In case of non-observance of the operating instruction the manufacturer cannot assume any liability or warranty. The copyright of this documentation remains with DILO Armaturen und Anlagen GmbH, D-87727 Babenhausen. Any use not permitted by the author (multiplication, leaving to third persons etc.) is subject to penalty and indemnification.
The corresponding symbol which is used cannot replace the text of the safety note. Therefore the complete text must always be read. This symbol is not for safety notes, but for information to understand the device better. www.dilo-gmbh.com C 3970-00 Page 3 / 12...
Installation Only operate the device indoors or in dry places outdoors. The ambient temperature during operation of the device should be in the range of -20°C to Warning +50 °C. www.dilo-gmbh.com C 3970-00 Page 4 / 12...
2.1 Technical data -LeakPointer Dimensions of the device: Length (L): 254 mm / 10” Width (B): 59 mm / 2.3 “ Height (H): 51 mm / 2” Weight: 0.5 kg / 1lbs. 2 oz. www.dilo-gmbh.com C 3970-00 Page 5 / 12...
Page 6
Sensitivity: Fully automatic sensitivity selection with 6 visible LED range indicators. Maximum sensitivity allows detection of approx. 5 g/yr. Audible alarms: Two user selectable audible alarm options (High and low intensity alarm tones) www.dilo-gmbh.com C 3970-00 Page 6 / 12...
See Maintenance section for LO BAT conditions. 2.2 Standard accessories Carrier case 1.5 V alkaline batteries Spare sensor Operating manual (multilingual) on CD-ROM www.dilo-gmbh.com C 3970-00 Page 7 / 12...
Enables user to pinpoint the exact location of a leak source. LED display for Leak Sizing Allows user to determine the approximate size of a leak. One Button Operation For reset, change audible alarm, and shut off. www.dilo-gmbh.com C 3970-00 Page 8 / 12...
OFF. LOW BATTERY INDICATOR When the batteries reach approximately 10% of capacity, the rightmost tachometer LED will stay lit above the LO BAT legend. www.dilo-gmbh.com C 3970-00 Page 9 / 12...
500 mm/s (1-2 in/s), and no more than 5 mm (0.2”) from the surface, completely around the position. Slower and closer movement of the probe greatly improves the likelihood of finding a leak. www.dilo-gmbh.com C 3970-00 Page 10 / 12...
If a suspected leak is indicated, verify several times by moving the sensor away from the leak area, resetting the unit and then back to the suspected leak. If the instrument indicates a leak three consecutive times, then you have found a leak. www.dilo-gmbh.com C 3970-00 Page 11 / 12...
Eventually, the sensor will need to be replaced. This condition is usually indicated by abnormal or erratic performance. With the unit OFF, the old sensor can be removed by screwing off and the new one can be screwed on afterwards. DILO recommends that you always have a spare sensor available for replacement in the field.