure and pulse rate are displayed.
If either the systolic or the diastolic reading is high (refer to subsection 1.3), the "
" symbol appears.
![Omron - 5 Series - Taking a Measurement Taking a Measurement]()
- Press the [START/STOP] button to turn your monitor off.
Note
- Your monitor automatically turns off after 2 minutes.
- Wait 2 - 3 minutes between measurements. The wait time allows the arteries to decompress and return to their pre-measurement state. You may need to increase the wait time depending on your individual physiological characteristics.
Using the Guest Mode
The guest mode can be used to take a single measurement for another user.
Note
- No readings are stored in the memory.
- Press and hold the [START/STOP] button for more than 3 seconds.
The date/time display will disappear.
![]()
- Release the [START/STOP] button when the date/time display turns off.
The arm cuff automatically starts to inflate.
If your systolic pressure is more than 210 mmHg
If your systolic pressure is more than 210 mmHg, take a measurement manually.
After the arm cuff starts to inflate, press and hold the [START/STOP] button until the monitor inflates 30 to 40 mmHg higher than your expected systolic pressure.
Note
- Do not inflate above 299 mmHg.
![]()
- Inflating to a higher pressure than necessary may result in bruising where the arm cuff is applied.
Using Your Monitor with a Smart Device
Pairing Your Monitor with a Smart Device
The date and time on your monitor will automatically be set when you pair your monitor with your smart device.
Review the list of compatible smart devices at OmronHealthcare.com
- Enable Bluetooth on your smart device.
- Download and install the free "OMRON connect US/CAN" app onto your smart device.
![www.apple.com]()
![play.google.com]()
If you already have the "OMRON connect US/CAN" app and have created your account, open the app and add your new monitor. - Open the app and follow the pairing instructions shown on your smart device.
- Confirm that your monitor is connected successfully.
When your monitor is connected successfully to your smart device, the "OK" symbol appears.
![]()
- Press [START/STOP] button to turn your monitor off.
Note
![]()
- If the "Err" symbol appears, follow the instructions in the "OMRON connect US/CAN" app.
- Your monitor automatically turns off after 2 minutes.
- Be aware that OMRON will not be responsible for the loss of data and/or information in the app.
- "OMRON connect US/CAN" is the only app that we recommend to use with your monitor to transfer data correctly.
Taking a Measurement
Refer to sub-section 3.1 to start a measurement. Open the app on your smart device and follow the instructions.
Note
- Make sure that your monitor is within 16 ft. (5m) of your smart device.
Transferring Data Automatically/Manually
- The readings will be transferred to your smart device within 1 hour after taking a measurement.
- To transfer the readings manually, press the
button.
Disabling/Enabling Bluetooth Warning
- This product emits radio frequencies (RF) in the 2.4 GHz band. DO NOT use this product in locations where RF is restricted, such as on an aircraft or in hospitals. Turn off the Bluetooth feature in this monitor and remove batteries when in RF restricted areas. For further information on potential restrictions refer to documentation on the Bluetooth usage by the FCC.
Bluetooth is enabled as default.
- When your monitor is off, press and hold the
button for more than 10 seconds. "oFF" flashes 3 times.
![]()
- Press the [START/STOP] button to turn your monitor off.
Note
![]()
- To enable the Bluetooth, press and hold the
button for more than 2 seconds. "on" appears on the display. - Your monitor automatically turns off 2 minutes after the completion.
- When Bluetooth is disabled, the "
" symbol appears.
Using Memory Function
Your monitor automatically stores up to 60 readings.
Note
- This monitor can be used by multiple users. When you use memory function with other users, track (or manage) your own readings appropriately.
Viewing the Readings Stored in Memory
- Press the
button.
![Omron - 5 Series - Viewing the Readings Stored in Memory Viewing the Readings Stored in Memory]()
The Memory number appears for one second before the pulse rate is displayed. The most recent reading stored in the memory will be numbered as "1".
Note
- If the reading is high (refer to sub-section 1.3), the "
" symbol appears. - If there are no readings stored in the memory, the screen to the right is displayed.
- If the memory is full, new readings will replace the old readings, starting with the oldest reading.
- Press the
or
button repeatedly to scroll through the previous readings stored in the memory. :
:To view the older readings
: To view the more recent readings
Viewing the Average Reading
Your monitor can calculate an average reading based on the latest 2 or 3 readings taken within the most recent 10 minute timeframe.
- When your monitor is off, press and hold the
button for more than 2 seconds.
The "
" symbol flashes.
Note
- If there are only 2 readings in the memory for that period, the average will be based on these 2 readings.
- If the average reading is high (refer to sub-section 1.3), the "
" symbol appears.
Other Settings
Deleting All Readings Stored in Memory
- Press the
button.
After a memory number appears, the latest reading will appear. - While holding the
button down, press and hold the [START/STOP] button for more than 3 seconds.
The display to the right flashes.
![Omron - 5 Series - Deleting All Readings Stored in Memory - Step 1 Deleting All Readings Stored in Memory - Step 1]()
- While holding the
button down, press and hold the [START/STOP] button for more than 3 seconds.
![Omron - 5 Series - Deleting All Readings Stored in Memory - Step 2 Deleting All Readings Stored in Memory - Step 2]()
- Press [START/STOP] button to turn your monitor off.
Note
- All readings will be deleted. You cannot partially delete the readings stored in the memory.
- Your monitor automatically turns off after 2 minutes.
Restoring Your Monitor to the Default Settings
To delete all the information stored in your monitor, follow the instructions below. Make sure that your monitor is turned off.
- While holding the
button down, press and hold the [START/ STOP] button for more than 3 seconds.
The display to the right flashes.
![Omron - 5 Series - Restoring to the Default Settings - Step 1 Restoring to the Default Settings - Step 1]()
- While holding the
button down, press and hold the [START/ STOP] button for more than 3 seconds.
![Omron - 5 Series - Restoring to the Default Settings - Step 2 Restoring to the Default Settings - Step 2]()
- Press [START/STOP] button to turn your monitor off.
Note
- Reverting to the default setting of your monitor does not delete the information in the app.
- Your monitor automatically turns off after 2 minutes.
- You will need to re-pair your monitor when using the monitor again. Without re-pairing, your readings will not be transferred to your app.
Error Messages and Troubleshooting
If any of the below problems occur during measurement, check to make sure that no other electrical device is within 12 inches (30 cm). If the problem persists, refer to the table below.
Display/Problem | Possible Cause | Solution |
![]() appears or the arm cuff does not inflate. | Air plug is not completely plugged into the monitor. | Insert the air plug securely. |
The arm cuff is not applied correctly. | Apply the arm cuff correctly, then take another measurement. |
Air is leaking from the arm cuff. | Replace the arm cuff with a new one. |
![]() appears or a measurement cannot be complete after the arm cuff inflates. | You move or talk during a measurement and the arm cuff does not inflate sufficiently. | Remain still and do not talk during a measurement. If "E2" appears repeatedly, inflate the arm cuff manually until the systolic pressure is 30 to 40 mmHg above your previous readings. |
The systolic pressure is above 210 mmHg and a measurement cannot be taken. | Inflate the arm cuff manually until the systolic pressure is 30 to 40 mmHg above your previous readings. |
![]() appears | The arm cuff is inflated exceeding the maximum allowable pressure. | Do not touch the arm cuff and/or bend the air tube while taking a measurement. If inflating the arm cuff manually. |
![]() appears | You move or talk during a measurement. Vibrations disrupt a measurement. | Remain still and do not talk during a measurement. |
![]() appears | The pulse rate is not detected correctly. | Apply the arm cuff correctly, then take another measurement. Remain still and sit correctly during a measurement. If the " " symbol continues to appear, we recommend you to consult with your physician. |
![]() appears |
![]() does not flash during a measurement |
![]() appears | The monitor has malfunctioned. | Press the [START/ STOP] button again. If "Er" still appears, contact customer service. |
![]() appears | The monitor cannot connect to a smart device or transmit data correctly. | Follow the instructions shown in the "OMRON connect US/CAN" app. If the "Err" symbol still appears after checking the app, contact customer service. |
![]() flashes | The monitor is waiting for pairing with the smart device. | Refer to sub-section 4.1 for pairing your monitor with your smart device. |
![]() flashes | The monitor is ready to transfer your readings to the smart device. | Open the "OMRON connect US/CAN" app to transfer your readings. |
![]() flashes | There are more than 40 readings in memory to be transferred. | Pair or transfer your readings to the "OMRON connect US/CAN" app so you can keep them in memory in the app, and this error symbol disappears. |
The date and time is not set. |
![]() appears | There are 60 readings in memory to be transferred. |
![]() flashes | Batteries are low. | Replacing all batteries with 4 new batteries is recommended. |
![]() appears or the monitor is turned off unexpectedly during a measurement. | Batteries are depleted. | Immediately replace all batteries with 4 new batteries. |
Nothing appears on the display of the monitor. | Batteries are depleted. |
Battery polarities are not properly aligned. | Check the battery installation for proper placement. |
Readings appear too high or too low. | Blood pressure varies constantly. Many factors including stress, time of day, and/or how you apply the arm cuff, may affect your blood pressure. |
Any other communication issue occurs. | Follow the instructions shown in the smart device, or visit the "Support" section in the "OMRON connect US/CAN" app for further help. If the problem still persists, contact customer service. |
Any other problems occur. | Press the [START/STOP] button to turn the monitor off, then press it again to take a measurement. If the problem continues, remove all batteries and wait for 30 seconds. Then re-install batteries. If the problem still persists, contact customer service. |
Maintenance
To protect your monitor from damage, follow the directions below:
- Changes or modifications not approved by the manufacturer will void the user warranty.
![]()
- DO NOT disassemble or attempt to repair this monitor or other components. This may cause an inaccurate reading.
Storage
Store your monitor and other components in a clean, safe location.
- Remove the arm cuff from the monitor.
![]()
- To unplug the air plug, pull on the plastic air plug at the base of the tube, not the tube itself.
- Gently fold the air tube into the arm cuff.
![Omron - 5 Series - Storage Storage]()
Note - Do not bend or crease the air tube excessively.
Do not store your monitor and other components:
- If your monitor and other components are wet.
- In locations exposed to extreme temperatures, humidity, direct sunlight, dust or corrosive vapors such as bleach.
- In locations exposed to vibrations or shocks.
Cleaning
- Do not use any abrasive or volatile cleaners.
- Use a soft dry cloth or a soft cloth moistened with mild (neutral) detergent to clean your monitor and arm cuff and then wipe them with a dry cloth.
- Do not wash or immerse your monitor and arm cuff or other components in water.
- Do not use gasoline, thinners or similar solvents to clean your monitor and arm cuff or other components.
Disposal
Dispose of your monitor, other components and optional accessories according to applicable local regulations.
Unlawful disposal may cause environmental pollution.
Optional Accessories
![]()
For further information, visit our website at OmronHealthcare.com
Note
- Do not throw the air plug away. The air plug can be applicable to the optional cuff.
Using the AC Adapter (optional accessory)
Note
- Make sure not to place your monitor in a location where it is difficult to plug and unplug the AC adapter.
- We recommend keeping batteries in your monitor at all times, even if you choose to use the AC adapter. If only the AC adapter is used without keeping batteries in your monitor, you may need to reset the date and time each time you unplug and plug back the AC adapter. The readings will not be deleted.
![Omron - 5 Series - Using the AC Adapter Using the AC Adapter]()
- Insert the AC adapter plug into the AC adapter jack on the rear side of your monitor.
- Plug the AC adapter into an electrical outlet.
To unplug the AC adapter, unplug the AC adapter from the electrical outlet and then remove the AC adapter plug from the monitor.
![]()
- Only use the AC adapter, the arm cuff, batteries and accessories specified for this monitor. Use of unsupported AC adapters, arm cuffs and batteries may damage and/or may be hazardous to this monitor.
Specifications
Model | BP7250 HEM-7150T-Z |
Display | LCD digital display |
Cuff pressure range | 0 to 299 mmHg |
Blood pressure measurement range | SYS: 60 to 260 mmHg DIA: 40 to 215 mmHg |
Pulse measurement range | 40 to 180 beats / min. |
Accuracy | Pressure: ±3 mmHg Pulse: ±5% of display reading |
Inflation | Automatic by electric pump |
Deflation | Automatic pressure release valve |
Measurement method | Oscillometric method |
Transmission method | Bluetooth ® low energy technology |
Wireless communication | Frequency range: 2.4 GHz (2400 - 2483.5 MHz) Modulation: GFSK Effective radiated power: < 20 dBm |
Operation mode | Continuous operation |
IP classification | Monitor: IP 20 Optional AC adapter: IP21 |
Rating | DC6 V 4.0 W |
Power source | 4 "AA" batteries 1.5 V or optional AC adapter (INPUT AC100 - 240 V 50 - 60 Hz 0.12 - 0.065 A) |
Battery life | Approximately 1,000 measurements (using new alkaline batteries) |
Durable period (Service life) | Monitor: 5 years Arm cuff: 5 years Optional AC adapter: 5 years |
Operating conditions | +50°F to +104°F (+10°C to +40°C) / 15 to 90% RH (non-condensing) / 800 to 1060 hPa |
Storage / Transport conditions | -4°F to +140°F (-20°C to +60°C) / 10 to 90% RH (non-condensing) |
Weight | Monitor: Approximately 10.2 oz. (290 g) not including batteries Arm cuff: Approximately 6 oz. (170 g) |
Dimensions | Monitor: Approximately 4.2" (w) × 3.4" (h) × 6.0" (l) (105 mm × 85 mm × 152 mm) Arm cuff: Approximately 5.7" × 23.4" (air tube: 29.5") (145 mm × 594 mm (air tube: 750 mm)) |
Cuff circumference applicable to the monitor | 7" to 17" (17 to 42 cm) (included arm cuff: 9" to 17" (22 to 42 cm)) |
Memory | Stores up to 60 readings |
Contents | Monitor, Wide-Range D-Ring Cuff, 4 "AA" batteries, instruction manual, quick start guide |
Protection against electric shock | Internally powered ME equipment (When using only batteries) Class II ME equipment (Optional AC adapter) |
Applied part | Type BF (arm cuff) |
Note
- These specifications are subject to change without notice.
- In the clinical validation study, K5 was used on 85 subjects for determination of diastolic blood pressure.
- This monitor complies with the requirements of ISO 81060-2:2013.
- This monitor has not been validated for use on pregnant patients.
- IP classification is degrees of protection provided by enclosures in accordance with IEC 60529. This monitor and optional AC adapter are protected against solid foreign objects of 0.5 inches (12.5 mm) diameter and greater such as a finger. The optional AC adapter is protected against vertically falling water drops which may cause issues during a normal operation.
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.