50 to 500 mL/min (0.01 to 0.1 gal/min) Flow rate for -21, 3.8 to 45.4 L/min (1 to 12 gal/min) PC9001-71, PC9001-81: 12 to 100 psi Sample pressure Sample temperature 0 to 60 °C (32 to 140 °F) Sample pressure...
Specification Details Warranty 1 year Certifications CE and FDA Accession No. 9320350-008 Viscosities tested at ambient temperature: 25 °C ±2 degrees; 77 °F ±2 degrees General information In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential damages resulting from any defect or omission in this manual.
Page 4
This symbol indicates a laser device is used in the equipment. Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/98/EC), European electrical equipment users must now return old or end-of-life equipment to the Producer for disposal at no charge to the user.
Page 8
Figure 5 Cable connections 1 9 to 33 VDC 2 Alarm driver Table 1 Male connector wires—DB15 cable Pin number Color Function White Power input (9 to 33 VDC) Brown Power ground (return) Green Alarm driver Yellow — Gray RS232-RXD (receive data) Pink RS232-TXD (transmit data) Blue...
Page 9
Table 2 Female connector wires—DB15 cable (continued) Pin number Color Function Pink RS232-TXD (transmit data) Blue — — Orange RS232-GND (signal return) RS485-A (signal A of differential serial driver pair) Black RS485-B (signal B of differential serial driver pair) Violet RS485-SGND (shield ground connection) —...
6–14 or 14–21). Operation Configure the instrument 1. Open the PC9001 Configuration Utility. A window opens (Figure 2. Select the communication (COM) port on the computer that is connected to the counter. 3. Change the Node ID setting to the Node ID of the instrument. The default Node ID setting for new instruments is 138.
2. Select the Alarms tab. 3. Make changes to these fields: • Standard (ISO, NAS or SAE) • Alarm Direction (Clean to Dirty or Dirty to Clean) • Code Limits (4, 6, 14 or 21 µm) If NAS is selected as the Standard, only the code limits for 6– 14 µm and 14–21 µm will be available.
The instrument cannot be calibrated by the user. Contact the manufacturer for instrument calibration. Troubleshooting Error indicator When an error occurs, an exclamation point appears on the main menu of the PC9001 Configuration Utility. 1. Double-click on the exclamation point. A window opens (Figure 11).
Table 9 To change Filtertechnik codes to ISO codes, delete all of the digits to the right of the decimal point. Do not change the value either up or down. For example, a Filtertechnik code readout of 7.2 is equivalent to an ISO code of 7.