II Sterile Tubing Welder
Example Standard Operating Procedure
Overview
This document is an example of a standard operating procedure (SOP) for the
TSCD II Sterile Tubing Welder. You may reference this document when writing an
SOP that meets your facility's requirements. For complete information about
using the TSCD II Sterile Tubing Welder, including Warnings and Cautions,
refer to the TSCD
with purchase.
Table of
Page
Contents
Principle of
The collection and processing of blood products often requires manipulation and
Operation
transfer of the blood products among various containers. The TSCD II device offers
technology for connecting sterile blood product containers while maintaining a sterile
fluid pathway and a closed system.
In this system, two lengths of tubing are secured parallel to each other in the clamps
of the TSCD II device. A TSCD wafer that has been heated to approximately 300 °C
(572 °F) cuts through the two lengths of tubing. The section of tubing held in the left
front clamp is moved along the wafer until it is aligned directly across from the
section of tubing held by the right rear clamp. At this point the wafer is retracted, the
clamps push the tubing together, and the two sections of tubing are welded together
by the fusion of polyvinyl chloride (PVC) material.
©2012 Terumo BCT, Inc.
®
II Sterile Tubing Welder Operating Instructions included
1
Principle of Operation
2
2
3
4
4
5
5
6
7
8
9
10
10
11
11
12
12
13
14
 
 
 
 
 
Page 1 of 14
 
 
 
PN 306620452
 

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TSCD II and is the answer not in the manual?

Questions and answers

Summary of Contents for Terumo BCT TSCD II

  • Page 1: Table Of Contents

    In this system, two lengths of tubing are secured parallel to each other in the clamps of the TSCD II device. A TSCD wafer that has been heated to approximately 300 °C (572 °F) cuts through the two lengths of tubing. The section of tubing held in the left front clamp is moved along the wafer until it is aligned directly across from the section of tubing held by the right rear clamp.
  • Page 2: Applications For Use

    “spiking” procedure. The TSCD II device is not to be used with tubing that is connected to a person. This device is for use by trained individuals in such settings as blood bank laboratories and hospitals.
  • Page 3: Operating Procedure Flow Chart

    Pinch the weld site to open the tubing and inspect the weld. Remove and discard the waste segments. Note: This Operating Procedure Flowchart is not intended to replace the procedures provided in the current TSCD II operating instructions.    ...
  • Page 4: Operating Procedures

    “Assembling and Attaching the Bag Support Shelves” on page 8. Connect the power cord to the AC inlet at the back of the TSCD II device, and plug the power cord into a grounded outlet.
  • Page 5: Replacing The Wafer Cassette (Cartridge)

    Place a length of fluid-filled or non-fluid-filled tubing into the rear clamp slot. Tubing must conform to the specifications listed in section 3, “Specifications,” of the operating instructions.                   ©2012 Terumo BCT, Inc. Page 5 of 14 PN 306620452...
  • Page 6: Performing A Weld

    Do not open the clamp covers until the message WELD COMPLETE/ OPEN CLAMP appears. This message indicates that the welding cycle is complete.                   ©2012 Terumo BCT, Inc. Page 6 of 14 PN 306620452...
  • Page 7: Inspecting The Integrity Of The Weld

    Carefully remove the waste segments and discard appropriately. Do not apply pressure or squeeze the segments, because the seals are temporary.                   ©2012 Terumo BCT, Inc. Page 7 of 14 PN 306620452...
  • Page 8: Removing And Attaching The Cover

    Slide the cover down into the hinge sockets until the cover snaps into place. Assembling and Attaching the Bag Support Shelves Note: Do not attempt to lift or transport the TSCD II device using the bag support shelves. Step Action...
  • Page 9 Pull up on the bag support shelf to release the rollers from the slot. Repeat step 1 for the second shelf.                   ©2012 Terumo BCT, Inc. Page 9 of 14 PN 306620452...
  • Page 10: Maintenance Procedures

      Maintenance Procedures To prolong equipment life and for maximum performance of the TSCD II device, recommended maintenance procedures should be performed. Before performing any maintenance, turn off the TSCD II device and disconnect the device from its power source.
  • Page 11: Cleaning The Clamps

    Using either a foam applicator or cotton swab dampened with isopropyl alcohol or an alcohol pad, wipe down the compartment. Ensure that all foreign matter is removed from the compartment.                   ©2012 Terumo BCT, Inc. Page 11 of 14 PN 306620452...
  • Page 12: Replacing The Air Filter

    (SDOT0467). Turn off the power and unplug the power cord from the device. Facing the TSCD II device, remove the filter case from the lower right side of the device by pulling it gently to the right. Remove the air filter from the filter case gently, being careful not to bend the four claws that hold the air filter in place.
  • Page 13: Emptying And Cleaning The Wafer Disposal Box

    Removing and emptying the wafer disposal box Pull on the insert on the front of the box to loosen it from the TSCD II device. Slide the box forward to remove it from the device.
  • Page 14: Setting Parameters

      Inserting the wafer disposal box Insert the lower edge of the disposal box into the front of the TSCD II device. Push the box into the device until it locks into place. Note: The TSCD II device has two photo sensors that align with the slots on the wafer disposal box when it is inserted properly.

Table of Contents

Save PDF