(BLOS), the surgical display also ensures a long lifetime and low power consumption. Ease of installation The MDSC-8532 comes with a smart cable management system that hides the cables for a clutter-free set-up. Equipped with VESA 100 interface to allow easy mounting on surgical booms and spring arms. Available in different models, this surgical display also features a host of connectivity options and remote control.
1.2 About this user guide Overview This manual provides support to the user during the installation, set up and utilization of the MDSC-8532 display. Depending on the specific version that has been purchased, some of the features and options described in this document may not apply to the display in user’s hands.
8. 3G-SDI out (**) 9. DVI-D in 10. HDMI 2 in 11. HDMI 1 in 12. DisplayPort 1.2 in 13. DisplayPort 1.2 out 14. USB 2.0 type A interface 15. USB 2.0 type B interface 12GP version only. R5914105 /01 MDSC-8532...
“Picture-by-Picture (PBP)”, page 33 for more information. Beside the video input connections, the MDSC-8532 also has video output capabilities allowing you to loop- through or duplicate all video inputs connected with the MDSC-8532 to another display, projector, video recorder, ...
A subset of the commands of the remote control protocol is also available on a new DDC protocol on DVI and DisplayPort1 auxiliary channel. 2.2.2 MDSC-8532 12GP version To connect the interfaces 1. Connect one or more video source(s) to the corresponding video inputs of the display. For a list of supported video inputs, see “Technical specifications”, page...
CAUTION: To avoid risk of electric shock, the external DC power supply must be connected to a mains with protective earth. The ground connection on the display’s DC power input connector has no protective earth function. The MDSC-8532 display protective earth connection is provided via a dedicate pin (see next steps).
Display installation Protective earth Earth the MDSC-8532 by connecting the protective earth pin to a grounded outlet by means of a yellow/green AWG18 wire (maximum admitted cable length according to national regulation requirements). Image 2–5 CAUTION: The display must be earthed.
CAUTION: Use an arm that is in compliance with VESA requirements. CAUTION: The monitor VESA interface has been designed for a safety factor 6 (to support 6 times the monitor weight). In the medical system, use an arm with suitable safety factor (IEC60601–1). R5914105 /01 MDSC-8532...
After unlocking the front keyboard, all keys will light up. Touching any of these keys while the backlight is on will execute the function of the key. However, if no further action is taken within the time defined by the OSD time-out parameter, the keys will dim again and the keyboard will re-lock. R5914105 /01 MDSC-8532...
About the front menu The MDSC-8532 has a front menu which is used to activate the OSD main menu but which also supports 3 user key functions. These user keys allow to immediately activate a commonly used function without having to access and browse the OSD menus.
Page 22
Select a gamma preset selectable options in the fron menu is greater than 4) Select a main input source Enable/disable horizontal flip Adjust the brightness level Select a Picture in Picture mode Select a color space preset R5914105 /01 MDSC-8532...
1024x768 @ 100 Hz Picture Pro le Picture Brightness advanced Contrast Input select Saturation Con g. Color Temperature 6500K Color Space ITU709 System Gamma Native Sharpness Image 3–4 1. Menu pages 2. Sub-menus (menu items) 3. Status bar R5914105 /01 MDSC-8532...
To unlock the menu 1. Unlock the front keyboard (see “Front keyboard locking/unlocking”, page 20). 2. When the OSD lock window appears, press the following key sequence to unlock the menu: left - right - right - up R5914105 /01 MDSC-8532...
Advanced operation Not all features described in the chapter “Advanced operation” are available. The features not available in specific software versions are shown on OSD menu in light-grey. R5914105 /01 MDSC-8532...
To adjust the contrast level 1. Bring up the OSD main menu. 2. Navigate to the Picture menu. 3. Enter the Contrast submenu. The command bar Contrast is highlighted. 4. Set the contrast level as desired and confirm. R5914105 /01 MDSC-8532...
BT.2020 reproducible colors are within the limit of the LCD panel color gamut. Factory calibration – Color space: RGB primary calibration, according to the selected standard, is performed within the physical limitation of the LCD panel used. R5914105 /01 MDSC-8532...
To adjust the sharpness level 1. Bring up the OSD main menu. 2. Navigate to the Picture menu. 3. Enter the Sharpness submenu. The command bar Sharpness is highlighted. 4. Set the sharpness level as desired and confirm. R5914105 /01 MDSC-8532...
2. Navigate to the Picture Advanced menu. 3. Enter the Zoom submenu. The zoom command bar is highlighted. 4. Adjust the zoom level as desired and confirm. 4.2.4 Pan This function is not available when DisplayPort mode is set to DP 1.2 MST. R5914105 /01 MDSC-8532...
About main sources The available main sources for your display are: • • SDI 4K (*) • • DisplayPort • HDMI-1 • HDMI-2 • Auto Search 12GP version only. Available main source options may differ depending on display model. R5914105 /01 MDSC-8532...
Please refer to the technical specifications for an overview of accepted video formats. To select the DisplayPort mode 1. Bring up the OSD main menu. 2. Navigate to the Input Select menu. 3. Enter the DP mode submenu. 4. Select one of the available DisplayPort modes and confirm. R5914105 /01 MDSC-8532...
12GP version only. To select the failover input 1. Bring up the OSD main menu. 2. Navigate to the Input Select menu. 3. Enter the Failover Input submenu. 4. Select one of the available failover inputs and confirm. R5914105 /01 MDSC-8532...
4K-SDI (**) No PBP with DP 1.2 MST (**) MDSC-8532 12GP version Primary input up to 4K resized to half screen on left side. Secondary input up to 4K resized to half screen on right side. The Video Parameters applied to the Primary input Source are also applied to the Secondary input.
4K-SDI (**) No PIP with DP 1.2 MST (**) MDSC-8532 12GP version Gamma and Color temperature for the PiP Source are always set to Native and 6500 K independently from the Transfer Function applied to the Primary input Source. To select Picture-in-Picture input 1.
User 1, User 2 or User 3 profile. The Factory and X Ray profiles can be modified, but the factory default can’t be overwritten and can always be recalled through the recall profile menu item. R5914105 /01 MDSC-8532...
4. Select one of the available profiles to save and confirm. 4.4.6 Front menu user keys About front menu user keys The MDSC-8532 has a front menu (see “Front menu ”, page 21) which supports 3 user key functions. These user keys allow to immediately activate a commonly used function without having to access and browse the OSD menus.
Front lock: Front Menu and OSD menu are accessible though the rear keyboard only To enable/disable control locking 1. Bring up the OSD main menu. 2. Navigate to the System menu. 3. Enter the Control Lock submenu. 4. Select one of the available options and confirm. R5914105 /01 MDSC-8532...
About display status This information shows the active values of the main display parameters. To consult display status 1. Bring up the OSD main menu. 2. Navigate to the System menu. 3. The display status information is shown. R5914105 /01 MDSC-8532...
The display can be used in failsafe mode as corner “Failsafe mode” is visible, even after temporary backup, running a factory FW (see the rebooting the monitor. FW version on OSD menu). Please contact the Service Center. R5914105 /01 MDSC-8532...
(no applied part). • The equipment shall not be used with life support equipment. • The user should not touch the equipment, nor its signal input ports (SIP)/signal output ports (SOP) and the patient at the same time. R5914105 /01 MDSC-8532...
Page 43
Never use the monitor near water - e.g. near a bathtub, washbasin, swimming pool, kitchen sink, laundry tub or in a wet basement. • The equipment is IP21 (IP45 front side only) compliant with a tilt of ± 10°. The PSU only is IP20 compliant. R5914105 /01 MDSC-8532...
Page 44
Sufficient expertise is required for installing this equipment, especially to determine the strength of the wall for withstanding the display's weight. Be sure to entrust the attachment of this equipment to the wall to licensed contractors of Barco and pay adequate attention to safety during the installation and usage. •...
6.2 Cybersecurity Hospital IT security To prevent unauthorized access to the device, the organization incorporating the MDSC-8532 in their IT network shall have the necessary state-of-the-art policies, processes, standards and other security measures in place to incorporate, support and protect the device into the IT network. This shall include the application of risk management (e.g.
Page 46
Mainland, marked with the Environmental Friendly Use Period (EFUP) logo. The number inside the EFUP logo that Barco uses (please refer to the photo) is based on the “General guidelines of environment-friendly use period of electronic information products” of Chinese Mainland.
The device is intended to display medical images from medical systems such as endoscopic or laparoscopic cameras, surgical microscopes, room and boom cameras, ultrasound, image guided therapy and intervention, PACS, anesthesiology, patient information and other compatible medical image systems. R5914105 /01 MDSC-8532...
Page 48
The manufacturing country of the product is indicated on the product label (“Made in …”). Importers contact information To find your local importer, contact one of Barco’s regional offices via the contact information provided on our website (www.barco.com). Factory address Fimi S.r.l., Via Saul Banfi 1, 21047 Saronno, VA, Italy...
8532, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result. Electromagnetic emissions The MDSC-8532 is intended for use in the electromagnetic environment specified below. The customer or the user of the MDSC-8532 should assure that it is used in such an environment. Electromagnetic environment –...
Page 50
Consult the dealer or an experienced technician for help. Electromagnetic immunity The MDSC-8532 is intended for use in the electromagnetic environment specified below. The customer or the user of the MDSC-8532 should assure that it is used in such an environment.
Indicates compliance with Part 15 of the FCC rules (Class A or Class B). Indicates the device is approved according to the UL Recognition regulations. E346057 MEDICAL – GENERAL MEDICAL EQUIPMENT AS TO ELECTRICAL SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH ANSI/AAMI AS60601-1:2005/(R)2012, CSA CAN/CSA-C22.2 E346057 NO. 60601-1:14 R5914105 /01 MDSC-8532...
Page 53
Caution: Federal law (United Stated of America) restricts this device to sale by or on the order of a licensed healthcare practitioner. Indicates the device is approved according to the BIS regulations. IS 13252 (Part 1) IEC 60950-1 R-xxxxxxxx www.bis.gov.in Indicates the device is approved according to the INMETRO regulations. R5914105 /01 MDSC-8532...
Page 54
Indicates the Unique Device Identifier. Indicates the Authorised Representative for the European Union. Indicates the Authorised Representative for Switzerland. Warning: dangerous voltage Caution Consult the Instructions For Use. Values for xx and yy can be found in the technical specifications paragraph. R5914105 /01 MDSC-8532...
Page 55
Indicates that the box should not be cut with a knife, a cutter or any other sharp object. Indicates the temperature limits to which the device can be safely exposed when being stored. +60 °C -20 °C R5914105 /01 MDSC-8532...
Barco software products are the property of Barco. They are distributed under copyright by Barco NV or Barco Inc., for use only under the specific terms of a software license agreement between Barco NV or Barco Inc. and the licensee. No other use, duplication, or disclosure of a Barco software product, in any form, is authorized.
Page 57
2-side anti-reflective alkali-aluminosilicate glass with anti-fingerprint coating Keyboard Front: 5-key capacitive touch - user programmable Rear: 5-key membrane Video inputs MDSC-8532 • 4K-UHD input SSTP 1x DP 1.2 SST / MST up to 3840 x 2160 @ 50/60 Hz (user configurable) 2x HDMI 2.0 up to 3840 x 2160 @ 50/60 Hz...
Page 58
CAN/CSA-C22.2 No. 60601-1: 14 • IEC 60601-1: 2012 Edition 3.1 • EN 60601-1: 2006 + A1:2013 • IEC 60601-1-2 (2014) • EN 60601-1-2 (2015) • FCC CFR 47 Part 15 Subpart B (Class B) • ICES-003 (Class B) R5914105 /01 MDSC-8532...
Page 59
3 years Supported timings Full HD and 4MP timings Format HDMI 720x487@59.94Hz (NTSC) 720x480p@59.94Hz 720x480p@60.00Hz 720x576i@50.00Hz (PAL) 720x576p@50.00Hz 800x600@60.00Hz 800x600@75.00Hz 1024x768@60.00Hz 1024x768@70.00Hz 1024x768@75.00Hz 1152x864@75.00Hz 1280x720p@50.00Hz 1280x720p@59.94Hz 1280x720p@60.00Hz 1280x1024p@60.0Hz 1400x1050p@60.00Hz 1600x1200p@60.00Hz 1680x1050p@60.00Hz 1920x1080p@29.97Hz 1920x1080p@30.00Hz 1920x1080p@50.00Hz 1920x1080p@59.94Hz 1920x1080p@60.00Hz 1920x1200p@60.00Hz R5914105 /01 MDSC-8532...
Need help?
Do you have a question about the MDSC-8532 and is the answer not in the manual?
Questions and answers