Problem
Bed does not function at all
Flashing Side-rails
Hand set not functioning
Bed movement is faulty
Bed movement does not stop
Bed is producing unusual sounds,
burning odors or movement
deviations observed in motors, bed
parts or limits of hand switch/wired
hand set functions
Pressing a function hand set results
in motion other than intended
Foot Lifter™ does not stop moving
in reaction to resistance
Bed continually beeping when
handset is pressed (not 2
beeps)
Bed does not move on wheels
Nurse Control failures:
For standing position Bed does not
move
Troubleshooting
Possible Cause
Power Cord not connected to
Socket or to Controller
Actuator plugs are not connected
to controller
Handset is not connected
Electrical Interference
Loose connection to Controller
Program Failure
Program failure
Plugs of Actuators are connected in
the wrong socket of the controller.
Continues pressing on Hand Set
Electric or Mechanical Problem
Program failure
Plugs of Actuators are connected in
the wrong socket of the controller
Scale is off / Micro- switch
failure (if no scale)
Low power or no power
Wheels are not down
The program was not unlocked
30
Rev.: 15.0
Rev. Date: Nov. 10, 2020
Solution
Clear Error - See page 24
Check connection of both sides of
Power cable.
Check if Socket has electric power
Check that all Actuator plugs are
connected well in the controller
socket.
Check power supply is connected
and has a green light
Check handset connection to the
controller.
Check cable integrity
If bed is tilted or backrest raised
and CPR needed - use Manuel
Mode - See page 24
Preform Clear error – see page 24
Clear Error - See page 24
Check connection of Handset wire
to controller.
Check that power cord is connect-
ed and see if the green light on the
power supply is on.
Call Technician
Initialize Bed (Technician Only)
Call Technician
Do Not operate bed.
Let go of switches
Call Technician - don't move bed
Check all connections
Initialize Bed (Technician Only)
Call Technician and do not operate
bed
Call Technician - don't move bed
Check that power cord is
connected
Press wheel function till hearing
two beeps.
Check unlocking - see Nurse
instructions