Alphanumeric Keyboard Replacement Procedure
NOTE:
The Alphanumeric Keyboard is supplied as an integral part of the OPIO Bottom Cover.
In the event that a Vivid S60/Vivid S70 ultrasound scanner that previously did not have the
Alphanumeric Keyboard option fitted and now requires replacement of the OPIO bottom cover, it is
necessary to order and fit the Alphanumeric Keyboard complete with OPIO bottom cover (supplied as
one FRU part).
Figure 8-57 Alphanumeric Keyboard/OPIO Bottom Cover Assembly
CAUTION
8-3-10-1
Tools
Phillips screwdriver.
FRU Part #
Table 9-12
Refer to
8-3-10-2
Time Required
30 min
8-3-10-3
Preparation
Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
8-3-10-4
Alphanumeric Keyboard Removal Procedure
1.) Remove the Operator Panel Keyboard Assembly:.
•
Operator Panel Keyboard Assembly Removal Procedure on page 8 - 44
2.) Carefully place the Operator Panel Keyboard Assembly upside-down on a flat, clean, stable surface.
1
When performing these procedures, take precautions to avoid damage of
electrostatic-sensitive components. Always have the ESD wrist strap
connected either to the DIB chassis or to the GND plug at the rear of the
scanner, and to your hand.
If a battery is present, first remove the battery as it contains stored energy.
Refer to
Battery Removal Procedure on page 8 -
.
on page 9-11
Chapter 8 - Replacement Procedures
V
S60N/V
S70N
B
IVID
IVID
145.
Power Shut Down on page 4 -
S
M
ASIC
ERVICE
ANUAL
7.
8-51