Rhythm Healthcare P2-E6 Instruction Manual

Rhythm Healthcare P2-E6 Instruction Manual

Portable oxygen concentrator
Hide thumbs Also See for P2-E6:

Advertisement

ITEM# P2-E6
P2-E6 PORTABLE
OXYGEN CONCENTRATOR

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the P2-E6 and is the answer not in the manual?

Questions and answers

Elizabeth Marnul
March 2, 2025

I turn on my Rhythm Healthcare Portable Oxygen Concentrator Model P2 and it goes on fine, I get oxygen and then the machine shuts down about 3 minutes later. I checked settings and they say everything is good

1 comments:
Mr. Anderson
March 2, 2025

The Rhythm Healthcare Portable Oxygen Concentrator Model P2 could shut down after 3 minutes (300 seconds) of operation due to the following reasons:

1. Oxygen Concentration Below 87% – If the oxygen concentration remains below 87% for more than 300 seconds, the device may shut down, and the user should contact the provider.
2. Oxygen Concentration Below 50% – If the oxygen concentration drops below 50% continuously for more than 300 seconds, the device will auto-shut down after 30 seconds, and the user should contact the provider.

These conditions indicate a potential issue with oxygen production or delivery, requiring professional inspection.

This answer is automatically generated

Alex Cabrera
April 3, 2025

can this MODEL E2-P6 be used while charging?

1 comments:
Mr. Anderson
April 3, 2025

Yes, the Rhythm Healthcare model P2-E6 can be used while charging with the AC power adapter.

This answer is automatically generated

Linda Waddington
April 21, 2025

My elderly neighbor has this unit. when turned on it reads valve check failure and then shuts off. She is limited income. Has a second machine but she needs this back up. Can it be fixed reasonably?

Summary of Contents for Rhythm Healthcare P2-E6

  • Page 1 ITEM# P2-E6 P2-E6 PORTABLE OXYGEN CONCENTRATOR...
  • Page 2: Table Of Contents

    TABLE OF CONTENTS Symbols ......................1-2 Intended Use ...................... 3 Contraindications ....................4 General Precaution ..................4-7 Parts Diagram ....................8 User Interface Buttons and Display Symbols .......... 9-12 Alerts ......................13-15 Power Supply ....................16-17 Accessories ....................17-18 General Operation ..................19-22 Troubleshooting ....................
  • Page 3: Symbols

    Symbol Description A warning indicates that the personal safety of the patient may be WARNING involved. Disregarding a warning could result in significant injury. A caution indicates that a precaution or service procedure must CAUTION be followed. Disregarding a caution could lead to a minor injury or damage to equipment.
  • Page 4 Symbol Description Serial Number This side up Fragile Storage Humidity (Non-condensing) Storage Temperature MR unsafe...
  • Page 5: Intended Use

    It supplies a high concentration of oxygen and is used with a nasal cannula to deliver oxygen from the concentrator to the patient. The P2-E6 is a small and portable device that may be used at the home and can be taken with you while performing your daily activities.
  • Page 6: Contraindications

    • In certain circumstances, oxygen therapy can be hazardous. Please seek medical advice before using this device. • The P2-E6 is not designed or specified to be used in conjunction with a humidifier, nebulizer or connected with any other equipment. Do not modify the P2-E6 Concentrator. Any modifications performed on the equipment may impair performance or damage equipment and will void your warranty.
  • Page 7 • Must only use the parts and accessories that are provided by the manufacturer. WARNING The settings of the P2-E6 might not correspond with a continuous flow of oxygen. WARNING The settings of other models or brands of oxygen therapy equipment do not correspond...
  • Page 8 WARNING To ensure that you receive the correct therapeutic amount of oxygen delivery according to your medical condition, the P2-E6: • Must be used only after one or more settings have been individually determined or prescribed for you at your specific activity levels.
  • Page 9 • Use only with P2-E6 batteries • Use only approved P2-E6 accessories WARNING Remove the battery from the device if the P2-E6 will not be used for an extended period of time. WARNING Device operation exceeding the voltage, breath rate, temperature, humidity and/or altitude...
  • Page 10: Parts Diagram

    Parts Diagram...
  • Page 11: User Interface Buttons And Display Symbols

    User Interface Buttons and Display Symbols Display Panel: Symbol Description Instructions ON I OFF Press once to turn "ON”. Button Press and hold for one second to turn "OFF" Press once to toggle between audible and silent mode. The panel will display the appropriate icon to indicate which mode is enabled: Audible mode----- Audio Alarm...
  • Page 12 User Interface Buttons and Display Symbols Home Screen Home screen will show the icon as: Icon Description Current flow setting (range is 1 to 6) Battery charge level: • Battery percentage remaining • Battery time remaining (in hours:minutes) Device runtime since powered on (in hours:minutes) Alerts are silenced Alerts are audible...
  • Page 13 User Interface Buttons and Display Symbols ICON DESCRIPTION Powered by AC or DC only Powered by battery only (not plugged in and charging) 1. 2. Battery level percentage and time remaining (in hours:minutes). 2. 3. Device ON Battery is charging. Battery level percentage and estimated time to fully charge the battery.
  • Page 14 User Interface Buttons and Display Symbols Audible Mode: The display below shows that the device has detected an active alarm while in Audible Mode (example below shows "Absence of Breath"): Alert in Silent Mode: The display below shows that the device has detected an active alarm while in Silent Mode (example below shows "Absence of Breath"):...
  • Page 15: Alerts

    Alerts Adapter plug / unplug: An adapter icon displays when adapter is plugged in and disappears when unplugged. An audible alarm (if enabled) will also be heard. Battery plug / unplug A battery icon displays when battery is connected and disappears when disconnected. An audible alarm (if enabled) will also be heard.
  • Page 16 Alerts-continued Display of screen Alarm item Alarm condition System process Battery cycle > 500 Battery Alarm only Battery Exhausted Exhausted Or health < 50% Replace Sieve Bed expired or Replace Sieve Bed Alarm only Sieve Bed Sieve Bed chip error Pls Contact provider Switch battery Adapter input...
  • Page 17 Alerts - continued Display of screen Alarm item Alarm condition System process shut down after Valve Check Fail: Valve Check Fail Valve switch failed Pls Contact Provider shut down after Cooling Fan Fail: Cooling Fan Fail Cooling fan failed Pls Contact Provider Battery Depleted: RSOC: ! 5% shut down after...
  • Page 18: Power Supply

    Power Supply Standard Lithium-Ion Battery # BA-P201 The P2-E6 is powered by a standard lithium-ion battery. When fully charged, the battery can last for up to 5.4 hours of operation. Recharge the battery with the AC or DC adapter. Recharging time is not more than 4 hours.
  • Page 19: Accessories

    Accessories Nasal Cannula The P2-E6 Oxygen Concentrator must use a single lumen nasal cannula to provide oxygen to the patient. WARNING Nasal cannulas should not be used by more than one person. DO NOT use a cannula that has a length exceeding 25ft (7,6m).
  • Page 20 Accessories List Item Nasal Cannula 1 pc NOC-DAW0721 Carry Bag 1 pc CB-P200 AC power supply 1 pc EM11012E DC power supply 1 pc ED1010C Intake Filter 5 pc FI-P201 Battery 1 pc BA-P201...
  • Page 21: General Operation

    General Operation Find a well-ventilated location to place the P2-E6, make sure it's turned off. Be sure the Intake and exhaust has clear access. Put the P2-E6 in a suitable place where any alarms can be heard. WARNING Do not use P2-E6 in the presence of flammable anesthetics, detergents, or other chemical vapors.
  • Page 22 Slide the battery into place until the latch returns to the upper position. There will be an audible sound when the battery is in position. Connect the AC or DC power to P2-E6. The green LED on the power adapter will turn on and there will be an audible sound when power is connected.
  • Page 23 CAUTION Oxygen concentration may not reach specification during the two-minute warm up time. CAUTION 30 seconds after startup, the P2-E6 will enter auto-pulse mode. During these 30 seconds, inhalation will not work.
  • Page 24 Set to the flow rate prescribed by your physician or clinician. Press the + or - setting buttons to adjust the P2-E6 to the desired flow rate. The current setting can be viewed on the display from 1 to 6.
  • Page 25: Troubleshooting

    Troubleshooting The table below lists some common problems and solutions. If you can't resolve a problem, please contact your provider. Problem Possible Cause Recommended Solution Battery is not Remove the battery and re-install it installed correctly correctly. Use the AC or DC power adapter to operate the device (with the battery Battery is depleted inserted) to recharge the battery.
  • Page 26: Cleaning The Case

    Nasal cannula should be rated for 5 liters per minute to ensure proper patient usage and oxygen delivery. Filter Cleaning and Replacement Filters are designed for adequate air flow through the device at the front of the P2-E6. Pre Filter # FI-P202 The particle screen/pre filter must be cleaned weekly to ensure adequate air flow.
  • Page 27: Battery Care And Maintenance

    Keep liquids away from batteries. If batteries get wet, stop using immediately and dispose of battery properly. Battery Replacement Press down on the latch and slide the battery out. Insert the P2-E6 battery by sliding battery into place until the latch returns to the upper position and it clicks into place.
  • Page 28 75°F (24°C) when batteries are being charged. Battery Time Remaining Clock The P2-E6 continuously displays the battery time remaining. This displayed time is only an estimate and the actual time remaining may vary from this value. Store battery in a cool, dry place. Store with a charge of CAUTION 40-50%.
  • Page 29: Disposal Of Equipment And Accessories

    Maintenance and Cleaning of P2-E6 Disposal of Equipment and Accessories Follow your local governing ordinances for disposal and recycling of the P2-E6 accessories. The battery contains lithium-ion cells and should be recycled and must not be incinerated. P2-E6 Replacement Items Item Model No.
  • Page 30: Specifications

    System Specifications P2-E6 Portable Oxygen Concentrator Specifications Dimensions L/W/H:8.70” x 3.35” x 6.30” 22.1cm x 8.5cm x 16.0cm Weight 4.37 pounds 1.98Kg (with battery) User Interface 2.8” LCD color display screen Sound Level 49 dB (A) (on setting 2) Time from Turning...
  • Page 31: Classifications

    System Specifications Concentrator Specifications-Continued Power: AC Power supply AC Input: 100 to 240VAC 50 to 60 Hz DC Input: 11-16VDC DC Output: 19V 6.3A DC Power supply Rechargeable Battery Voltage: 14.4VDC Rated capacity:6.7Ah Up to 5.4 hours at setting 1 Battery Duration Up to 1.5 hours at setting 6 Battery Charging Time...
  • Page 32: Standards Compliance

    System Specifications Standards Compliance The device is designed to conform to the following standards: - IEC 60601-1-2, 2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests - IEC 60601 - 1: Medical Electrical Equipment - part 1: General Requirements for Basic safety &...
  • Page 33: Emc Information

    System Specifications EMC Information The device has been designed to meet EMC standards throughout its Service Life. Guidance and Manufacturer's Declaration - Electromagnetic Immunity: The Concentrator is intended for use in the electromagnetic environment specified below. The user of the Concentrator should make sure it is used in such an environment. Compliance Electromagnetic Immunity Test...
  • Page 34: Electromagentic Emissions

    2.7 GHz Guidance and Manufacturer's Declaration - Electromagnetic Emissions: The P2-E6 is intended for use in the electromagnetic environment specified below. The user of the P2-E6 should ensure it is used in such an environment. Emissions Test Electromagnetic Environment -...
  • Page 35: Warranty

    Warranty The P2-E6 Oxygen Concentrator warranty covers the repair or replacement of the unit. The warranty term is 36 months from the date of shipment. Please contact us by telephone or email to return defective equipment under warranty and to resolve any problems. Our trained technicians will help you with any questions or problems with your POC.
  • Page 36 Rhythm Healthcare, LLC 3200 Tyrone Blvd N St. Petersburg, FL 33710 Email: contactus rhythmhc.com Website: www.rhythmhc.com V1: 08/08/22...

Table of Contents

Save PDF