4.29 PC-100/PC-200 Procedural Stretcher Caster

Tools required:

Removal

1. Pull the base shroud free of its Velcro®
shroud (A) to the upper frame (B) (see figure 4-47 on page 4-91).
Figure 4-47. PC-100/PC-200 Procedural Stretcher Shroud and Upper Frame
1. Torx® is a registered trademark of Textron, Inc.
2. Loctite® is a registered trademark of Loctite Corporation.
3. Velcro® is a registered trademark of Velcro Industries, BV (a Dutch corporation).
TranStar® PC-100/PC-200/PC-300/PC-350/PC-400/PC-450/
PC-500/PC-550/TR-500/SG-500/OB-450/OB-550/EL-550
Series Stretcher Service Manual (man115re)
Chapter 4: Removal, Replacement, and Adjustment Procedures
Ratchet
1
T25 Torx®
head bit
Phillips head screwdriver
Torque wrench 0-250 in-lb
Blue Loctite®
8" x 2" x 4" (203 mm x 51 mm x 102 mm) lumber
If patient security straps are available, use them to secure the base
shroud to the upper frame.
If the patient security straps are not available, you can use rope, or
you can remove the base shroud (refer to procedure 4.25).
4.29 PC-100/PC-200 Procedural Stretcher Caster
7/16" socket
Hammer
5/16" pin punch
2
(P/N SA3618) adhesive
3
fastenings, and secure the base
4
Page 4 - 91