System Operation
Optional External X-Ray Receiver Battery Replacement
4-60
STEP
The battery compartment cover will not close
properly if the battery was replaced facing the wrong
way. If cover doesn't close, repeat step 3 to place the
battery correctly.
Close battery compartment cover and finger tighten
4
cover screw.
Description
Follow this procedure to replace the standard 9V transistor radio
battery only if your system is equipped with the optional
externally-mounted X-Ray Receiver unit.
Prerequisites
Verify the external X-Ray Receiver is turned off.
Procedure RTR0030
Replace the 9V battery as follows:
STEP
Using the thumb, press and slide the battery cover
1
off the opposite end of the X-Ray Receiver from the
Velcro strip.
Remove the discharged battery from the battery
2
connector.
SAIC Proprietary
RTR-4 Operator's Manual
ACTION
NOTE
ACTION
RTR0030a
120300 Rev. D
Need help?
Do you have a question about the RTR-4 and is the answer not in the manual?
Questions and answers