Summary of Contents for lec medical Pharmacy Plus Series
Page 1
USER MANUAL Pharmacy Plus - Laboratory Plus - Pharmacy Connect Operating Instructions 083870100 - Rev05 - 27/07/22 www.lec-medical.co.uk The Experts in Medical Refrigeration...
7: Before Switching On ............................. 15 8: Operation ................................16 IJ Controller Overview ............................16 9: Basic Operations using the LEC Medical App ....................17 9:1 Download and install the LEC Medical App ....................17 9:2 Connecting to an appliance ........................17 9:3 App Home Page ............................
2: Safety Instructions In this manual, the following symbols and conventions are used: - This symbol when used alone indicates important operating instructions which reduce the risk of injury or poor performance of the unit. WARNING: A situation which if not avoided, could result in serious injury or death. CAUTION: A situation which, if not avoided, may result in minor or moderate injury, impaired performance, or damage to the equipment.
3: General Recommendations Intended Use and Special Warning This equipment is tested in compliance with established regulations and then shipped ready for use. This equipment is intended for cold storage and is not to be used in the presence of explosive gases or mixtures and/or close to sources with high magnetic or electric fields/ WARNING: These units are not medical WARNING: Do not damage the...
4: Product Specifications This range of LEC refrigerators and freezers have been designed and built with the intention of the product being used within pharmacies, hospitals, and laboratories. Please see below for specific product information. Pharmacy Plus and Neonatal Range Refrigerators Depth Height Width...
Page 8
Product Specifications cont. Pharmacy Connect Range Refrigerators Depth Height Width Weight Rating Power Temperature Ambient SKU No Model Refrigerant Inc Handle (mm) (mm) (Kg) (Amp) Range Temperature (mm) 444411319 PCSR158CT-UK 18.5 0.78 2⁰C to 8⁰C 16⁰C to 32⁰C R600a 444411321 PCSR158CT-EU 18.5 0.78...
Product Specifications cont. Laboratory Plus Fridge/Freezer Combi Depth Height Width Weight Rating Power Temperature Ambient SKU No Model Refrigerant Inc Handle (mm) (mm) (Kg) (Amp) Range Temperature (mm) Fr +2⁰C to +10⁰C 444411352 LSFSF280BT-UK 16⁰C to 32⁰C R600a Fz -18⁰C to -25⁰C Fr +2⁰C to +10⁰C 444411353 LSFSF280BT-EU...
BLUE: NEUTRAL • DO NOT modify system components, BROWN: LIVE especially the controller. Use LEC Medical exact replacement equipment or parts. Before These colours might not correspond with the use, confirm that the product has not been colour markings identifying the terminals in your altered in anyway plug.
Electrical Connections 1: Green Yellow 2: Blue 3: Brown 4: 13 Amp Fuse 5: Cord Clamp Note: First Connection to the mains. The equipment provides information about the higher temperature (Hi alarm) till the moment of cooling. The alarm is signalized by a sound signal and “Hi” caption on the display. To mute the alarm touch anywhere on the front surface of the controller.
7: Before Switching On Thank you for purchasing this LEC Pharmacy/Laboratory Refrigerator Before you first use this pharmacy refrigerator, please carry out the following actions: - • Check that the refrigerator has not been damaged in any way during transportation. •...
8: Operation IJ Controller Overview Air Temp Displayed Compressor Running Scroll Up Arrow Toggle Air/Load Display Display Max Temp Navi Button (Enter) Reset Air & Load Temp Unit ON/OFF Display Min Temp Scroll Down Arrow Evaporator Fan Running Load Temp Displayed Controller in alarm Condition Main Display Symbol...
9: Basic Operations using the LEC Medical App 9:1 Download and install the LEC Medical App LEC Medical recommends using the dedicated LEC Medical App to access the data recorded in the controller, scan the appropriate QR code to Download the LEC Medical App.
APP Home Page and controller displays as shown below, we recommend changing the Connected to: Carel_00000***** to a name of your choice see page **. For other app functions see section 10: Controller will flash bLE when connected to the LEC Medical App Current Air Temperature Current Load Temperature Control Temperature Min/Max Temperatures...
LEC Medical Bluetooth enabled appliance is present. To rename the Appliance press the WHITE TEXT button next to “Connected to:” this is usually the controller serial number then enter the appliance name (Customer choice).
The factory set point will maintain the air temperature of the refrigerator and the load temperature. In some extreme ambient temperatures, it may be necessary to adjust the set point. Using the LEC Medical App this a very simple procedure.
9:9 Switching the appliance on and off Control of the appliance can be switched on and off via the app by pressing on/off symbol as shown here. WARNING! This switch acts as a toggle switch, for example If the appliance is ON at the time of pressing the on/off symbol, control of the appliance will be switched OFF. If the appliance is OFF at the time of pressing the on/off symbol, control of the appliance will be switched ON.
11: Operations using the controller touch screen 11:1 Unlocking the controller 3 Secs 11:2 Viewing and re-Setting Min/Max Temperatures on the controller Air Temp Displayed Scroll Up Arrow Toggle Air/Load Display Max Temp Displayed Navi Button (Enter) Reset Air & Load Temp Min Temp Displayed Scroll Down Arrow 11:2:1 Select air or load temperature...
Both air and load maximum and minimum temperatures will be “RESET” Please note: The Minimum and Maximum temperatures are stored in the controller for approximately 3 months on a FIFO bases. Use the LEC Medical app to retrieve the temperatures. (Sections 10: *)
11:3 Changing the Set Point Unlock the controller 11:1 Press Navi button Press Navi button Press Navi button Press to enter USER PASSWORD “11” Press Navi button Press Navi button Press Navi button Press to change to set point Press Press Navi button Press...
14: Disposal of your refrigerator WEEE: Disposal at end of life. WARNING: This appliance contains insulation, gases and refrigerant and must be disposed of in the appropriate manner. Please ensure the appliance is disconnected, emptied and clearly labelled by an authorized person.
Spares and After Sales Service: Email: info@gdha.com Tel: 0344 815 3742 Fax: 0844 248 4123 Address: Lec Medical Stoney Lane Prescot Merseyside, L35 2XW Warranty For details on the warranty of this appliance and any other information please visit: - https://www.lec-medical.co.uk...