Model 14C Survey Meter Table of Contents Introduction Getting Started Unpacking and Repacking Battery Installation Connecting a Detector to the Instrument Battery Test Instrument Test Reading the Meter-Face Dial Operational Check Specifications Identification of Controls and Functions Safety Considerations Environmental Conditions for Normal Use...
Page 4
Model 14C Survey Meter Parts List Model 14C Survey Meter Main Board, Drawing 464 × 454 HV Power Supply Board, Drawing 464 × 302 Wiring Diagram, Drawing 464 × 175 Drawings and Diagrams...
Model 14C Survey Meter 1. Introduction he Model 14C is a rugged, portable survey instrument that operates on two standard "D" cell alkaline batteries. The instrument features a regulated high voltage power supply set at 900 volts and provides five linear ranges for measurement from 0 to 2000 mR/hr.
Check individual item serial numbers and ensure calibration certificates match. The Model 14C serial number is located on the front panel below the battery compartment. Most Ludlum Measurements, Inc. detectors have a label on the base or body of the detector for model and serial number identification.
) is supplied to the detector via the detector input connector. A mild electric shock may occur if you make contact with the center pin of the input connector. Switch the Model 14C range selector switch to the position before connecting or disconnecting the cable or detector.
Page 8
Model 14C Survey Meter The count rate scale reads 0-5K (kcpm or 1000s of counts per minute) and has COUNTS MINUTE on the dial. BAT TEST If the needle is pointing as indicated below and the instrument range selection switch is on the ×0.1 scale multiple, then the reading is 3.5 kcpm (multiplied by) ×0.1 = 350 cpm.
Page 9
1.0 mR/hr mark on the middle arc lines up with 3.3 kcpm on the upper arc. The meter face in this example works with a detector that receives 3.3 kcpm per mR/hr (the Ludlum Model 44-9 pancake detector.) Additional detectors may be used with this meter face, but only the cpm dial is valid for these detectors.
Model 14C Survey Meter In the following picture, the needle is on the first tick mark past the 4 kcpm mark. Therefore, if the instrument selector switch is on the ×0.1 range, the reading is 4.2 kcpm (multiplied by) ×0.1 = 420 cpm.
Model 14C Survey Meter 3. Specifications : GM, Scintillation Compatible Detectors Meter Dial : typically 0-2 mR/hr and cpm, bat test (others available) : rotary range multiplier switch selects multiples of ×0.1, ×1, ×10, ×100, and ×1000 Ranges : reading within 10% of true value with detector connected Linearity : series "C"...
Model 14C Survey Meter 4. Identification of Controls and Functions : a six-position switch marked “ ”, ×1000, ×100, ×10, ×1, and Range Multiplier Selector Switch ×0.1. Moving the range selector switch to one of the range multiplier positions (×1000, ×100, ×10, ×1, and ×...
YMBOLS Caution! The operator or responsible body is cautioned that the protection provided by the equipment may be impaired if the equipment is used in a manner not specified by Ludlum Measurements, Inc. Caution! Verify instrument voltage input rating before connecting to a power converter. If the wrong power converter is used, the instrument and/or power converter could be damaged.
AINTENANCE RECAUTIONS The Model 14C may be cleaned externally with a damp cloth, using only water as the wetting agent. Do not immerse the instrument in any liquid. Observe the following precautions when cleaning or performing maintenance on the instrument: 1.
6. Calibration and Maintenance ALIBRATION Note: Measure high voltage with a Ludlum Model 500 Pulser or a high-impedance voltmeter with a high-meg probe. If one of these instruments is unavailable, use a voltmeter with a minimum input resistance of 1000 megohms.
Instrument maintenance consists of keeping the instrument clean and periodically checking the batteries and the calibration. The Model 14C instrument may be cleaned with a damp cloth (using only water as the wetting agent). Do not immerse instrument in any liquid. Observe the following precautions when cleaning: 1.
Page 17
Model 14C Survey Meter Note: Never store the instrument over 30 days without removing the batteries. Although this instrument will operate at very high ambient temperatures, battery seal failure may occur at temperatures as low as 38 °C (100 °F).
Note that the first troubleshooting tip is for determining whether the problem is with the electronics or with the detector. A Ludlum Model 500 Pulser is invaluable at this point because of its ability to simultaneously check high voltage, input sensitivity or threshold, and the electronics for proper counting.
Page 19
Model 14C Survey Meter POSSIBLE SOLUTION nonlinear readings (continued) 3. Check for “sticky” meter movement. Does the reading change when you change the meter? Does the meter needle “stick” at any spot? 4. Check the “meter zero.” Turn the power OFF.
2. Check the HV. For most GM tubes, the voltage is normally 900 Vdc, or 460-550 Vdc for “peanut” tubes (Ludlum Model 133 series). 3. If the input sensitivity is too low, the user could see some double pulsing.
Model 14C Survey Meter 8. Technical Theory of Operation NPUT The external detector pulses are coupled from the detector through C4 to amplifier U9/U15/Q2. CR1 protects the amplifier from input shorts. R40 couples the detector to the high-voltage supply. The internal detector (V1) is located on the HV Power Supply circuit board. V1 pulses are coupled through C6 to comparator U13 on the main circuit board.
Page 22
Model 14C Survey Meter ETER RIVE The meter is driven by the collector of Q1, coupled as a constant current source in conjunction with pin 1 of U10. For battery test, U18A opens and U18B closes, and the meter movement is directly coupled to the battery through R31.
To this end, Ludlum Measurements, Inc. strives to supply the consumer of its goods with information regarding reuse and recycling of the many different types of materials used in its products.
Model 14C Survey Meter Parts List Model 14C Survey Meter Reference Description Part Number UNIT Completely Assembled Model 14C Survey Meter 48-1611 Main Board, Drawing 464 × 454 BOARD Completely Assembled Main Circuit Board 5464-454 APACITORS 47pF, 50V 04-5740 100pF, 100V...
Need help?
Do you have a question about the 14C and is the answer not in the manual?
Questions and answers