Instructions for Use
RIK-1 Med - ENG - Rev03
RIK-117 RIK-122

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the lina and is the answer not in the manual?

Questions and answers

Summary of Contents for W&H Med lina

  • Page 1 Instructions for Use RIK-1 Med - ENG - Rev03 RIK-117 RIK-122...
  • Page 3: Table Of Contents

    Ordinary maintenance Administrator Monthly or 50-cycle maintenance User management (optional) 400-cycle maintenance Traceability options (optional) 800-cycle or biannual maintenance Managing printers 800-cycle maintenance LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 4 Recommendations for validation Diagrams Water quality Accessories, spare parts, consumables Authorized W&H service partners Documentation forms W&H installation check-list Helix test documentation form LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 5: Conformity

    IEC 61326-1 Electrical equipment for measurement, control and laboratory use - EMC requirements; general requirements. IEC 61770 Electric appliances connected to the water mains - Avoidance of backsiphonage and failure of hose-sets. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 6: Symbols And Messages

    PROPERTY DAMAGE MESSAGES Notice: Indicates information considered important, but not hazard-related. Typically to avoid damage to the product. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 7: Introduction

    For any suggestions or remarks please send an email to Please carefully read the safety information (see "Safety warnings" office.sterilization@wh.com. on page  9). LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 8: Use Restrictions

    See "Sterilization cycles" on page  89 for the full list of key program features, including sterilization time, temperature and recommended load type. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 9: Safety Information

    (see "Sterilization cycle management" on page  46). Do not drink any water that has been inside the sterilizer. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 10: Responsibility

    Note: a list of security information is mentioned in the MDS2 document, which is available on request. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 11: Getting Started

    LINA 17: 41.5 kg (91.5 lbs) User interface menu LINA 22: 43.5 kg (95.9 lbs) Sterilizer setup User authentication (optional) USB pen drive Standby mode LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 12 CONTENTS OF THE PACKAGING Emergency door opening tool Reversible rack Tray holder Mains cable Sterilizer Trays (three) LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 13: Handling

    Service W&H. USB pen drive loaded with This manual, declaration of Instructions for Use conformity, warranty card, work test report, maintenance sheet LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 14: Product Description

    Touch screen Service door Sterilization chamber Dust filter Reset button of the thermostat switch Emergency door opening tool Door seal Door pin LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 15 Bacteriological filter Mains switch Identification label Used water drain port (grey) Clean water drain port (blue) Drain tube release button USB port LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 16 Test connection Pressure safety valve cover Used water drain (optional) Water supply inlet (optional) Condenser grid Mains cable guide Mains cable plug socket LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 17 In the normal position, it can host 5 trays horizontally or 3 cassettes/containers vertically. In a 90° degrees rotated position, it can host 3 trays or 3 cassettes/containers horizontally. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 18: Installing The Sterilizer

    Connect the sterilizer to a dedicated line. Do not use cable extensions nor multiple sockets/adapters. Ensure that external and internal surfaces are free from moisture or condensation before connecting to power. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 19 Note: for details see also the "Wi-Fi key & APP - Quick start". LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 20: Operating The Sterilizer

    Fast Cycle activation code, see "Sterilization cycle management" on page  46. Available tests Additional buttons used to navigate the menu. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 21: User Interface Menu

    Opens the pages to manage cycles. During a cycle, shows the cycle parameters in real time. Measurement Sets the unit of measure. Units LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 22 Management Settings Only with a network connection set. Sets the parameters Administrator only. Adds a user. Add User of the network location. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 23 Dust Filter Shows the status of the dust filter for replacement and resets its counter to zero. Formats the USB pen drive. Format LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 24 Shows the number of cycles performed and left before the 4000 cycle necessary maintenance. service Updates the software. Software Update LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 25 Opens a screen with other daily repetition. settings/options. Indicates that the option is ON and allows to set it OFF by touching it. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 26 Copy the system info to the USB pen drive. Shows an animation about the replacement procedure. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 27: Sterilizer Setup

    Wi-Fi dongle key, follow the following procedure under supervision of your IT manager / network administrator. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 28: User Authentication (Optional)

    Configuration and in the DNS Configuration fields. Tap on each entry field and enter the IP addresses (ask your IT manager for details). Then tap to confirm. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 29: Usb Pen Drive

    Notice: Periodically remove the USB pen drive to save the cycle or open or close the chamber door. data on a computer or on another safe support. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 30 CHANGING STANDBY MODE DELAY TIME On the homepage tap > > > to change the delay. to confirm and go back to the previous page. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 31: Administrator

    WHO CAN MANAGE USERS AND RESET THEIR PIN Only a user with administrator rights can create and delete users and reset the PIN code of a user to "0000". LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 32: Traceability Options (Optional)

    If you want the user to check the load and release it as valid at the end of the cycle, tap to confirm and go back to the previous page. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 33: Managing Printers

    If necessary, tap to print another test label and repeat step to confirm the settings and go back to the previous page. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 34: Label Printer Usage (Optional)

    (see "Label content description" on the next On the homepage tap > > page). Activate Disabled. to confirm and go back to the previous page. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 35: Label Content Description

    Date and time of cycle start Expiry Expiry date of the bag/package. The cycle outcome if a storage time is not set. date LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 36: Sterilizer Tests

    Bowie and Dick test CAUTION! Follow local/national guidelines on the frequency of testing. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 37 Remove the indicator sheet from the center of the test pack and check the change in colour. See "Interpret the test result" on the next page LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 38 Certain ares of the indicator sheet have not changed colour since there was an air Repeat the test. If it fails pocket during the cycle due to sterilizer malfunction. repeatedly, call technical service. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 39: Helix Test

    Unscrew the tube capsule and place inside it an indicator strip according to the instructions of the test manufacturer. Screw the capsule. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 40 Part of the chemical indicator strip has not turned dark. The air removal Repeat the test. If it fails repeatedly, from the capsule was not complete. call technical service. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 41: Vacuum Test

    It consists of a vacuum phase, followed by a stabilization period of 5 minutes and a testing period of 10 minutes. The internal pressure is monitored during the testing period. The pressure rise must be less than 0.013 bar (0.19 psi). LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 42 Clean the chamber face side and the chamber filter. Repeat the Vacuum test. See "Carry out the test" above. If the test fails repeatedly, call technical service. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 43: Sterilization Cycles

    Risk of burns. Before touching, ensure the sterilization sterilizer. The manufacturer’s warranty is void in case of damage chamber is cold. caused by chemicals. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 44 Ensure that packs do not touch the walls of the chamber. Place sterilization pouched items with the paper side facing up. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 45: Prepare The Sterilizer

    Follow these recommendations to obtain the most from the drying: Ensure the paper side of the sterilization bags faces up, and that the space in-between bags is enough. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 46: Sterilization Cycle Description

    WARNING! For your safety and the safety of your patients never exceed the maximum load weight limits as this could impair the sterilization process. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 47 You can set the cycle start date and time and enable or disable it for each cycle. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 48 Following is an example: While the sterilization cycle is running or when cycle ends tap : the cycle information page opens. to scroll the pages. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 49 DRY phase. A cycle can be manually aborted at any time. To stop a cycle: Tap STOP: a confirmation request appears. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 50: Unloading

    USB pen drive. The summarized reports are stored in HTML format and the analytical reports in SCL format. All parameters are recorded every second. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 51 USB capacity. To save all the cycle reports: On the homepage tap > > : after the confirmation all sterilization cycle reports are stored in the USB. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 52 To check if the parameters entered are valid, see "Test the data storage (optional)" below. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 53 Plateau/Sterilization phase start MIN, MAX temperature MIN, MAX pressure Plateau/Sterilization phase end Drying phase start Drying phase end Pressure leveling phase Cycle end conditions LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 54 Description Cycle time Cycle time Date (below) Cycle end date and time "Cycle completed" Cycle outcome Trk. Tracking code for traceability management LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 55: Maintenance

    215 lx (±15 lx) to 1500 lx (±15 lx). biannual maintenance" on page  64. Before making any operation, ward off unauthorized personnel from the working area. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 56 When the maximum number of cycles has been met, a message to replace the consumable will be displayed. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 57: Monthly Or 50-Cycle Maintenance

    Rinse with clean water. Note: when the seal is new it might be necessary to hold the door gently closed at the sterilization start. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 58 WARNING! Ensure that the steam diffuser plate is correctly placed and engaged, as this is essential for the sterilization process. Note: The trays, the tray holder and the steam diffuser plate may also be cleaned in a washer disinfector. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 59 Empty the sterilizer chamber by removing the trays and the rack. Turn the filter cap at the back of the chamber (bottom/center) counter-clockwise. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 60 Rinse the cartridge filter with tap water. Insert the cartridge filter in the filter cap. Insert the filter cap with the cartridge filter in its original position. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 61 Proceed as follows: Clean all external sterilizer covers with a slightly damp cloth moistened with water. Notice: Never use disinfectants, detergents or abrasive products. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 62: 400-Cycle Maintenance

    Open the service door. Unscrew the bacteriological filter by hand (counter-clockwise). Screw on the new bacteriological filter (clockwise) and tighten it snug. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 63 Open the chamber door. Pull out the dust filter handle from underneath the sterilizer. Lift the used filter from the handle and remove it. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 64: 800-Cycle Or Biannual Maintenance

    "Prepare the sterilizer for cleaning the water tanks" on the next page. "Access to the water tanks" on the next page. "Clean the water tanks" on page  67 LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 65 To drain the detergent solution during the subsequent cleaning, leave the drain tube attached to the drain port of the tank you want to clean. ACCESS TO THE WATER TANKS Proceed as follows: Lift the water tank cover. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 66 Clean and dry the internal tank cover and its rubber membrane to eliminate any condensate. Notice: Never use disinfectants, strong detergents or abrasive products. Use a slightly damp cloth moistened with water. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 67 Detach the drain tube, attach it to the drain port of the other tank and repeat steps 1, 2, 3. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 68 Maintenance Remove the internal filters. Clean the metal cartridges of the internal filters with tap water. Reposition the internal filters. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 69 Maintenance Reposition the internal tank cover and then the water tank cover. Detach the drain tube. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 70: 800-Cycle Maintenance

    Carefully clean the seal seat and the chamber face side with a damp, lint-free cloth. Moisten the new seal with water to facilitate its placement. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 71 Wipe any residual water and run a Vacuum and Helix test to check for perfect tightness of the seal. See "Vacuum test" on page  41 and "Helix test" on page  39. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 72: 4000 Cycle Or Five-Year Maintenance

    For each element, the actions to carried out are the following: Element Replace Clean Check Solenoid valves Vacuum pump internal parts Sterilization chamber and external surfaces Chamber filter LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 73: Extraordinary Maintenance

    To drain the clean water, insert the drain tube connector in the blue port. When the water has been completely drained, press the release button to remove the drain tube and close the service door. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 74: Disposal

    The same instructions apply to disposal of all used consumable parts. MATERIALS The sterilizer is mainly built from fiber-reinforced polymers, metals and electric / electronic components. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 75: Diagnostics

    Check the dust filter and ensure that the sterilizer fan is not blocked. See error "13x to 16x" on the next page. Repeat the cycle. If the problem persists, call service. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 76 If the problem persists, call service. Wait for the chamber to cool down. Reset the safety thermostat (see "Extraordinary maintenance" on page  73). LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 77 Please close the door. The door must be locked, but you didn't close it. Close the door so it can be locked. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 78 File lost. (see "Sterilization cycle report" on page  50). LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 79 Call technical service for the 4000 cycle maintenance. cycles has been achieved and the relevant Please call your technical support maintenance step is to be performed. to organize service. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 80: Troubleshooting

    If an anomaly is detected, the cycle is aborted automatically, and the sterilizer starts a reset phase. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 81 Push the reset button of the thermostat switch: a click sound indicates that the thermostat switch has been reset. Insert the dust filter back into its original position. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 82 Wait for the sterilizer to finish the error reset phase and follow the instructions on the display. Note: If the thermostat operates repeatedly, call technical service. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 83 Mount the chamber filter cap properly (see "Ordinary maintenance" on page  55) Load incorrectly placed. See "Load maintenance and preparation" on page  43 LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 84 Remove any objects interfering with the chamber door. Check the door does not force message appears. itself. against the load or the chamber furniture. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 85 Call service to order the requested component (door gasket, dust filter, bacteriological maintenance. maintenance of the sterilizer. filter…). See "Ordinary maintenance" on page  55 LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 86: Emergency Door Opening

    The door locking system is electrically activated. In case the door remains locked due to a black-out or an electric fault, an auxiliary unlocking procedure is available. For this reason, two 9V batteries size PP3 or 1604 are required. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 87 Take the emergency door opening tool included in the sterilizer box. Firmly open the service door. Unscrew the bacteriological filter by hand (counter-clockwise). LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 88 Holding the service door pulled, plug the plastic connector into the socket behind the bacteriological filter. As soon as the door opens, unplug the plastic connector to prevent system overload and consequent damage. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 89: Technical Data

    Processing bagged items with the S Naked 134 cycle will result in wet bags/pouches at the end of the cycle, exposing the items to contamination due to improper storage. The display reminds the maximum permitted load before starting a cycle. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 90 2.03 bar (g) 1.04 bar (g) 2.03 bar (g) 5' 30" 20' 30" 20' 30'' 5' 30" Duration of the Plateau/Sterilization phase LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 91 B Universal 134 B Prion 134 B Universal 121 S Naked 134 Note: Values and cycle names could be different depending on country requirements. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 92 Note 4: Duration reported in the table refers to 0.5 kg. It is possible to sterilize up to 2.0 kg by drying time extension. WARNING! Never exceed the maximum load weight limits as specified in the table as this could impair the sterilization process. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 93: Sterilization Cycle Phases

    Rise to the Plateau/Sterilization phase Process (Plateau/Sterilization phase) Vacuum drying Leveling. Pressure inside the sterilization chamber is leveled to the atmospheric pressure. End of the cycle LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 94 All type B sterilization cycles feature the same basic pressure profile as shown in the graph below. The duration and the temperature sterilization phase differ between the various cycles. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 95 The S Naked 134 cycle is specifically designed to sterilize unwrapped instruments, for immediate use on patients, not requiring a complete drying. Thus, the drying phase of this cycle is short and not assisted by vacuum, making this cycle the fastest available. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 96: Technical Data

    Max. RH 80% up to 31 °C (88 °F), linearly decreasing to 50% at 40 15.4’’) °C (104 °F) Bacteriological filter 0.3 µm Note*: Usable space with standard rack and trays. With optional racks and LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 97 Average water consumption 0.5 to 0.66 litres/cycle (0.13 to 0.17 gal/cycle) Tank volume Clean water 4.8 l (1.27 gal), 2.8 l (0.74 gal) with air gap Used water 4.8 l (1.27 gal) LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 98: Recommendations For Validation

    LINA sterilizers can be validated in accordance to EN17665-1. For further details please refer to the Qualification/Validation guide for sterilization cycles of the manufacturer. Part Description Hottest points Coldest points LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 99: Diagrams

    Note (*): The water filtration system must be fitted with a backflow preventing device complying to IEC 61770 and to national and local regulations. For water requirements see "Technical data" on page  96. Note (**): Rear serial connection. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 100: Water Quality

    < 0.1 mg/l cadmium, lead) Chloride < 2 mg/l Phosphate < 0.5 mg/l Conductivity (at 20°C) < 15 µs/cm pH value 5–7 LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 101: Accessories, Spare Parts, Consumables

    190 x 50 x 400 Usable space - Cassette size (mm): 190 x 50 x 312 190 x 50 x 312 190 x 50 x 312 LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 102 190 x 32 x 380 205 x 35 x 380 210 x 32 x 380 210 x 32 x 380 190 x 32 x 380 LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 103 Usable space - Cassette size 190 x 70 x 380 (mm): 190 x 50 x 312 190 x 50 x 312 190 x 50 x 312 LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 104 (186 x 19.5 x 287 mm) Standard aluminium tray for LINA 22 F523205X (186 x 19.5 x 379 mm) Tray holder F523001X LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 105 1 roll of 2100 labels Emergency door opening tool F372106X 1 wax/resin ribbon activation code instructions Kit helix test (PCD plus 30 strips) T801003X LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 106 Note*: The rack, rotated 90°, accepts 5 standard aluminium trays. 2 dust filters 250 helix strips T800205X Before stocks are exhausted LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 107: Authorized W&H Service Partners

    Note: use only accessories recommended by W&H. Authorized W&H service partners A list and a map with your nearest W&H service partner are available at www.wh.com. LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 108: Documentation Forms

    Yes No abort a cycle? Have you removed all the protection covers from the sterilizer Yes No and from all the ship-with? LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 109 28 Have you executed a B Universal 134 cycle program with the tray Yes No rack and trays inserted? 29 Are all connections to the sterilizer well positioned and plugged Yes No (accessories, etc…)? LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 110: Helix Test Documentation Form

    Use this page to create a logbook tracing the effectiveness of the sterilization cycle during the whole lifespan of your sterilizer. FORM Date Cycle n. Operator Released Signature Chemical indicator LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 111 Documentation forms Date Cycle n. Operator Released Signature Chemical indicator LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 112 Documentation forms Date Cycle n. Operator Released Signature Chemical indicator LINA  | Instructions for Use |  RIK-1 Med ENG Rev03 |  11/05/2020 |  © 2020 W&H Sterilization Srl...
  • Page 114 Manufacturer RIK-1 Med Instructions for Use W&H Sterilization Srl via Bolgara, 2 Rev03 Brusaporto (BG) 11/05/2020 Italy Subject to changes www.wh.com +39 035 66 63 000...