TSCD
II Sterile Tubing Welder
Example Standard Operating Procedure
Overview
This document is an example of a standard operating procedure (SOP) for the
TSCD II Sterile Tubing Welder. You may reference this document when writing an
SOP that meets your facility's requirements. For complete information about
using the TSCD II Sterile Tubing Welder, including Warnings and Cautions,
refer to the TSCD
with purchase.
Table of
Page
Contents
Principle of
The collection and processing of blood products often requires manipulation and
Operation
transfer of the blood products among various containers. The TSCD II device offers
technology for connecting sterile blood product containers while maintaining a sterile
fluid pathway and a closed system.
In this system, two lengths of tubing are secured parallel to each other in the clamps
of the TSCD II device. A TSCD wafer that has been heated to approximately 300 °C
(572 °F) cuts through the two lengths of tubing. The section of tubing held in the left
front clamp is moved along the wafer until it is aligned directly across from the
section of tubing held by the right rear clamp. At this point the wafer is retracted, the
clamps push the tubing together, and the two sections of tubing are welded together
by the fusion of polyvinyl chloride (PVC) material.
©2012 Terumo BCT, Inc.
®
II Sterile Tubing Welder Operating Instructions included
1
Principle of Operation
2
2
3
4
4
5
5
6
7
8
9
10
10
11
11
12
12
13
14
Page 1 of 14
PN 306620452
Need help?
Do you have a question about the TSCD II and is the answer not in the manual?
Questions and answers