SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL 40 Albion Rd, Suite 100 Lincoln, RI 02865 Tel: 401-333-8580 Fax: 401-333-5550 ALPHA OMEGA Email: contact@aoi-corp.com salescontact@aoi-corp.com INSTRUMENTS VER 1.3 11/06...
Page 2
Omega Instruments Corp. be liable for any damages, whether direct or indirect, special, consequential, or incidental arising from the use of this manual. Some states in the USA do not allow the exclusion of incidental or consequential damages. Alpha Omega Instruments...
Page 3
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL WARNINGS Explanation of graphic symbols THIS SYMBOL IS INTENDED TO ALERT THE USER TO THE PRESENCE OF IMPORTANT OPERATING AND MAINTENANCE INSTRUCTIONS. To avoid the risk of fire or electric shock, do not expose the Series 9510 Carbon Dioxide Analyzer to rain or water spray.
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL TABLE OF CONTENTS PAGE WARRANTY WARNINGS SECTION 1.0 INTRODUCTION General Description Optional Equipment Descriptions 1.2.1 Open Diffuser Configuration 1.2.2 Battery Operation 1.2.3 Serial Communications / RS232C 1.2.4 Serial Communications / RS485 1.2.5...
Page 5
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL SECTION 3.0 OPERATION Preparation for Operation 3.1.1 AC Input Voltage Selection 3.1.2 Mounting Configurations 3.1.3 Initial Check Operating Procedures 3.2.1 Power ON 3.2.2 Front Panel Controls and Indicators 3.2.3...
Page 6
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL SECTION 4.0 CALIBRATION PROCEDURES Routine Calibration Check Calibration Gas No Zero Check Required Procedure for Checking Calibration with a Flow Through System Procedure for Checking Calibration with an Open Diffuser System...
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL SECTION 1.0 INTRODUCTION 1.1 General Description The Series 9510 Carbon Dioxide (CO ) Analyzer is a microprocessor based instrument providing accurate and repeatable trace or percent carbon dioxide measurements. The sensor used in the Series 9510 is a non-dispersive infrared type, which does not react chemically with the sample gas.
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL 1.2.2 Optional Battery Operation P/N 95BAT The Series 9510 can be equipped so that it may be powered from a normal AC/DC source as well as an optional built-in battery pack.The batteries are installed at the factory and are designed for applications where the user desires to power the analyzer for short periods using the batteries.
INSTRUCTION MANUAL Non-Dispersive Infrared Sensor for Carbon Dioxide Alpha Omega Instruments uses a non-dispersive infrared (NDIR) sensor to determine carbon dioxide concentrations in gases. Infrared gas sensors, which are comprised of solid state devices, do not chemically react with the gas.
Page 10
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL (-KCL Sensing lamp signal (-KCL Reference lamp signal K= absorption constant C= gas concentration I= IR intensity emitted by the source {-KC(L The ratio “R” (S ) = (l...
2.1 Unpacking the Analyzer Upon receiving the shipping container, carefully examine the packaging to check if the outer surfaces have been damaged. If so, report the findings immediately to Alpha Omega Instruments who will, in turn, provide further instructions. NOTE: IF DAMAGE TO THE SHIPPING CONTAINER HAS BEEN FOUND, DO NOT PROCEED FURTHER.
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL 2.2.2 CO Sensor Wiring of the CO sensor has been done at the factory. 2.2.3 Wiring Alarm Relays Access to the control signals generated from the Series 9510 Carbon Dioxide Analyzer is accomplished using customer-supplied signal cable.
Page 15
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL TABLE 1.0 illustrates the various wiring configurations for the four alarms in the a nalyzer based on whether the alarms are going to be configured for fail-safe or non fail-safe operation.
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL 2.2.4 Wiring 4-20 mADC and 0-2 VDC Outputs The Series 9510 Carbon Dioxide Analyzer has two linear outputs, 4-20 mADC and 0-2 VDC over the instrument's range. These outputs can be measured simultaneously. To wire to the 4-20 mADC output, wire to terminals 12 (4-20) and 13 (AGND).
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL SECTION 3.0 OPERATION 3.1 Preparation for Operation 3.1.1 AC Input Voltage Selection The Series 9510 Carbon Dioxide Analyzer has an ON/OFF power switch on the side of the instrument.
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL Wall Mounted Electronics with Remote Sensor Enclosure. Please refer to Appendix D 3.1.3 Initial Check The Series 9510 Carbon Dioxide Analyzer is ready to use out of the shipping container. The analyzer has been calibrated at the factory and, under normal conditions, re -calibration is not required at initial start-up.
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL 3.2 Operating Procedures 3.2.1 Power ON Power to the instrument is applied when the power cord is plugged into an AC outlet and the power switch on the left side of the instrument is switched ON. The Series 9510 Carbon Dioxide Analyzer is equipped with battery backed RAM so when the instrument is initialized, all values set by the user via the front panel or optional RS-232C or RS-485 inputs will be maintained if switch #8 of SW6 is set to the "ON"...
Page 20
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL 3.2.3 Scaling Analog Outputs of the Analyzer For certain applications, it may be desirable to change the analog output of the Series 9510 Carbon Dioxide Analyzer over a narrower range of measurement. Users wishing to do so can make this change quite easily using the instrument’s front panel adjustment.
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL Alarm System 3.3.1 General Description of Alarm Processing The Series 9510 Carbon Dioxide Analyzer is equipped with four single pole double throw (SPDT) relays with Form C contacts rated at 10 amperes @ 30 VDC and 115/230 VAC. All alarm relays are user configurable with the front panel controls of the instrument and the configuration switch SW6.
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL 3.3.3 Fail Safe Switches #1 through #4 control the fail-safe operation for each of the four alarm relays. Each switch must be turned "ON" to be fail-safe (refer to FIGURE 3.0). In the fail-safe mode, the normal or non-alarm state of each of the four (4) alarm relays is energized.
2 pounds per square inch (PSI). If the pressure is expected to be in excess of 2 psi, it is advisable to use a pressure regulator (Alpha Omega Instruments Model Number 9PRR or 9LPR). Flow rates through the analyzer should be between 100 to 300 milliliters per minute (ml/min.) for precent systems and approximately 150 millimeters a minute for trace systems.
The frequency of these checks is often determined by in-house calibration protocols. If none exists, Alpha Omega Instruments Corp. recommends that a calibration check be made on an average of once every 6 months.
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL Note: calibration gas flowing through the carbon dioxide sensor is discharged inside the analyzer enclosure. Therefore, it is imperative that no combustible gases i.e. methane, hydrogen, etc. be used for calibration.
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL SECTION 5.0 MAINTENANCE The Series 9510 Carbon Dioxide Analyzer requires very little maintenance or checking. 5.1 Calibration Check The user should check the calibration of the analyzer in accordance with established protocol, and adjust the sensor output as required (please refer to Section 4.0).
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL SECTION 6.0 BASIC SERIAL COMMUNICATIONS FOR OPTIONAL RS-232C OR RS-485 6.1 Baud Rates The following baud rates may be used with the Series 9510 CO Analyzer: Baud Rates Available...
Page 28
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL An example of each command is as follows: 'A' Command - Alarm set point with low or high alarm option. Alarm #1 will be set to go off in the case of the CO level dropping below 20 ppm.
Page 29
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL In this example, the user would just type 'E' by itself. This would keep people from inadvertently changing the system settings. However, if the need should arise to change a setting, all that would be necessary would be a 'D' command with no pass code.
Page 30
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL The next morning, the technicians find the analyzer reading 50.0 ppm, but alarms are going off! Alarm 1 and Alarm 2 are both on. This is very meaningful to them, because it means the CO level in the room increased to 52 ppm, but stayed below 53 ppm.
Page 31
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL SL100 <Enter> Sets the High CO scaled output level to 100 O.K. S <Enter> Outl = 10 Outh = 100 For this example, at a CO2 level of 10 (PPM or %), the analog output reads 0.0 VDC and 4 ma, and at 100 (PPM or %, the analog output reads 2 VDC and 20 ma.
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL RS232/485 Connections Connect the Series 9510 Carbon Dioxide Analyzer to a terminal or computer as follows: Series 9510 Computer...
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL SECTION 7.0 ENHANCED REMOTE CONTROL COMMANDS WITH RS-485 General Please read the section on serial communications first to familiarize yourself to the command structure. All commands under RS485 are either exactly the same or have slight enhancements over RS232C. This section is devoted to specifying only the enhancements associated with the RS485 protocol.
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL Please refer to Section 6.2 for information on the following commands. 'A' Command - Same as RS232 'B' Command - Same as RS232 'C' Command - Same as RS232...
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL GFS1=1 <Enter> GFS2=1 <Enter> GFS3=1 <Enter> GFS4=1 <Enter> Note how each Alarm is treated individually. The following commands are not available: GH <Enter> does nothing. GI <Enter> does nothing.
Page 36
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL You can substitute the above, "Unit#1" with any string you like. Usually the shorter the better for purposes of saving keystrokes when accessing the unit. Now type the following: U <Enter>...
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL RS-485 Enhanced Commands 'M' Command - Manual / Automatic clearing of alarms. The enhancement to this command is using the 'name' as a variable. For instance, no matter what unit is...
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL APPENDIX B PANEL MOUNTING CONFIGURATION PART NUMBERS PNL, PNR, & PTX FRONT VIEW...
Page 40
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL APPENDIX B PANEL MOUNTING CONFIGURATION PART NUMBERS PNL, PNR, & PTX SIDE VIEW...
Page 41
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL APPENDIX B PANEL MOUNTING CONFIGURATION PART NUMBERS PNL, PNR, & PTX REAR VIEW...
Page 42
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL APPENDIX B PANEL MOUNTING CONFIGURATION PART NUMBERS PNL, PNR, & PTX PANEL CUT-OUT DETAILS...
Page 43
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL APPENDIX C EXPLOSION PROOF ELECTRONICS/SENSOR CONFIGURATION PART NUMBER EXX FRONT VIEW Note: The item shown above is an explosion proof oxygen measuring system. The Series 9510 Carbon Dioxide Analyzer is identical to that shown above in terms of mechanical layout.
Page 44
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL APPENDIX C EXPLOSION PROOF ELECTRONICS/SENSOR CONFIGURATION PART NUMBER EXX SIDE VIEW-ELECTRICAL CONNECTIONS Note: The item shown above is an explosion proof oxygen measuring system. The Series 9510 Carbon Dioxide Analyzer is identical to that shown above in terms of mechanical layout. Due to options ordered with the Series 9510 Carbon Dioxide Analyzer, the overall depth of the enclosure may be approximately 2”...
Page 45
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL APPENDIX C EXPLOSION PROOF ELECTRONICS/SENSOR CONFIGURATION PART NUMBER EXX SIDE VIEW-GAS CONNECTIONS Note: The item shown above is an explosion proof oxygen measuring system. The Series 9510 Carbon Dioxide Analyzer is identical to that shown above in terms of mechanical layout.
ALPHA OMEGA INSTRUMENTS CORP. SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL APPENDIX G EXPLOSION PROOF SENSOR ENCLOSURE (REMOTE SENSOR TYPE) Note: The explosion proof remote enclosure shown is identical to the one used for the Series 9510 Carbon Dioxide Analyzer with one exception. The Series 9510 enclosure will not have the...