Atracsys fusionTrack 500 User Manual page 48

Table of Contents

Advertisement

Name
Image
Compression
Threshold
Image Integration Time
Temperature sensor in-
dex
Temperature sensor lo-
cation
Temperature
sensor
value
Temperature sensor ref-
erence
Calibration type
Calibration
processing
datetime
Enables IR strobe
Strobe mode
Type
Description
int32,
Allows to access the picture compression thresh-
read / write,
old, i.e. minimum pixel intensity for a pixel to
permanent
be considered white. Increasing the value de-
creases the compressed picture size.
int32,
Allows to access the sensor integration time, i.e.
read / write,
the exposure time for the pictures. If the strobing
permanent
is enabled, the strobe time is automatically syn-
chronised.
int32,
Access the index of the temperature sensor to
read / write
be accessed with the 'Temperature sensor loca-
tion', 'Temperature sensor value' and 'Tempera-
ture sensor reference' options. Reading a value
before setting a valid index will result in an error.
ftkBuffer,
Allows to read the location of the temperature
read-only
sensor for the index set with the 'Temperature
sensor index' option. Reading a value before set-
ting a valid index will result in an error.
float32,
Allows to read the current temperature sensor
read-only
reading for the index set with the 'Temperature
sensor index' option. Reading a value before set-
ting a valid index will result in an error.
float32,
Allows to read the median temperature sensor
read-only
value for the index set with the 'Temperature sen-
sor index' option, during the calibration. Reading
a value before setting a valid index will result in an
error.
int32,
Allows to read the type of the calibration file
read-only
stored in the device. The value corresponds to
the CalibrationType enumeration.
int32,
Allows to read the date and time of the calibration
read-only
processing (given in UTC).
int32,
Allows to toggle on/off the IR strobe on each
read / write,
frame.This option is only available until firmware
permanent
version 0x1.1.6.d2.
int32,
Allows to switch between the different strobing
read / write,
modes:
permanent
0: strobe and picture acquisition are synchro-
nised;
1: strobe and picture acquisition are desyn-
chronised;
2: strobe and picture acquisition are synchro-
nised on every second acquired frame.
This option is only available with firmware version
0x1.1.8.3 and newer.
User manual
Version: v4.5.2

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the fusionTrack 500 and is the answer not in the manual?

Questions and answers

Table of Contents

Save PDF