Instructions for use
Status 1080 TM
Always be on the safe side.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Status 1080 TM and is the answer not in the manual?

Questions and answers

Summary of Contents for KaVo Status 1080 TM

  • Page 1 Instructions for use Status 1080 TM Always be on the safe side.
  • Page 2 Distributed by: Manufacturer: KaVo Dental GmbH Kaltenbach & Voigt GmbH Bismarckring 39 Bismarckring 39 D-88400 Biberach D-88400 Biberach Tel. +49 7351 56-0 www.kavo.com Fax +49 7351 56-1488...
  • Page 3: Table Of Contents

    Instructions for use Status 1080 TM Table of Contents Table of Contents User instructions ............................3 1.1 User guide ............................3 1.1.1 Abbreviations ........................3 1.1.2 Symbols ..........................3 1.1.3 Target group ......................... 3 1.2 Service .............................. 3 1.3 Warranty terms and conditions ......................3 1.4 Transportation and storage .......................
  • Page 4 Instructions for use Status 1080 TM Table of Contents 4.7.1 Select memory level Dentist 1 or Dentist 2 ................ 47 4.7.2 Fill the tumbler and rinse the spittoon ................47 4.7.3 Turn the x-ray image viewer on and off ................48 4.7.4...
  • Page 5: User Instructions

    Additional information can be obtained at: www.kavo.comAdditional information can be obtained at: www.kavo.com 1.3 Warranty terms and conditions KaVo provides the end customer with a warranty that the product cited in the hand‐ over certificate will function properly and guarantees zero defects in the material or 3/74...
  • Page 6: Transportation And Storage

    Claims from this warranty can only be asserted when the transfer form (copy) be‐ longing to the product has been sent to KaVo, and the original can be presented by the operator or user.
  • Page 7: Information On The Packaging: Storage And Transportation

    ADSp. Art. 28). Outside Germany Note KaVo is not liable for damage arising from transportation. Immediately inspect the delivery after receipt! If the packaging is visibly damaged on delivery, please proceed as follows: 1. The recipient of the package must record the loss or damage on the delivery receipt.
  • Page 8 Instructions for use Status 1080 TM 1 User instructions | 1.4 Transportation and storage Transport upright with the arrows pointing upwards Fragile - protect against impact. Protect from moisture. Permissible stacking load Temperature range Humidity Air pressure 6/74...
  • Page 9: Safety

    The user must ensure that that the unit works properly and is in a satisfactory con‐ dition before each use. The KaVo Status 1080 TM equipment system is a dental treatment unit in accord‐ ance with ISO 7494, with a dental chair in accordance with ISO 6875. This KaVo...
  • Page 10 The overarching guidelines and/or national laws, national regulations and the rules of technology applicable to medical devices for start-up and use of the KaVo product for the intended purpose are to be applied and complied with.
  • Page 11 This must be done in strict com‐ pliance with all applicable national regulations. Questions on proper disposal of the KaVo product can be answered by the KaVo branch. Disposal of electronic and electrical devices Note According to EC directive 2002/96 concerning used electrical and electronic devi‐...
  • Page 12: Product-Specific

    2.2.2 Product-specific Use and target group KaVo Status 1080 TM is used for the dental treatment of children and adults. The device may only be used by medical professionals. For technician jobs that require greater pressure than when working in the oral cavity such as grinding prostheses, etc., a special technician's machine must be used.
  • Page 13: Product Specific

    Instructions for use Status 1080 TM 2 Safety | 2.3 Safety instructions CAUTION Insufficient hygiene. Transmission of pathogens. ▶ To maintain the hygienic system, observe the care and servicing instructions. WARNING Injury or damage from damaged functional parts. When functional parts are damaged, it can cause additional damage or personal injury.
  • Page 14 Instructions for use Status 1080 TM 2 Safety | 2.3 Safety instructions CAUTION Germ formation. Infections. ▶ Before starting, rinse all the water drain lines without instruments. ▶ Before startup and after the device has not been used for a while (weekends, holidays, vacations, etc.), rinse or purge the air and water lines.
  • Page 15 Instructions for use Status 1080 TM 2 Safety | 2.3 Safety instructions CAUTION Danger of injury from overload. The patient chair can collapse. ▶ Do not subject the patient chair to a load above its limit (135 kg). CAUTION Hair may become caught in the patient's chair when the headrest is moved.
  • Page 16: Product Description

    Instructions for use Status 1080 TM 3 Product description | 3.1 Treatment unit 3 Product description 3.1 Treatment unit ① Standard headrest ⑤ Armrest (supplementary equipment) ② SOMATIC headrest ⑥ Basic chair ③ 2-joint headrest ⑦ Comfort Backrest ④ Bench ⑧...
  • Page 17: Unit Base

    Instructions for use Status 1080 TM 3 Product description | 3.2 Unit base 3.2 Unit base ① Pivotable swing arm for holding the ④ Main fuse patient unit, device body and assis‐ tant unit ② Supply unit, customer connection of ⑤...
  • Page 18: Unit Body With Patient's Part

    Instructions for use Status 1080 TM 3 Product description | 3.3 Unit body with patient's part 3.3 Unit body with patient's part ① Patient's part ④ Unit body The central control is housed in the unit body. ② Tumbler filler ⑤...
  • Page 19: Dentist's Unit

    Instructions for use Status 1080 TM 3 Product description | 3.4 Dentist's unit 3.4 Dentist's unit ① Keypad and display panel ⑥ Three-function or multi-function handpiece ② Memospeed, Memodent ⑦ PIEZOlux ③ Locking brake ⑧ Either INTRA Lux K 200 motor or IN‐...
  • Page 20: Assistant Unit

    Instructions for use Status 1080 TM 3 Product description | 3.5 Assistant unit 3.5 Assistant unit ① Three-function handpiece or multi‐ ④ Saliva ejector functional handpiece ② Spray mist suction ⑤ Satelec Mini LED ③ Key and display field 18/74...
  • Page 21: Keypads

    Instructions for use Status 1080 TM 3 Product description | 3.6 Keypads 3.6 Keypads The key fields on the dentist unit and assistant unit are the same. ① Patient chair ② function keys Group of keys for the patient chair...
  • Page 22: Footswitch

    Instructions for use Status 1080 TM 3 Product description | 3.7 Footswitch Designation Display LED Tumbler filler Cold light (on instruments)/ green Treatment unit ON when the instruments are moun‐ Spray preselection (on re‐ green/yellow moved instruments) Counterclockwise motor rotation 3.7 Footswitch...
  • Page 23: Technical Data

    Instructions for use Status 1080 TM 3 Product description | 3.8 Technical Data 3.8 Technical Data Drilling template and setup plan Drilling template (paper) Page 001 (Mat. no. 1.000.3001) Drilling template (sheet met‐ al)Mat. no. ). 10.002492 Layout plan (Mat. no. 1.000.3001)
  • Page 24 >45 mbar, recommended: 60 mbar Suction vacuum flow 500 Nl/min. The values apply to the KaVo measuring set (Mat. no. 0.411.8500) Operating environment Floor quality The quality of the floor must comply with the load bearing capacity for buildings DIN 1055 page 3 and the pressure re‐...
  • Page 25: Rating And Name Plates

    Instructions for use Status 1080 TM 3 Product description | 3.9 Rating and name plates Transportation and storage conditions Ambient temperature -20 to +55°C Relative humidity 5% to 95% non-condensing Air pressure 700 to 1,060 hPa Weight Treatment unit 282 kg gross ,242 kg net 3.9 Rating and name plates...
  • Page 26 Instructions for use Status 1080 TM 3 Product description | 3.9 Rating and name plates serial number Read and note the content of accompanying documents. Classification The handpieces are type BF application parts The patient chair is a type B application part...
  • Page 27: Operation

    Instructions for use Status 1080 TM 4 Operation | 4.1 Switching the device on and off 4 Operation 4.1 Switching the device on and off The main switch has the following tasks: ▪ Plug in or disconnect all the pins of the unit's plug into the customer's power supply.
  • Page 28: Converting From Right Handed To Left Handed

    Instructions for use Status 1080 TM 4 Operation | 4.2 Converting from right handed to left handed 4.2 Converting from right handed to left handed CAUTION Collision and damage to the unit. ▶ Ensure sufficient space for swinging. ▶ Turn off the device before conversion.
  • Page 29 Instructions for use Status 1080 TM 4 Operation | 4.2 Converting from right handed to left handed ▶ Remove the seat. ▶ Carefully place the foot control in the spittoon bowl. ▶ Use protective base. ▶ Remove the cover ①.
  • Page 30 Instructions for use Status 1080 TM 4 Operation | 4.2 Converting from right handed to left handed ▶ Lift the lock ② and simultaneously rotate the support arm toward the foot end. CAUTION Let the instrument hose hang on the patient chair or another object.
  • Page 31 Instructions for use Status 1080 TM 4 Operation | 4.2 Converting from right handed to left handed ▶ Remove the lock ③, and simultaneously rotate the device body toward the backrest. 29/74...
  • Page 32 Instructions for use Status 1080 TM 4 Operation | 4.2 Converting from right handed to left handed ▶ The support arm and device body are fixed when they audibly lock in place. ▶ Swing the dentist's unit into position, and rotate the patient's unit ① 360°.
  • Page 33 Instructions for use Status 1080 TM 4 Operation | 4.2 Converting from right handed to left handed WARNING Safety lever not changed over. Collision of the armrests with the device body. ▶ Change over the safety lever. ▶ Change over the safety lever ①.
  • Page 34: Adjusting The Dental Chair

    Instructions for use Status 1080 TM 4 Operation | 4.3 Adjusting the dental chair 4.3 Adjusting the dental chair 4.3.1 Swing the armrest (extra) To make it easier to get in and out, the armrest can be swung outward. ▶ Lift the armrest upward and swing it outward.
  • Page 35: Adjust Head Rest

    Instructions for use Status 1080 TM 4 Operation | 4.3 Adjusting the dental chair 4.3.3 Adjust head rest Standard headrest CAUTION Adjusting the headrest. Injury of neck muscles. ▶ Make sure that the patient is aware of the headrest setting.
  • Page 36 Instructions for use Status 1080 TM 4 Operation | 4.3 Adjusting the dental chair Swing the headrest ▶ Push the catch up ②, and swing the headrest into position. ▶ To lock the headrest, apply slight pressure (arrowA). SOMATIC headrest CAUTION Adjusting the headrest.
  • Page 37 Instructions for use Status 1080 TM 4 Operation | 4.3 Adjusting the dental chair Adjust the brake for the lengthwise headrest shift ▶ Remove the cover ③. ▶ Adjust the braking force using hexagon socket screw WAF 3 mm. ▶ Remount the cover ③.
  • Page 38 Instructions for use Status 1080 TM 4 Operation | 4.3 Adjusting the dental chair 2-joint headrest CAUTION Adjusting the headrest. Injury of neck muscles. ▶ Make sure that the patient is aware of the headrest setting. ▶ Patients need to raise their head slightly during adjustment.
  • Page 39: Turning The Head Cushion

    Instructions for use Status 1080 TM 4 Operation | 4.3 Adjusting the dental chair Swing the headrest ▶ Push the lock ⑥ together, and swing headrest into the desired position. When swinging the headrest back into position, make sure that there is nothing in area B.
  • Page 40: Adjust Chair Position

    Instructions for use Status 1080 TM 4 Operation | 4.3 Adjusting the dental chair 4.3.5 Adjust chair position Gradually adjust the chair position Note The functions of the control keys on the dentist unit and assistant unit are the same.
  • Page 41 Instructions for use Status 1080 TM 4 Operation | 4.3 Adjusting the dental chair Note The seat and backrest can be moved at the same time. When the operating voltage for the patient chair is below 200 V, the movements in the automatic program can only be sequential.
  • Page 42 Instructions for use Status 1080 TM 4 Operation | 4.3 Adjusting the dental chair The chair positions can be saved on two footswitches; the standard setting is as follows: ▪ "Spray default" footswitch: "LP" automatic position (last position) ▪ "Blown air" footswitch: "SP" automatic position (rinsing position) The footswitch assignment can be changed.
  • Page 43 Instructions for use Status 1080 TM 4 Operation | 4.3 Adjusting the dental chair Retrieve the chair positions using the foot control Note If an instrument is removed, the chair functions of the foot control are blocked. The blocking can be removed by briefly pressing the stirrup switch. The functions are then available.
  • Page 44: Safety Shut-Off

    Instructions for use Status 1080 TM 4 Operation | 4.3 Adjusting the dental chair The chair moves into the automatic position. You can also use the cross switch. Lock the chair functions when the instrument is removed ▶ Press the foot pedal.
  • Page 45 Instructions for use Status 1080 TM 4 Operation | 4.3 Adjusting the dental chair The safety cut-offs can be found at the following places on the treatment unit. Check the patient chair and foot control safety shutoffs ① Backrest ④ Armrest ②...
  • Page 46 Instructions for use Status 1080 TM 4 Operation | 4.3 Adjusting the dental chair Safety shut-off on patient's part ① Patient's part pivoted over dental chair The safety shutoff occurs went a movement angle has been exceeded, or part of the treatment unit collides with an object.
  • Page 47: Special Chair Functions

    Instructions for use Status 1080 TM 4 Operation | 4.4 Moving dentist's unit 4.3.7 Special chair functions Note With operating voltages greater than 200 V, the spindle motors can operate simul‐ taneously, i.e., the chair and backrest can move simultaneously in the automatic program.
  • Page 48: Adjust The Height Of The Assistant Unit

    Instructions for use Status 1080 TM 4 Operation | 4.5 Adjust the height of the assistant unit 4.5 Adjust the height of the assistant unit The assistant's unit can be positioned vertically in three levels. ▶ To set a higher level, pull the assistant unit upward gently until it audibly locks in place.
  • Page 49: Adjusting Functions

    Instructions for use Status 1080 TM 4 Operation | 4.7 Adjusting functions Note When the patient unit is swung over the patient chair, the safety shutoff is activated. 4.7 Adjusting functions 4.7.1 Select memory level Dentist 1 or Dentist 2...
  • Page 50: Turn The X-Ray Image Viewer On And Off

    Instructions for use Status 1080 TM 4 Operation | 4.7 Adjusting functions The spittoon is rinsed. ▶ To stop rinsing before the set time, press the "Spittoon" button again. 4.7.3 Turn the x-ray image viewer on and off ▶ Press the "X-ray viewer" button.
  • Page 51: Use The Function Keys (Only With The Memospeed)

    Instructions for use Status 1080 TM 4 Operation | 4.7 Adjusting functions Using the timer Setting the timer The minimum timer time is 30 seconds, and the maximum timer time is 8 minutes. ▶ Press the "Timer" button until you hear a beep.
  • Page 52: Using Instruments

    Instructions for use Status 1080 TM 4 Operation | 4.8 Using instruments 4.8 Using instruments The following sections describe the use and setting of the instruments. A distinction is drawn between operation with and without the installed Memospeed. The instruments are prevented from simultaneous use by software that determines when they are mounted.
  • Page 53: Operating Turbine

    Instructions for use Status 1080 TM 4 Operation | 4.8 Using instruments ▶ Press the foot pedal several times when the instrument is mounted until the desired level is displayed. Memospeed Display ① Level display ② Time If stored values are changed during operation, these changed values will be lost when the unit is switched off and back on again.
  • Page 54 Instructions for use Status 1080 TM 4 Operation | 4.8 Using instruments E123 ▶ Press the "Level selection" button until you hear a beep. ▶ To switch between maximum and minimum mode, press the "Increase value" or "Reduce value" button.
  • Page 55: Using The Intra Lux Motor K 200 And Intra Lux Motor Kl 701

    Instructions for use Status 1080 TM 4 Operation | 4.8 Using instruments ▶ "Preselected spray" footswitch. Two LEDs shine when the spray cooling level is activated. Storing values ▶ Press the "Level selection" button until you hear a beep. E123 The set values are saved for the set memory level and the set dentist level.
  • Page 56 Instructions for use Status 1080 TM 4 Operation | 4.8 Using instruments Setting the rotational direction of the motor Note Only change the rotary direction of the motor when the motor is not running. ▶ Remove the motor from the holder.
  • Page 57: Operating The Piezolux

    Instructions for use Status 1080 TM 4 Operation | 4.8 Using instruments The display switches to the setting menu for maximum. ▶ Press the "Increase value" or" Decrease value" button until the desired value is displayed. ▶ Press the "Preselect level" key.
  • Page 58 Instructions for use Status 1080 TM 4 Operation | 4.8 Using instruments Note The torque wrench is subject to natural wear and should be changed when it longer works properly. (Mat. no. 1.000.4887) The following settings can be changed: ▪ Water emission ▪...
  • Page 59 Instructions for use Status 1080 TM 4 Operation | 4.8 Using instruments E123 ▶ Press the "Preselect level" button until you hear a tone. ▶ Press the "Increase value" or" Decrease value" button until the desired value is set. Changing the high-pressure lamp CAUTION Injury hazard from hot valve body.
  • Page 60: Operating Three-Function And Multi-Function Handpiece

    Instructions for use Status 1080 TM 4 Operation | 4.8 Using instruments 4.8.6 Operating three-function and multi-function handpiece ① "Air" button ③ "Water" button ② Cannula CAUTION Danger of injury from cannulas that are worn or not locked into place.
  • Page 61 Instructions for use Status 1080 TM 4 Operation | 4.8 Using instruments Removing the cannulas ▶ Remove the cannula with a slight amount of rotation while holding the handpiece on the grip sleeve. Setting cold light intensity The cold light intensity is adjusted in the same manner as with the turbine.
  • Page 62: Setting The Spray Mist Suction Device And Saliva Ejector

    Instructions for use Status 1080 TM 4 Operation | 4.8 Using instruments 4.8.7 Setting the spray mist suction device and saliva ejector CAUTION Unintentionally activating the saliva ejector or spray fog suction. Injury in the area of the mouth. ▶ When using the vacuum stop to interrupt suction, do not leave the saliva ejector or spray fog suction in the patient's mouth.
  • Page 63: Preparation Methods Din En Iso 17664

    Instructions for use Status 1080 TM 5 Preparation methods DIN EN ISO 17664 5 Preparation methods DIN EN ISO 17664 Note The preparation methods can be found in the care instructions. 61/74...
  • Page 64: Adaptable Accessories And Kits

    Instructions for use Status 1080 TM 6 Adaptable accessories and kits | 6.1 Unit base 6 Adaptable accessories and kits 6.1 Unit base Designation Description DVGW water block Water block, compact Without DVGW Water bottle DVGW Intensive disinfection kit Steel support base...
  • Page 65: Dentist Unit

    Instructions for use Status 1080 TM 6 Adaptable accessories and kits | 6.4 Dentist unit 6.4 Dentist unit Designation Description Multiflex LUX hose For connecting the turbine and SONICflex. LUX motor hose and IN‐ For connecting the INTRA LUX motor K 200 and INTRA TRA K and INTRA KL mo‐...
  • Page 66: Troubleshooting

    Instructions for use Status 1080 TM 7 Troubleshooting 7 Troubleshooting Note In case of malfunctions, consult the separate instructions for the use and care of the individual instruments (such as the turbine, motor, camera, Satelec Mini LED, etc.). Malfunction Cause Remedy Nothing works.
  • Page 67 Instructions for use Status 1080 TM 7 Troubleshooting Malfunction Cause Remedy Sieves in suction connector ▶ Exchange the sieve. blocked. Foot switch for vacu-stop has been ▶ Release the foot switch. pressed. Suction machine not running. ▶ Turn on the suction machine.
  • Page 68: Carrying Out Safety Checks

    ▪ Type BF comment 2 ▪ Measurement according to EUL/EPA KaVo offers a medical device book for keeping an inventory and recording essential master data on the medical device. The medical device book is only required in Germany and is therefore only in German (Mat. no. 0.789.0480).
  • Page 69 Instructions for use Status 1080 TM 8 Carrying out safety checks | 8.1 Measuring protective cable resistance Note Take into account the additional measuring points of additional equipment such as connection to equipment by a different manufacturer, treatment lamps, multimedia system, etc.
  • Page 70: Measure Equivalent Unit Leakage Current

    Instructions for use Status 1080 TM 8 Carrying out safety checks | 8.2 Measure equivalent unit leakage current 8.2 Measure equivalent unit leakage current Threshold: < 10 mA The equivalent unit leakage current (EUL) can be measured using an EPA meas‐...
  • Page 71: Measuring Dummy Patient Leakage Current

    Instructions for use Status 1080 TM 8 Carrying out safety checks | 8.3 Measuring dummy patient leakage current Note Take into account the additional measuring points of additional equipment such as connection to equipment by a different manufacturer, treatment lamps, multimedia system, etc.
  • Page 72 Instructions for use Status 1080 TM 8 Carrying out safety checks | 8.3 Measuring dummy patient leakage current Note Take into account the additional measuring points of additional equipment such as connection to equipment by a different manufacturer, treatment lamps, multimedia system, etc.
  • Page 73 Instructions for use Status 1080 TM 8 Carrying out safety checks | 8.3 Measuring dummy patient leakage current ▶ Check the EPA at the measuring points or with an EPA measuring line Mat. no. 1.001.9904. 71/74...
  • Page 74: Details About Electromagnetic Compatibility According To En60601-1-2

    9.1 Electromagnetic Transmissions The Status 1080 TM treatment unit is for use in an environment like the one cited below. The customer or user of the Status 1080 TM should ensure that it is used in the correct environment. Measurements of noise transmis‐...
  • Page 75: Recommended Safe Distance Between Portable And Mobile Hf Telecommunications Equipment And The Treatment Unit

    9.3 Recommended safe distance between portable and mobile HF telecommunications equipment and the treatment unit The Status 1080 TM is intended for use in an electromagnetic environment in which HF disturbances are controlled. The customer or the user of the Status 1080 TM can help prevent electromagnetic disturbances by maintaining the minimum dis‐...
  • Page 76 Instructions for use Status 1080 TM 9 Details about electromagnetic compatibility according to EN60601-1-2 | 9.3 Recommended safe distance between portable and mobile HF telecommunications equipment and the treatment unit NOTE 2: These guidelines may not be applicable in every case. The spread of electromagnetic waves is absorbed and reflected by buildings, objects and people.