Hart Scientific 9100 HDRC User Manual

Handheld dry-well

Advertisement

Model 9100 HDRC
Handheld Dry-well
User Manual
Rev. 8A1601

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 9100 HDRC and is the answer not in the manual?

Questions and answers

Summary of Contents for Hart Scientific 9100 HDRC

  • Page 1 Model 9100 HDRC Handheld Dry-well User Manual Rev. 8A1601...
  • Page 2 © Copyright,1997 Hart Scientific, Inc. 799 E. Utah Valley Drive American Fork, Utah 84003-9775 Telephone (801) 763-1600 • Fax (801) 763-1010 Internet: http://www.hartscientific.com...
  • Page 3 WARNING To ensure the safety of operating personnel, and to avoid damage to this equipment: DO NOT operate this unit without a properly grounded, properly polarized power cord. DO NOT connect this unit to a non-grounded, non-polarized outlet. DO USE a ground fault interrupt device. WARNING To ensure the safety of personnel, and to avoid damage to equipment: DO NOT use this unit for any application other than calibration work.
  • Page 5: Table Of Contents

    Table of Contents Introduction..... . 1 Specifications and Environmental Conditions . . . 3 2.1 Specifications ..... . . 3 2.2 Environmental Conditions .
  • Page 6 10 Maintenance ....25 11 Troubleshooting ....27 11.1 Troubleshooting.
  • Page 7: Introduction

    1 Introduction Introduction Hart Scientific’s Model 9100 HDRC Dry-well Calibrator is a small, portable in- strument designed for quick on-site checks and calibration of thermocouple and RTD temperature probes. It features an aluminum temperature block well. Three different blocks are available. Block “A” has six holes, four that accept probe diameters of 1/4", 5/32", 1/8", and 1/16", and two that accept probes of...
  • Page 9: Specifications And Environmental Conditions

    2 Specifications and Environmental Conditions Specifications and Environmental Conditions Specifications Operating Range 33°C to 300°C (92°F to 572°F) Accuracy 33 to 100°C: ±0.5°C 100 to 300°C: ±1°C holes greater than 0.25": ±2°C Stability ±0.2°C Heating Times 35 to 300°C: 9.5 min. Stabilization 3 min.
  • Page 10: Warranty

    Warranty Hart Scientific, Inc. (Hart) warrants this product to be free from defects in ma- terial and workmanship under normal use and service for a period as stated in our current product catalog from the date of shipment.
  • Page 11: Safety Guidelines

    3 Safety Guidelines Safety Guidelines • Operate the instrument in room temperatures between 5–40°C (41–104°F). Allow sufficient air circulation by leaving at least 6 inches of space between the instrument and nearby objects. Overhead clearance needs to allow for safe and easy insertion and removal of probes for cali- bration.
  • Page 13: Quick Start

    4 Quick Start Quick Start Unpacking Unpack the dry-well carefully and inspect it for any damage that may have oc- curred during shipment. If there is shipping damage, notify the carrier immedi- ately. Verify that the following components are present: •...
  • Page 15: Parts And Controls

    5 Parts and Controls Parts and Controls The user should become familiar with the dry-well calibrator and its parts: (See Figures 2, 3, and 4). Rear Panel Power Cord - The removable power cord, (Figure 2) attaches to the back side of the 9100.
  • Page 16: Prop View

    The prop was not designed to be used as a handle for carrying the instrument. To avoid damage, it should not be forced be- yond the inclined position of the calibrator. Manual Rev. 8A1601 Hart Scientific...
  • Page 17: Front Panel

    When the desired set-point temperature is reached, releasing button immediately changes the display back to the actual block temperature and the controller begins to heat or cool to the new set-point temperature. 9100 HDRC Figure 4. Front Panel 9100 Manual Rev. 8A1601...
  • Page 19: General Operation

    6 General Operation General Operation Set-up Place the calibrator on a flat surface with at least 6 inches of free space in front of and behind the instrument. The prop may be swung down to raise the front of the instrument from a horizontal position. Plug the power cord into an AC grounded outlet of the appropriate voltage.
  • Page 21: Controller Operation

    7 Controller Operation Controller Operation The controller is fully programmable. Each of the parameters such as propor- tional band, integration time, display units, and calibration offset can be viewed and altered. The controller has been configured at the factory for optimum ac- curacy, stability, and response time.
  • Page 22 7 Controller Operation Figure 5. Controller Programming Chart Manual Rev. 8A1601 Hart Scientific...
  • Page 23 7 Controller Operation Display Alternates Parameter Value dEr.5 dEr.5 Derivative sensitivity Derivative sensitivity di5.5 di5.5 Display sensitivity Display sensitivity no.AL no.AL Disable SP2 Alarm Disable SP2 Alarm annunciator annunciator Move ProG ProG Auto Auto StAY StAY Program mode Program mode auto-exit switch auto-exit switch up and...
  • Page 24 3 seconds or wait approximately 1 minute and the controller will auto- matically return. This is how all of the adjustable parameters for the controller work on the 9100 HDRC Calibrator. The controller may be locked again after changing parameters to avoid accidentally changing one of them. Follow the instructions above and replace the nOnE at the LoCK menu with ALL and then exit out of the program menu as explained.
  • Page 25: Test Probe Calibration

    8 Test Probe Calibration Test Probe Calibration For optimum accuracy and stability, allow the calibrator to warm up for 10 min- utes after power-up and then allow adequate stabilization time after reaching the set-point temperature. After completing calibration, allow the block to cool before switching the power off.
  • Page 26: Calibration Of Multiple Probes

    Inserting a cold probe into a well will require another period of stabilization de- pending on the magnitude of the disturbance and the required accuracy. For example, inserting a ¼ inch diameter probe at room temperature into a sleeve Manual Rev. 8A1601 Hart Scientific...
  • Page 27 8 Test Probe Calibration at 300°C will take approximately 5 minutes to be within 0.1°C of its set-point and will take 10 minutes to achieve maximum stability. Decreasing the time required for the calibration process can be accomplished by knowing how soon to make the measurement. It is recommended that typi- cal measurements be made at the desired temperatures with the desired test probes to establish these times.
  • Page 29: Calibration Procedure

    9 Calibration Procedure Calibration Procedure Pre-Calibration Make sure all parameters in the controller are set according to the parameter table provided in the operation manual. a) Turn the unit on. b) Set all parameters to the values shown in the chart. Calibration Accuracy Specification: ±...
  • Page 30: Accuracy Check

    Record the actual temperature of the dry-well. Repeat for set-points of 100°C, 150°C, 200°C, 250°C and 300°C. If any of the measured temperatures are not within specification, repeat the Span Adjust and the Zero Adjust section of the procedure as needed. Manual Rev. 8A1601 Hart Scientific...
  • Page 31: 10 Maintenance

    Hart Scientific Customer Service for more information. • Before using any cleaning or decontamination method except those rec- ommended by Hart, users should check with Hart Scientific Customer Ser- vice to be sure that the proposed method will not damage the equipment. •...
  • Page 33: 11 Troubleshooting

    Check that the power cord is plugged in. • If the unit has power and the display is off, the fuse may be blown and the unit needs to be returned to Hart Scientific for repair. Contact Hart Scien- tific Customer Service. The display flashes “Input Fail”...
  • Page 34: Ce Comments

    • With the unit stable, slowly rotate the unit. If no change occurs, the unit may need to be calibrated. Contact Hart Scientific Customer Service. If the display changes more than twice the normal display deviation, another unit in the area could be emitting therefore energy. Move the unit to a differ- ent location and rotate the unit again.
  • Page 35: Wiring Diagram

    11 Troubleshooting 11.3 Wiring Diagram Figure 7 Wiring Diagram 9100 Manual Rev. 8A1601...

Table of Contents