Page 1
Infusomat ® Instructions for Use Software IFME/IFMe...
Page 2
Patient Safety Attention: Consult accompanying documents! Read Instructions for Use prior to use. Application only under regularly supervision by specially trained staff. ➤ Operation Connected electrical equipment must comply ➤ Ensure the unit is properly positioned and with the relevant IEC/EN-publications secured.
Overview Handle For easy transport. Display All important information in plain text. Green background illumination only if connected to mains or a key is pushed at battery operation. Keypad for Input Correct Input/CLEAR Display reset to 000.0 Red LED indicates alarm condition Additionally "AAA.A”...
Page 5
Guide for Short Infusion Pole Multi-Function- Connector (MFC) Mains/Power Connection Connection for staff (protect against ingress of moisture). In case of call, ambulances mains/power failure, the pump switches to (12 V DC) and inter- battery operation. Battery operation time: face for fm anaes- >...
Operation Infusion upper clip 1. Ensure reliable installation Never position infusion bottle below pump level. ➤ Connect staff call. ➤ Insert spike vertically into infusion bottle. Fill lower part of drop chamber to max. 2/3. ➤ Open roller clamp. Original 2.
Additional Settings To Change the Delivery Rate volume has to be set. ➤ Press START/STOP. Green operating indicator goes out. ➤ Press C. Target Time (Time Pre-selection) Display reset to 000.0 ➤ Press key below TIME. ➤ Enter new delivery rate. (No setting possible, Enter target time via keypad (50 min = 50;...
Recording the data: In connection with the software “IFME” the The protocol of events only can be transferred Infusomat fmS is equipped with a history func- to a computer via interface (MFC interface tion (memory of events). This permanent lead). For this a terminal programme installed in...
Page 9
Special Functions Select Special Mode SF Depending on the version, several functions may Interval bolus be deactivated. Contact service. Automatic bolus in set time interval. In case of ➤ To set special functions press key below SF manual bolus administration the interval bolus repeatedly until desired special function is is skipped.
Page 10
Special Functions ➤ Drug Display Exit from selection: Press key below END. 9 selectable drug names can be stored (input via service program only). Check Battery Capacity Remaining battery life time is displayed, ➤ Select with key below (+). Drug is displayed, e.g.
Page 11
Dosage Calculation (Overview) values (rate or dosage) are displayed. The dosage calculation automatically calculates ➤ First check rate displayed for plausibility, the delivery rate in ml/h. then confirm RATE. Value is displayed. Setting parameters: ➤ Start infusion. 1. Concentration Note: Dosage value with unit is displayed (down - per ml or on the right).
Page 12
Special Functions ➤ Check entered or calculated rate for plausi- Secondary infusion bility and confirm with key below RATE. Rate is e.g. 100 ml Primary Infusion bag volume with displayed. e.g. 1000 ml delivery rate 10 ml/h bag volume with ➤...
Alarms Operating alarms - audible alarm every 4 sec., - operating indicator off, - staff call. Remedy failure and restart infusion. ➤ Switch off device. Drop Alarm / Pressure Alarm ➤ Connect to mains/power or 12 V DC power Infusion bottle empty? source.
Page 14
The delivery rate and a drug can be entered in matically will be switched to integrated the Infusomat® fmS as "proposal” via an rechargeable battery. external computer. Both items of data must be checked on the Infusomat fmS and acknow- Charge Battery ledged. ➤ Charge battery in case of: ➤...
Technical Data Type of unit Volumetric infusion pump Classification (acc. to IEC/EN 60601-1) defibrillator-proof, CF equipment Protection Class I; IP 22 (Moisture protection: drip protected) Class (acc. to Directive 93/42 EEC) II b Rated voltage / current 230 V AC~ (0.06 A), 50/60 Hz Mains fuse T 0.16 A or 200 V/230 V/240 V AC~ (0.06 A), 50/60 Hz...
Page 17
Air detector Technical sensitivity Air bubbles > 0.01 ml Alarm triggering: With air bubble size of typ. 0.3 ml (limit value 0.4 ml) or 1.5 ml/h (cumulative value of 1 h as of air bubble volume 0.01 ml) Can be set from 0.01 to 0.3 ml via service program only Can be set from 0.5 to 3.5 ml/h via service program only Accuracy of set delivery rate typ.
Warranty provided concerning waste disposal and hygiene for batteries and disposables. B. Braun provides as from the date of delivery a warranty of 2 years for every Infusomat® fmS. This covers repair or replacement of parts dama- ged as a result of design/manufacturing errors or material defects.
Ordering Art.-Nr. Infusomat® fmS 230 V 871 5548 Infusomat® fmS 200 - 240 V 871 5440 Infusomat® fmS 100 – 120 V 871 5416 Recommended accessories for Infusomat® fmS Connecting lead for potential equalisation 870 1628 MFC-Connecting lead for staff call 871 1682 MFC-Connecting lead for ambulance cars (12 V DC) 871 1674...