Set the 【Left Scale】 for arterial pressure.
Set the 【Right Scale】 for venous pressure.
Set the 【Grid】 of the overlapped waveform area.
Set the 【Wave Speed】 of the overlapped display waveform.

13.6. IBP troubleshooting

This section describes problems you may encounter during using. You can refer to
the following table for troubleshooting. If the problem persists, please contact
maintenance staff.
NOTE: See D Alarm Information for physiological alarms and technical
alarm information.
Problem
IBP parameter area and waveform
area cannot be found on the
interface
P1/P2 does not display systolic and
diastolic pressure measurements
 
S12 Vet Monitor User's Manual 
Solution
1. Check
parameters is set in the 【Screen Layout】
menu or not. For details, see 3.6.1 Setting
Parameters to be protected.
2. Check whether the IBP parameter switch is
turned on or not. For details, see 3.6.1 Setting
Parameters
3. Check the IBP cable, IBP sensor and
module are connected or not.
4. Check the valve position of the 3-way
stopcock is correct or not.
5. Confirm that the sensor has been zeroed.
For details, see 13.3.2 IBP Sensor zero.
Set the displayed pressure to 【All】 . For details,
see 13.5.3 Setting up display types about
Extended Pressure.
13-9 