Table of contents Table of contents 1 General information ........................1.1 Read the instructions for use ................... 1.2 Read the instructions for use of compatible products ............ 1.3 Scope ..........................1.4 General signs and symbols ..................... 1.5 Description of warning messages ................... 2 Normal use ...
Page 4
Table of contents 9 Accessories and spare parts ....................21 9.1 Recommended accessories .................... 21 9.2 Spare parts ........................22 10 Appendix ..........................23 10.1 Appendix: Scope ......................23 11 Subsidiaries ..........................25 Instructions for use • CLICKLINE Instruments 10 mm • TBF107_EN_V1.0_10-2021_IFU_CE-MDR...
General information 1 General information 1.1 Read the instructions for use If the instructions for use are not followed, patients, users, and third parties may be injured or the product may be damaged. Read the instructions for use carefully and follow all the safety notes and warnings. Read the reprocessing instructions carefully and follow all the safety notes and warnings.
General information – Unnumbered list, 1st level – Unnumbered list, 2nd level 1.5 Description of warning messages To prevent any injury to persons or damage to property, the warnings and safety notes in the instructions for use must be observed. The warnings use the following levels of danger: WARNING WARNING Designates a possible imminent risk.
Normal use 2 Normal use 2.1 Intended purpose CLICKLINE forceps (complete instrument) Forceps are used for grasping, preparing, dissecting and, if necessary, coagulating tissue in minimally invasive procedures. Forceps are surgically invasive and meant for short-term use. CLICKLINE scissors (complete instrument) Scissors are used for cutting and, if necessary, coagulating tissue in minimally invasive procedures.
Normal use 2.2 Indications The medical devices are tool-like instruments that are suitable for use during minimally invasive examinations and treatments. 2.3 Contraindications The medical devices must not be used for procedures in direct contact with the central nervous system (CNS) and central cardiovascular system. Beyond that, there are no contraindications for the use of the medical devices directly associated with the product.
Safety 3 Safety 3.1 Serious incidents A ‘serious incident’ includes incidents which, directly or indirectly, had, could have had or could have any of the following consequences: – Death of a patient, user, or another person – Temporary or permanent serious deterioration in the medical condition of a patient, user, or another person –...
Safety 3.6 Peak voltage output In combination with HF devices, this product must only be exposed to a specific, recurring maximum peak voltage output. If the maximum permissible peak voltage output is exceeded, the product may be damaged. Damaged products can result in injury to patients, users, and third parties.
(such as pacemakers). Observe the information on electromagnetic compatibility. In case of uncertainties, seek expert advice from KARL STORZ. 3.13 Disposal of pointed and sharp products If pointed and sharp products are not disposed of in suitable waste containers, there is a risk of injury and infection.
Product description 4 Product description 4.1 Product overview CLICKLINE handle*, insulated, with HF connection CLICKLINE handle*, uninsulated, without surgical ratchet, without HF connection CLICKLINE outer sheath, insulated, with irrigation port CLICKLINE outer sheath, uninsulated CLICKLINE working insert HF connection Locking button for releasing the working insert with outer sheath Rotating wheel for turning the jaws LUER-Lock connector for irrigation with sealing cap 10 Example: Fully assembled 3-part CLICKLINE instrument...
Product description Interface Interface description Outer sheath – working in- The working insert and outer sheath can be connected to each sert other via the bayonet lock. Please note that only working inserts and outer sheaths of the same length and size may be com- bined.
EU directives. A code number after the CE mark indicates the responsible notified body. The EU legislation relevant to the product can be found in the EU Decla- ration of Conformity, which can be requested from KARL STORZ. Instructions for use • CLICKLINE Instruments 10 mm • TBF107_EN_V1.0_10-2021_IFU_CE-MDR...
Preparation 5 Preparation 5.1 Unpacking the product Carefully remove the product and accessories from the packaging. Check the delivery for missing items and any possible damage. In the case of damage, hidden defects, and short deliveries, document their nature and extent and contact the manufacturer or supplier immediately.
Preparation Lock the working insert in the bayonet lock of the outer sheath by turning by a quarter turn. Introduce the outer sheath with closed jaws into the completely opened handle as far as it will go. Close the handle completely. The outer sheath is now locked in the handle. ð...
These residues must be removed before HF applications take place. The CLICKLINE instrument can be connected to an HF generator via the HF connection. KARL STORZ recommends using the CLICKLINE instruments with HF devices of the AUTOCON® series from KARL STORZ.
Disassembly 7 Disassembly 7.1 Disassembling CLICKLINE instruments WARNING Hot HF electrodes! Risk of burns! Burns to the user. Let the HF electrodes cool down after use. Open the handle. Press and hold the locking button (do not turn). Open the handle to the maximum in order to pull the outer sheath out toward the front. Hold the jaw tightly and open the bayonet lock by rotating the outer sheath by a quarter turn.
8 Maintenance, servicing, repairs, and disposal 8.1 Repairing the product Repair work may only be performed by KARL STORZ or by a company authorized by KARL STORZ. The interventions described in this instruction manual are exempt from this rule. Please contact your local KARL STORZ subsidiary or authorized dealer (see the list of subsidiaries).
Art. no. Designation 26002 M Unipolar high-frequency cable, with 4 mm plug, length 300 cm, for use with KARL STORZ, Erbe type T devices of older de- sign, and Ellman 26004 M Unipolar high-frequency cable, with 4 mm plug connector, length 300 cm,...
Accessories and spare parts 9.2 Spare parts Art. no. Designation 29100 Sealing cap for LUER lock irrigation connector, black, autoclavable, package of 7909890 Half-shell 7944790 Silicone ring Instructions for use • CLICKLINE Instruments 10 mm • TBF107_EN_V1.0_10-2021_IFU_CE-MDR...