Please read this instruction manual carefully before using the US Pro 2000
2
Edition Portable Ultrasound Unit.
nd
When the preset temperature is reached or the maximum preheat time has
ended (3 minutes), all of the indicators lights will flash five times. Once
complete, the device enters standby mode. This head warming feature takes
approximately three minutes from a cold/room temperature start to nish.
If the warming feature is not needed, press both the "MODE" button and the
"TIME" button simultaneously. The device will go back to standby mode.
When the device is in standby mode, the modulation duty cycle is defaulted
at 5% and the (L) indicator light will be illuminated.
WARNING: During the head warming period, the following items
should be noted:
The device will automatically exit the head warming feature if any
load is detected in the preheating process. Therefore, do not apply
the ultrasound head to the patient during the warming period.
To restart the warming feature, you will have to power o the device
and turn it back on again.
2. Apply transmission gel:
Wash the area to be treated so that it is free of oil and
dirt. Apply a generous layer of ultrasound transmission
gel on the treatment area. The gel acts as a coupling
substance and ensures e ectiveness. The area treated
should be two times the diameter of the sound head.

INSTRUCTIONS FOR USE

1. Turning on the device and head warming feature:
Turn the device on by sliding the power switch upwards
(towards "ON"). The power indicator light will illuminate.
The device will automatically enter the preheat mode.
The six indicator lights will ash alternately during this
period.
9

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PRO Series ELITE DU3037 and is the answer not in the manual?

Questions and answers

Gerard
February 19, 2025

The time indicator doesn’t stop moving also I never noticed it warming up-how does the power mode relate to the time indicator ie; the troubleshooting info states there could be settings issues between time and mode settings

1 comments:
Mr. Anderson
February 19, 2025

The power mode on the LGMedSupply PRO Series ELITE DU3037 is related to the time indicator through the modulation duty cycle settings. The available power modes are Low (5%), Medium (50%), and High (100%), corresponding to output power levels of 0.32W, 3.20W, and 6.40W, respectively. The time indicator allows the user to choose working durations of 5, 10, or 15 minutes.

For troubleshooting settings issues:
1. Ensure the AC/DC adapter is securely connected to both the ultrasound unit and the wall outlet.
2. Verify that the correct working time and modulation duty cycle are selected.
3. Check if the indication lights for time and duty cycle are functioning properly.
4. Confirm that the unit is being used according to the prescribed guidelines (no more than 2-3 times per day for 15-minute intervals).
5. If issues persist, refer to the instruction manual or contact the distributor for further assistance.

This answer is automatically generated