To Adjust Options:

From the HOME screen, touch the Adjust Options button to adjust the handpiece options (buttons and trigger, such as
handswitch). The available handpiece options will depend on the type of handpiece that is identified at the top of the
screen.
NOTE:
To program the footswitch pedals and pads, touch the Footswitch Options button.
Adjust Acceleration
• Touch the plus and minus buttons or slider bar to
adjust the percent of acceleration [0 to 100%].
• At 100%, the handpiece speed accelerates quickly.
Acceleration slows as the setting approaches zero
percent.
Adjust options - FORMULA HANDPIECE - 3rd Port
46 %
Accelerate
36 %
Brake
Osc
10
Turns
Adjust Braking
• Touch the plus and minus buttons or slider bar
to adjust the percent of braking [0 to 100%].
• At 100%, the handpiece stops abruptly. The
amount of deceleration decreases as the setting
approaches zero percent.
Adjust Oscillate Turns
• Touch the plus and minus buttons to select the number of turns
per direction when running a handpiece in the oscillate mode.
• Touch the minus button until the TIME-BASED icon is visible to
operate in a TIME-BASED mode where the number of turns
are based on the speed of the handpiece.

Figure 10. Adjust Options Screen

19
Select Run Mode
• FIXED: handpiece runs at the full set speed or power.
Varying pressure on the handswitch (or trigger) will not
vary the handpiece speed or power.
• VARIABLE: handpiece speed responds to the degree
of pressure applied to the handswitch (or trigger).
• ONE TOUCH: Handpiece activation responds to one
touch of a trigger device (handswitch) and continues to
run when the trigger device is released. Handpiece is
deactivated by touching any trigger
device. In this mode, the handpiece
operates at constant set-point
speed/power level.