A P P L I C A T I O N S; M E A S U R I N G Voltage Between T W O Points On A Waveform - Kenwood CS-4125 Instruciton Manual

40/20mhz dual trace oscilloscope
Table of Contents

Advertisement

Because both the v e r t i c a l and horizontal a x i s of the oscilloscope are calibrated,
the oscilloscope i s capable of not only d i s p l a y i n g w a v e f o r m s but c a n also q u a n t i t a -
t i v e l y m e a s u r i n g voltage or t i m e . When performing these latter measurements,
rotate the three V A R I A B L E controls ( C H I [ Y - a x i s ] , C H 2 [ X - a x i s ] a n d h o r i z o n -
tal) a l l the w a y i n the c l o c k w i s e direction to the C A L setting. A l l the o s c i l l o -
scope's V A R I A B L E controls w i l l c l i c k w h e n rotated into their C A L settings.
I n addition, the oscilloscope comes w i t h probes. T h e s e probes should a l l be
plugged into their proper j a c k s i n order to assure a m i n i m u m of interference to
the s i g n a l s y o u w a n t to measure.
M e a s u r i n g V o l t a g e B e t w e e n Two P o i n t s on a W a v e f o r m
Use the following procedures for m e a s u r i n g voltage, etc. between t w o points or
from peak to peak on a w a v e f o r m .
1.
A p p l y a s i g n a l to the I N P U T t e r m i n a l and adjust the V O L T S / D I V a n d
S W E E P T I M E / D I V controls. A l s o reset the trigger point i f n e c e s s a r y . Set
the A C - G N D - D C control at A C .
2.
Work the
^
measured falls on one of the horizontal graduation lines, w h i l e the other
points ( B ) can still be observed on the C R T screen.
3.
Work the 4\ •
at the center of the C R T screen.
4.
Measure the v e r t i c a l distance between the two points and m u l t i p l y that
value by the V O L T S / D I V setting. When u s i n g a probe, also m u l t i p l y the
value by the probe's attenuation rate.
Line up on the center vertical
scale using At-
control
Voltage between 2 points = Vertical distance (div) x VOLTS/DIV setting x Probe attenuation rate
F i g u r e 9. M e a s u r i n g the Voltage Between T w o Points

A P P L I C A T I O N S

P O S I T I O N control so that one of the points ( A ) to be
P O S I T I O N control so that point B falls on the v e r t i c a l scale
POSITION
Two points to be measured
Line up on a horizontal gradation
using * POSITION control
21